We compared the accuracy, operating time and radiation exposure
of the introduction of iliosacral screws using O-arm/Stealth Navigation
and standard fluoroscopy. Iliosacral screws were introduced percutaneously into the first
sacral body (S1) of ten human cadavers, four men and six women.
The mean age was 77 years (58 to 85). Screws were introduced using
a standard technique into the left side of S1 using C-Arm fluoroscopy
and then into the right side using O-Arm/Stealth Navigation. The
radiation was measured on the surgeon by dosimeters placed under
a lead thyroid shield and apron, on a finger, a hat and on the cadavers.Aims
Materials and Methods
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article: