header advert
Results 1 - 3 of 3
Results per page:
Bone & Joint Research
Vol. 6, Issue 2 | Pages 113 - 122
1 Feb 2017
Scholes SC Hunt BJ Richardson VM Langton DJ Smith E Joyce TJ

Objectives

The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.

Methods

In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 37 - 43
1 Jan 2009
Hassaballa M Mehendale S Poniatowski S Kalantzis G Smith E Learmonth ID

Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of > 10 mm in five (7.2%).

The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 3 | Pages 450 - 452
1 May 1993
Sarangi P Ward A Smith E Staddon G Atkins R

We made a prospective study of the incidence and natural history of algodystrophy and associated changes in bone mineral density in the ankles and feet of 60 consecutive patients who had suffered unilateral fractures of the tibial shaft. At bone union, 18 patients showed signs of algodystrophy. Its development was independent of the type of fracture management and of the severity of injury. Patients with algodystrophy lost significantly more bone mineral than did those without but the degree of this loss was independent of the type of treatment and of the time to fracture union. In most cases the symptoms resolved within six months of fracture union but in four patients they were still present at one year and two of these had not returned to work.