header advert
Results 1 - 10 of 10
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 185 - 191
1 Feb 2015
Kendrick BJL Kaptein BL Valstar ER Gill HS Jackson WFM Dodd CAF Price AJ Murray DW

The most common reasons for revision of unicompartmental knee arthroplasty (UKA) are loosening and pain. Cementless components may reduce the revision rate. The aim of this study was to compare the fixation and clinical outcome of cementless and cemented Oxford UKAs.

A total of 43 patients were randomised to receive either a cemented or a cementless Oxford UKA and were followed for two years with radiostereometric analysis (RSA), radiographs aligned with the bone–implant interfaces and clinical scores.

The femoral components migrated significantly during the first year (mean 0.2 mm) but not during the second. There was no significant difference in the extent of migration between cemented and cementless femoral components in either the first or the second year. In the first year the cementless tibial components subsided significantly more than the cemented components (mean 0.28 mm (sd 0.17) vs. 0.09 mm (sd 0.19 mm)). In the second year, although there was a small amount of subsidence (mean 0.05 mm) there was no significant difference (p = 0.92) between cemented and cementless tibial components. There were no femoral radiolucencies. Tibial radiolucencies were narrow (< 1 mm) and were significantly (p = 0.02) less common with cementless (6 of 21) than cemented (13 of 21) components at two years. There were no complete radiolucencies with cementless components, whereas five of 21 (24%) cemented components had complete radiolucencies. The clinical scores at two years were not significantly different (p = 0.20).

As second-year migration is predictive of subsequent loosening, and as radiolucency is suggestive of reduced implant–bone contact, these data suggest that fixation of the cementless components is at least as good as, if not better than, that of cemented devices.

Cite this article: Bone Joint J 2015; 97-B:185–91.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1216 - 1220
1 Sep 2012
Weston-Simons JS Pandit H Jenkins C Jackson WFM Price AJ Gill HS Dodd CAF Murray DW

The Oxford unicompartmental knee replacement (UKR) is an established treatment option in the management of symptomatic end-stage medial compartmental osteoarthritis (MCOA), which works well in the young and active patient. However, previous studies have shown that it is reliable only in the presence of a functionally intact anterior cruciate ligament (ACL). This review reports the outcomes, at a mean of five years and a maximum of ten years, of 52 consecutive patients with a mean age of 51 years (36 to 57) who underwent staged or simultaneous ACL reconstruction and Oxford UKR. At the last follow-up (with one patient lost to follow-up), the mean Oxford knee score was 41 (sd 6.3; 17 to 48). Two patients required conversion to TKR: one for progression of lateral compartment osteoarthritis and one for infection. Implant survival at five years was 93% (95% CI 83 to 100). All but one patient reported being satisfied with the procedure. The outcome was not significantly influenced by age, gender, femoral or tibial tunnel placement, or whether the procedure was undertaken at one- or two-stages.

In summary, ACL reconstruction and Oxford UKR gives good results in patients with end-stage MCOA secondary to ACL deficiency.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 622 - 628
1 May 2011
Pandit H Jenkins C Gill HS Smith G Price AJ Dodd CAF Murray DW

The contraindications for unicompartmental knee replacement (UKR) remain controversial. The views of many surgeons are based on Kozinn and Scott’s 1989 publication which stated that patients who weighed more than 82 kg, were younger than 60 years, undertook heavy labour, had exposed bone in the patellofemoral joint or chondrocalcinosis, were not ideal candidates for UKR. Our aim was to determine whether these potential contraindications should apply to patients with a mobile-bearing UKR. In order to do this the outcome of patients with these potential contraindications was compared with that of patients without the contraindications in a prospective series of 1000 UKRs. The outcome was assessed using the Oxford knee score, the American Knee Society score, the Tegner activity score, revision rate and survival.

The clinical outcome of patients with each of the potential contraindications was similar to or better than those without each contraindication. Overall, 678 UKRs (68%) were performed in patients who had at least one potential contraindication and only 322 (32%) in patients deemed to be ideal. The survival at ten years was 97.0% (95% confidence interval 93.4 to 100.0) for those with potential contraindications and 93.6% (95% confidence interval 87.2 to 100.0) in the ideal patients.

We conclude that the thresholds proposed by Kozinn and Scott using weight, age, activity, the state of the patellofemoral joint and chondrocalcinosis should not be considered to be contraindications for the use of the Oxford UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 470 - 475
1 Apr 2011
Kendrick BJL Simpson DJ Kaptein BL Valstar ER Gill HS Murray DW Price AJ

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively.

The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year).

The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 198 - 204
1 Feb 2011
Pandit H Jenkins C Gill HS Barker K Dodd CAF Murray DW

This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 (sd 7.2), the mean American Knee Society Objective Score 86.4 (sd 13.4), mean American Knee Society Functional Score 86.1 (sd 16.6), mean Tegner activity score 2.8 (sd 1.1). For the entire cohort, the mean maximum flexion was 130° at the time of final review.

The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100).

This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 367 - 373
1 Mar 2010
Kendrick BJL Longino D Pandit H Svard U Gill HS Dodd CAF Murray DW Price AJ

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/− signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001).

These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 374 - 379
1 Mar 2010
Kendrick BJL Rout R Bottomley NJ Pandit H Gill HS Price AJ Dodd CAF Murray DW

With medial unicompartmental osteoarthritis (OA) there is occasionally a full-thickness ulcer of the cartilage on the medial side of the lateral femoral condyle. It is not clear whether this should be considered a contraindication to unicompartmental knee replacement (UKR). The aim of this study was to determine why these ulcers occur, and whether they compromise the outcome of UKR.

Case studies of knees with medial OA suggest that cartilage lesions on the medial side of the lateral condyle are caused by impingement on the lateral tibial spine as a result of the varus deformity and tibial subluxation. Following UKR the varus and the subluxation are corrected, so that impingement is prevented and the damaged part of the lateral femoral condyle is not transmitting load. An illustrative case report is presented.

Out of 769 knees with OA of the medial compartment treated with the Oxford UKR, 59 (7.7%) had partial-thickness cartilage loss and 20 (2.6%) had a full-thickness cartilage deficit on the medial side of the lateral condyle. The mean Oxford Knee Score (OKS) at the last follow-up at a mean of four years was 41.9 (13 to 48) in those with partial-thickness cartilage loss and 41.0 (20 to 48) in those with full-thickness loss. In those with normal or superficially damaged cartilage the mean was 39.5 (5 to 48) and 39.7 (8 to 48), respectively. There were no statistically significant differences between the pre-operative OKS, the final review OKS or of change in the score in the various groups.

We conclude that in medial compartment OA, damage to the medial side of the lateral femoral condyle is caused by impingement on the tibial spine and should not be considered a contraindication to an Oxford UKR, even if there is extensive full-thickness ulceration of the cartilage.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1597 - 1601
1 Dec 2007
Beard DJ Pandit H Gill HS Hollinghurst D Dodd CAF Murray DW

Patellofemoral joint degeneration is often considered a contraindication to medial unicompartmental knee replacement. We examined the validity of this preconception using information gathered prospectively on the intra-operative status of the patellofemoral joint in 824 knees in 793 consecutive patients who underwent Oxford unicondylar knee replacement for anteromedial osteoarthritis. All operations were performed between January 1998 and September 2005. A five-point grading system classified degeneration of the patellofemoral joint from none to full-thickness cartilage loss. A subclassification of the presence or absence of any full-thickness cartilage loss was subsequently performed to test selected hypotheses. Outcome was evaluated independently by physiotherapists using the Oxford and the American Knee Society Scores with a minimum follow-up of one year.

Full-thickness cartilage loss on the trochlear surface was observed in 100 of 785 knees (13%), on the medial facet of the patella in 69 of 782 knees (9%) and on the lateral facet in 29 of 784 knees (4%). Full-thickness cartilage loss at any location was seen in 128 knees (16%) and did not produce a significantly worse outcome than those with a normal or near-normal joint surface. The severity of the degeneration at any of the intra-articular locations also had no influence on outcome.

We concluded that, provided there is not bone loss and grooving of the lateral facet, damage to the articular cartilage of the patellofemoral joint to the extent of full-thickness cartilage loss is not a contraindication to the Oxford mobile-bearing unicompartmental knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 940 - 945
1 Jul 2005
Pandit H Ward T Hollinghurst D Beard DJ Gill HS Thomas NP Murray DW

Abnormal sagittal kinematics after total knee replacement (TKR) can adversely affect functional outcome. Two important determinants of knee kinematics are component geometry and the presence or absence of a posterior-stabilising mechanism (cam-post). We investigated the influence of these variables by comparing the kinematics of a TKR with a polyradial femur with a single radius design, both with and without a cam-post mechanism.

We assessed 55 patients, subdivided into four groups, who had undergone a TKR one year earlier by using an established fluoroscopy protocol in order to examine their kinematics in vivo. The kinematic profile was obtained by measuring the patellar tendon angle through the functional knee flexion range (0° to 90°) and the results compared with 14 normal knees. All designs of TKR had abnormal sagittal kinematics compared with the normal knee. There was a significant (p < 0.05) difference between those of the two TKRs near to full extension. The presence of the cam-post mechanism did not influence the kinematics for either TKR design. These differences suggest that surface geometry is a stronger determinant of kinematics than the presence or absence of a cam-post mechanism for these two designs. This may be because the cam-post mechanism is ineffective.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 8 | Pages 1098 - 1103
1 Nov 2001
Beard DJ Murray DW Gill HS Price AJ Rees JL Alfaro-Adrián J Dodd CAF

We have assessed the effectiveness of reconstruction of the anterior cruciate ligament (ACL) in reducing functional tibial translation (TT).

The gait of 11 ACL-deficient patients was studied using Vicon equipment before and after surgery. Measurements of the angle between the patellar tendon and the long axis of the tibia were obtained in order to calculate TT in the sagittal plane relative to the uninjured limb during standing and walking.

Before surgery, patients did not show abnormal TT on the injured side, but after surgery significant anterior TT was found in the operated limb for every parameter of gait.

Abnormal anterior TT occurring during activity does not seem to be reduced by reconstruction; rather, it increases. It may be that the increased translation results from relaxation of excess contraction of the hamstring muscles, since compensatory muscle activity no longer is required in a reconstructed knee. The reduction of TT may not be an appropriate objective in surgery on the ACL.