We reviewed the long-term clinical and radiological
results of 63 uncemented Low Contact Stress (LCS) total knee replacements
(TKRs) in 47 patients with rheumatoid arthritis. The mean age of
the patients at the time of surgery was 69 years (53 to 81). At
a mean follow-up of 22 years (20 to 25), 12 patients were alive
(17 TKRs), 27 had died (36 TKRs), and eight (ten TKRs) were lost
to follow-up. Revision was necessary in seven patients (seven TKRs, 11.1%)
at a mean of 12.1 years (0 to 19) after surgery. In the surviving
ten patients who had not undergone revision (15 TKRs), the mean
Oxford knee score was 30.2 (16 to 41) at a mean follow-up of 19.5
years (15 to 24.7) and mean active flexion was 105° (90° to 150°).
The survival rate was 88.9% at 20 years (56 of 63) and the Kaplan–Meier
survival estimate, without revision, was 80.2% (95% confidence interval
37 to 100) at 25 years. Cite this article:
We analysed the long-term clinical and radiological results of 63 uncemented Low Contact Stress total knee replacements in 47 patients with rheumatoid arthritis. At a mean follow-up of 12.9 years (10 to 16), 36 patients (49 knees) were still alive; three patients (five knees) were lost to follow-up. Revision was necessary in three knees (4.8%) and the rate of infection was 3.2%. The mean clinical and functional Knee Society scores were 90 (30 to 98) and 59 (25 to 90), respectively, at final follow-up and the mean active range of movement was 104° (55° to 120°). The survival rate was 94% at 16 years but 85.5% of patients lost to follow-up were considered as failures. Radiological evidence of impending failure was noted in one knee.