header advert
Results 1 - 3 of 3
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 615 - 618
1 May 2012
ten Have BLEF Brouwer MD RW van Biezen FC Verhaar JAN

The purpose of this prospective study was to evaluate the long-term clinical and radiological outcomes of revision of the femoral component of a total hip replacement using impaction bone grafting. Femoral revision with an impacted allograft was performed on 29 patients (31 hips). In all, 21 hips (68%) had grade III or IV femoral defects according to the Endo-Klinik classification. A total of 11 patients (12 hips) died before the ten-year follow-up period. Of the remaining patients, 18 patients (19 hips) were followed for 10 to 15 years; three further patients died during this time. None of the 31 stems underwent further revision of their stem. However, four stems showed extensive subsidence (> 15 mm). One of these patients had a femoral fracture that required fixation. Three other patients had a femoral fracture, two of which required fixation and the other was treated conservatively. Patients with a femoral fracture and/or severe subsidence had significantly more grade IV defects (six of seven hips; p = 0.004). One patient needed a closed reduction for dislocation.

Impaction allografting in revision hip surgery gives good long-term results for femora with grades I, II and III Endo-Klinik-classified defects. Extensive subsidence and femoral fractures were seen mainly in patients with grade IV damaged femora.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1454 - 1459
1 Nov 2006
Brouwer RW Bierma-Zeinstra SMA van Raaij TM Verhaar JAN

A prospective, randomised, controlled trial compared two different techniques of high tibial osteotomy with a lateral closing wedge or a medial opening wedge, stabilised by a Puddu plate. The clinical outcome and radiological results were examined at one year.

The primary outcome measure was the achievement of an overcorrection of valgus of 4°. Secondary outcome measures were the severity of pain (visual analogue scale), knee function (Hospital for Special Surgery score), and walking distance.

Between January 2001 and April 2004, 92 patients were randomised to one or other of the techniques. At follow-up at one year the post-operative hip-knee-ankle angle was 3.4° (± 3.6° sd) valgus after a closing wedge and 1.3° (± 4.7° sd) of valgus after an opening wedge. The adjusted mean difference of 2.1° was significant (p = 0.02). The deviation from 4° of valgus alignment was 2.7° (± 2.4° sd) in the closing wedge and 4.0° (± 3.6° sd) in the opening-wedge groups. The adjusted mean difference of 1.67° was also significant (p = 0.01).

The severity of pain, knee score and walking ability improved in both groups, but the difference was not significant.

Because of pain, the staples required removal in 11 (23%) patients in the closing-wedge group and a Puddu plate was removed in 27 (60%) patients in the opening-wedge group. This difference was significant (p < 0.001).

We conclude that closing-wedge osteotomy achieves a more accurate correction with less morbidity, although both techniques had improved the function of the knee at one year after the procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1227 - 1232
1 Sep 2005
Brouwer RW Bierma-Zeinstra SMA van Koeveringe AJ Verhaar JAN

Our aim was to compare the degree of patellar descent and alteration in angle of the inclination of the tibial plateau in lateral closing-wedge and medial opening-wedge high tibial osteotomy (HTO) in 51 consecutive patients with osteoarthritis of the medial compartment and varus malalignment. Patellar height was measured by the Insall-Salvati (IS) and the Blackburne-Peel (BP) ratios. The tibial inclination was determined by the Moore-Harvey (MH) method. Multivariate linear regression analysis was used to determine the influence of the type of HTO (closing vs opening wedge) on the post-operative patellar height or tibial inclination. The intra- and interobserver variability of these methods was determined before operation and at follow-up at one year.

After an opening-wedge HTO the patellar height was significantly more decreased (mean post-operative difference: IS = 0.15; 95% confidence interval (CI) 0.06 to 0.23; BP = 0.11; 95% CI 0.05 to 0.18) compared with a closing-wedge HTO.

The angle of tibial inclination differed significantly (mean post-operative difference MH = −6.40°; 95% CI −8.74 to −4.02) between the two HTO techniques, increasing after opening-wedge HTO and decreasing after closing-wedge HTO.

There was no clinically-relevant difference in the intra- and interobserver variability of measurements of patellar height either before or after HTO.