Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
The clinical results of bilateral total knee replacement staged at a one-week interval during a single hospital admission were compared with bilateral total knee replacements performed under the same anaesthetic and with bilateral total knee replacements performed during two separate admissions. The data were retrospectively reviewed. All operations had been performed by the same surgeon using the same design of prosthesis at a single institution. The operative time and length of stay for the one-week staged group were comparable with those of the separate admission group but longer than for the patients treated under one anaesthetic. There was a low rate of complications and good clinical outcome in all groups at a mean follow-up of four years (1 to 7.2). The group staged at a one-week interval had the least blood loss (p = 0.004). With appropriate patient selection, bilateral total knee replacement performed under a single anaesthetic, or staged at a one-week interval, is a safe and effective method to treat bilateral arthritis of the knee.