Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known.

At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%).

The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1299 - 1302
1 Oct 2007
Ramiah RD Ashmore AM Whitley E Bannister GC

We determined the ten-year life expectancy of 5831 patients who had undergone 6653 elective primary total hip replacements at a regional orthopaedic centre between April 1993 and October 2004. Using hospital, general practitioner and the local health authority records, we recorded the dates of death for those who died following surgery.

The mean age at operation was 67 years (13 to 96) with a male:female ratio of 2:3. Of 1154 patients with a ten-year follow-up 340 (29.5%) had died a mean of 5.6 years (0 to 10) after surgery. Using Kaplan-Meier curves, the ten-year survival was 89% in patients under 65 years at surgery, 75% in patients aged between 65 and 74 years, and 51% in patients over 75.

The standardised mortality rates were considerably higher for patients under 45 years, 20% higher for those between 45 and 64 years, and steadily reduced in patients aged 65 and over.

The survival of cemented hip replacement derived from the Swedish Hip Arthroplasty Register Annual Report 2004 exceeds the life expectancy of patients over the age of 60 in our area, suggesting that cemented hip replacement is the procedure of choice in this population.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1584 - 1590
1 Dec 2006
Hook S Moulder E Yates PJ Burston BJ Whitley E Bannister GC

We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up.

The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips.

All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure.

Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001).

The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis.

Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.