The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1561 - 1567
1 Nov 2005
Janssen D Aquarius R Stolk J Verdonschot N

The Capital Hip implant was a Charnley-based system which included a flanged and a roundback stem, both of which were available in stainless steel and titanium. The system was withdrawn from the market because of its inferior performance. However, all four of the designs did not produce poor rates of survival. Using a simulated-based, finite-element analysis, we have analysed the Capital Hip system. Our aim was to investigate whether our simulation was able to detect differences which could account for the varying survival between the Capital Hip designs, thereby further validating the simulation.

We created finite-element models of reconstructions with the flanged and roundback Capital Hips. A loading history was applied representing normal walking and stair-climbing, while we monitored the formation of fatigue cracks in the cement.

Corresponding to the clinical findings, our simulation was able to detect the negative effects of the titanium material and the flanged design in the Capital Hip system. Although improvements could be made by including the effect of the roughness of the surface of the stem, our study increased the value of the model as a predictive tool for determining failure of an implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 3 | Pages 436 - 444
1 Apr 2000
van Loon CJM de Waal Malefijt MC Buma P Stolk PWT Verdonschot N Tromp AM Huiskes R Barneveld A

The properties of impacted morsellised bone graft (MBG) in revision total knee arthroplasty (TKA) were studied in 12 horses. The left hind metatarsophalangeal joint was replaced by a human TKA. The horses were then randomly divided into graft and control groups. In the graft group, a unicondylar, lateral uncontained defect was created in the third metatarsal bone and reconstructed using autologous MBG before cementing the TKA. In the control group, a cemented TKA was implanted without the bone resection and grafting procedure. After four to eight months, the animals were killed and a biomechanical loading test was performed with a cyclic load equivalent to the horse’s body-weight to study mechanical stability. After removal of the prosthesis, the distal third metatarsal bone was studied radiologically, histologically and by quantitative and micro CT.

Biomechanical testing showed that the differences in deformation between the graft and the control condyles were not significant for either elastic or time-dependent deformations. The differences in bone mineral density (BMD) between the graft and the control condyles were not significant. The BMD of the MBG was significantly lower than that in the other regions in the same limb. Micro CT showed a significant difference in the degree of anisotropy between the graft and host bone, even although the structure of the area of the MBG had trabecular orientation in the direction of the axial load. Histological analysis revealed that all the grafts were revascularised and completely incorporated into a new trabecular structure with few or no remnants of graft. Our study provides a basis for the clinical application of this technique with MBG in revision TKA.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 6 | Pages 1052 - 1057
1 Nov 1999
Giesen EBW Lamerigts NMP Verdonschot N Buma P Schreurs BW Huiskes R

The use of impacted, morsellised bone grafts has become popular in revision total hip arthroplasty (THA). The initial stability of the reconstruction and the effectiveness of any subsequent process of revitalisation and incorporation will depend on the mechanical integrity of the graft. Our aim in this study was to document the time-dependent mechanical properties of the morsellised graft. This information is useful in clinical application of the graft, in studies of migration of the implant and in the design of the joint.

We used 16 specimens of impacted, morsellised cancellous bone from the sternum of goats to assess the mechanical properties by confined compression creep tests. Consideration of the graft material as a porous, permeable solid, filled with fluid, allowed determination of the compressive modulus of the matrix, and its permeability to fluid flow.

In all specimens the compression tests showed large, irreversible deformations, caused by flow-independent creep behaviour as a result of rolling and sliding of the bone chips. The mean permeability was 8.82 *10−12 m4/Ns (SD 43%), and the compressive modulus was 38.7 MPa (SD 34%). No correlation was found between the apparent density and the permeability or between the apparent density and the compressive modulus. The irreversible deformations in the graft could be captured by a creep law, for which the parameters were quantified.

We conclude that in clinical use the graft is bound to be subject to permanent deformation after operation. The permeability of the material is relatively high compared with, for example, human cartilage. The confined compression modulus is relatively low compared with cancellous bone of the same apparent density. Designs of prostheses used in revision surgery must accommodate the viscoelastic and permanent deformations in the graft without causing loosening at the interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 665 - 669
1 Jul 1997
Verdonschot N Huiskes R

It has been suggested that the endurance of cemented femoral reconstructions in total hip arthroplasty is affected by the creep of acrylic cement, but it is not known to what extent cement creeps under loading conditions in vivo, or how this affects load transfer. We have simulated the long-term creep properties of acrylic cement in finite-element models of femoral stem constructs and analysed their effects. We investigated whether subsidence rates measured in vivo could be explained by creep of acrylic cement, and if polished, unbonded, stems accommodated creep better than bonded stems.

Our findings showed that polished prostheses subsided only about 50 μm as a result of cement creep. The long-term prosthetic subsidence rates caused by creep of acrylic cement are therefore very small and do not explain the excessive migration rates which have sometimes been reported. Cement creep did, however, relax cement stresses and create a more favourable stress distribution at the interfaces. These trends were found around both the bonded and unbonded stems. Our results did not confirm that polished, unbonded, stems accommodated creep better than bonded stems in terms of cement and interface stress patterns.