In an interdisciplinary project involving electronic
engineers and clinicians, a telemetric system was developed to measure
the bending load in a titanium internal femoral fixator. As this
was a new device, the main question posed was: what clinically relevant
information could be drawn from its application? As a first clinical
investigation, 27 patients (24 men, three women) with a mean age
of 38.4 years (19 to 66) with femoral nonunions were treated using the
system. The mean duration of the nonunion was 15.4 months (5 to
69). The elasticity of the plate-callus system was measured telemetrically
until union. Conventional radiographs and a CT scan at 12 weeks
were performed routinely, and healing was staged according to the
CT scans. All nonunions healed at a mean of 21.5 weeks (13 to 37).
Well before any radiological signs of healing could be detected,
a substantial decrease in elasticity was recorded. The relative
elasticity decreased to 50% at a mean of 7.8 weeks (3.5 to 13) and
to 10% at a mean of 19.3 weeks (4.5 to 37). At 12 weeks the mean
relative elasticity was 28.1% (0% to 56%). The relative elasticity
was significantly different between the different healing stages
as determined by the CT scans. Incorporating load measuring electronics into implants is a promising
option for the assessment of bone healing. Future application might
lead to a reduction in the need for exposure to ionising radiation
to monitor fracture healing.
The precise localisation of osteoarthritic changes is crucial for selective surgical treatment. Single photon-emission CT-CT (SPECT-CT) combines both morphological and biological information. We hypothesised that SPECT-CT increased the intra- and interobserver reliability to localise increased uptake compared with traditional evaluation of CT and bone scanning together. We evaluated 20 consecutive patients with pain of uncertain origin in the foot and ankle by radiography and SPECT-CT, available as fused SPECT-CT, and by separate bone scanning and CT. Five observers assessed the presence or absence of arthritis. The images were blinded and randomly ordered. They were evaluated twice at an interval of six weeks. Kappa and multirater kappa values were calculated. The mean intraobserver reliability for SPECT-CT was excellent (κ = 0.86; 95% CI 0.81 to 0.88) and significantly higher than that for CT and bone scanning together. SPECT-CT had significantly higher interobserver agreement, especially when evaluating the naviculocuneiform and tarsometatarsal joints. SPECT-CT is useful in localising active arthritis especially in areas where the number and configuration of joints are complex.