header advert
Results 21 - 40 of 48
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 198 - 204
1 Feb 2011
Pandit H Jenkins C Gill HS Barker K Dodd CAF Murray DW

This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 (sd 7.2), the mean American Knee Society Objective Score 86.4 (sd 13.4), mean American Knee Society Functional Score 86.1 (sd 16.6), mean Tegner activity score 2.8 (sd 1.1). For the entire cohort, the mean maximum flexion was 130° at the time of final review.

The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100).

This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1527 - 1534
1 Nov 2010
Grammatopoulos G Pandit H Murray DW Gill HS

Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients).

Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 (sd 0.10) vs mean 1.30 (sd 0.08) p = 0.001). At operation the patients with pseudotumours had a greater reduction in the size of their femoral heads (p = 0.035) and subsequently had greater neck narrowing (mean 10.1% (sd 7.2) vs mean 3.8% (sd 3.2) p < 0.001). No female patient with a pre-operative HNR ≤ 1.3 developed a pseudotumour.

We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1356 - 1362
1 Oct 2010
Simpson DJ Kendrick BJL Hughes M Glyn-Jones S Gill HS Rushforth GF Murray DW

We have evaluated the difference in the migration patterns over two years of two cementless stems in a randomised, controlled trial using radiostereophotogrammetric analysis (RSA). The implants studied were the Furlong HAC stem, which has good long-term results and the Furlong Active stem, which is a modified version of the former designed to minimise stress concentrations between the implant and bone, and thus to improve fixation.

A total of 23 Furlong HAC and 20 Furlong Active stems were implanted in 43 patients. RSA examinations were carried out immediately post-operatively and at six, 12 and 24 months post-operatively.

The subsidence during the first year in the Furlong HAC stem, was approximately one-third that of the Furlong Active stem, the measured mean subsidence of the femoral head at six months being 0.27 mm (95% confidence interval (CI) 0.03 to 0.51) and 0.99 mm (95% CI 0.38 to 1.60), respectively (p = 0.03). One Active stem continued to subside during the second year. All hips, regardless of the type of stem were clinically successful as judged by the Oxford hip score and a derived pain score without any distinction between the two types of stem.

The initial stability of the Furlong Active stem was not as good as the established stem which might compromise osseo-integration to the detriment of long-term success. The changes in the geometry of the stem, to minimise stress have affected the attainment of initial stability.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1072 - 1078
1 Aug 2010
Grammatopoulos G Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gill HS Murray DW

Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation.

We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone.

In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 787 - 793
1 Jun 2010
Steffen RT Athanasou NA Gill HS Murray DW

The cause of fracture of the femoral neck after hip resurfacing is poorly understood. In order to evaluate the role of avascular necrosis we compared 19 femoral heads retrieved at revision for fracture of the femoral neck and 13 retrieved for other reasons.

We developed a new technique of assessing avascular necrosis in the femoral head by determining the percentage of empty osteocyte lacunae present. Femoral heads retrieved as controls at total hip replacement for osteoarthritis and avascular necrosis had 9% (sd 4; n = 13) and 85% (sd 5; n = 10, p < 0.001) empty lacunae, respectively.

In the fracture group the percentage of empty lacunae was 71% (sd 22); in the other group it was 21% (sd 13). The differences between the groups were highly significant (p < 0.001).

We conclude that fracture after resurfacing of the hip is associated with a significantly greater percentage of empty osteocyte lacunae within the trabecular bone. This indicates established avascular necrosis and suggests that damage to the blood supply at the time of surgery is a potent risk factor for fracture of the femoral neck after hip resurfacing.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 356 - 361
1 Mar 2010
Kwon Y- Glyn-Jones S Simpson DJ Kamali A McLardy-Smith P Gill HS Murray DW

The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure.

We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred.

Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 335 - 336
1 Mar 2010
De Smet K Campbell PA Gill HS

We report the consensus of surgical opinions of an international faculty of expert metal-on-metal hip resurfacing surgeons, with a combined experience of over 18 000 cases, covering required experience, indications, surgical technique, rehabilitation and the management of problematic cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 367 - 373
1 Mar 2010
Kendrick BJL Longino D Pandit H Svard U Gill HS Dodd CAF Murray DW Price AJ

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/− signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001).

These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 374 - 379
1 Mar 2010
Kendrick BJL Rout R Bottomley NJ Pandit H Gill HS Price AJ Dodd CAF Murray DW

With medial unicompartmental osteoarthritis (OA) there is occasionally a full-thickness ulcer of the cartilage on the medial side of the lateral femoral condyle. It is not clear whether this should be considered a contraindication to unicompartmental knee replacement (UKR). The aim of this study was to determine why these ulcers occur, and whether they compromise the outcome of UKR.

Case studies of knees with medial OA suggest that cartilage lesions on the medial side of the lateral condyle are caused by impingement on the lateral tibial spine as a result of the varus deformity and tibial subluxation. Following UKR the varus and the subluxation are corrected, so that impingement is prevented and the damaged part of the lateral femoral condyle is not transmitting load. An illustrative case report is presented.

Out of 769 knees with OA of the medial compartment treated with the Oxford UKR, 59 (7.7%) had partial-thickness cartilage loss and 20 (2.6%) had a full-thickness cartilage deficit on the medial side of the lateral condyle. The mean Oxford Knee Score (OKS) at the last follow-up at a mean of four years was 41.9 (13 to 48) in those with partial-thickness cartilage loss and 41.0 (20 to 48) in those with full-thickness loss. In those with normal or superficially damaged cartilage the mean was 39.5 (5 to 48) and 39.7 (8 to 48), respectively. There were no statistically significant differences between the pre-operative OKS, the final review OKS or of change in the score in the various groups.

We conclude that in medial compartment OA, damage to the medial side of the lateral femoral condyle is caused by impingement on the tibial spine and should not be considered a contraindication to an Oxford UKR, even if there is extensive full-thickness ulceration of the cartilage.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1566 - 1574
1 Dec 2009
Glyn-Jones S Pandit H Kwon Y Doll H Gill HS Murray DW

Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery.

Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p < 0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p < 0.001).

We recommend that resurfacings are undertaken with caution in women, particularly those under 40 years of age but they remain a good option in young men. Further work is required to understand the aetiology of pseudotumours so that this complication can be avoided.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1019 - 1024
1 Aug 2009
Grammatopolous G Pandit H Kwon Y Gundle R McLardy-Smith P Beard DJ Murray DW Gill HS

Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements.

We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant.

At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 (sd 9.3) and was significantly worse (p < 0.001) than the outcome for fracture with a mean Oxford hip score of 40.2 (sd 9.2) or that for other causes with a mean Oxford hip score of 37.8 (sd 9.4). The clinical outcome of revision for pseudotumour was also significantly worse (p < 0.001) than the outcome of matched primary total hip replacements. By contrast, the outcome for fracture and other causes was not significantly different from that of matched primary total hip replacements (p = 0.065). After revision for pseudotumour there were three cases of recurrent dislocation, three of palsy of the femoral nerve, one of stenosis of the femoral artery and two of loosening of the component. Five hips required further revision. In three of these there was evidence of recurrent pseudotumour, and one is currently awaiting further revision. The incidence of major complications after revision for pseudotumour (50%) was significantly higher (p = 0.018) than that after revision for other causes (14%).

The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1291 - 1297
1 Oct 2008
De Haan R Pattyn C Gill HS Murray DW Campbell PA De Smet K

We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55°.

There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p < 0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55°, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 847 - 851
1 Jul 2008
Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gibbons CLM Ostlere S Athanasou N Gill HS Murray DW

We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken.

All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision.

We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 556 - 561
1 May 2008
Glyn-Jones S McLardy-Smith P Gill HS Murray DW

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 436 - 441
1 Apr 2008
Steffen RT Pandit HP Palan J Beard DJ Gundle R McLardy-Smith P Murray DW Gill HS

Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series.

A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points (sd 7.7) and the mean University of California Los Angeles activity score was 6.6 points (sd 1.9). There were no patients with definite radiological evidence of loosening or of narrowing of the femoral neck exceeding 10% of its width. There were 23 revisions (3.8%), giving an overall survival of 95% (95% confidence interval 85.3 to 99.2) at seven years. Fractured neck of femur in 12 hips was the most common indication for revision, followed by aseptic loosening in four. In three hips (three patients) (0.5%), failure was possibly related to metal debris.

Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 494 - 499
1 Apr 2008
Howells NR Gill HS Carr AJ Price AJ Rees JL

The aim of this study was to investigate the effect of laboratory-based simulator training on the ability of surgical trainees to perform diagnostic arthroscopy of the knee.

A total of 20 junior orthopaedic trainees were randomised to receive either a fixed protocol of arthroscopic simulator training on a bench-top knee simulator or no additional training. Motion analysis was used to assess performance objectively. Each trainee then received traditional instruction and demonstrations of diagnostic arthroscopy of the knee in theatre before performing the procedure under the supervision of a blinded consultant trainer. Their performance was assessed using a procedure-based assessment from the Orthopaedic Competence Assessment Project and a five-point global rating assessment scale.

In theatre the simulator-trained group performed significantly better than the untrained group using the Orthopaedic Competence Assessment Project score (p = 0.0007) and assessment by the global rating scale (p = 0.0011), demonstrating the transfer of psychomotor skills from simulator training to arthroscopy in the operating theatre. This has implications for the planning of future training curricula.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1597 - 1601
1 Dec 2007
Beard DJ Pandit H Gill HS Hollinghurst D Dodd CAF Murray DW

Patellofemoral joint degeneration is often considered a contraindication to medial unicompartmental knee replacement. We examined the validity of this preconception using information gathered prospectively on the intra-operative status of the patellofemoral joint in 824 knees in 793 consecutive patients who underwent Oxford unicondylar knee replacement for anteromedial osteoarthritis. All operations were performed between January 1998 and September 2005. A five-point grading system classified degeneration of the patellofemoral joint from none to full-thickness cartilage loss. A subclassification of the presence or absence of any full-thickness cartilage loss was subsequently performed to test selected hypotheses. Outcome was evaluated independently by physiotherapists using the Oxford and the American Knee Society Scores with a minimum follow-up of one year.

Full-thickness cartilage loss on the trochlear surface was observed in 100 of 785 knees (13%), on the medial facet of the patella in 69 of 782 knees (9%) and on the lateral facet in 29 of 784 knees (4%). Full-thickness cartilage loss at any location was seen in 128 knees (16%) and did not produce a significantly worse outcome than those with a normal or near-normal joint surface. The severity of the degeneration at any of the intra-articular locations also had no influence on outcome.

We concluded that, provided there is not bone loss and grooving of the lateral facet, damage to the articular cartilage of the patellofemoral joint to the extent of full-thickness cartilage loss is not a contraindication to the Oxford mobile-bearing unicompartmental knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1293 - 1298
1 Oct 2007
Steffen R O’Rourke K Gill HS Murray DW

In 12 patients, we measured the oxygen concentration in the femoral head-neck junction during hip resurfacing through the anterolateral approach. This was compared with previous measurements made for the posterior approach. For the anterolateral approach, the oxygen concentration was found to be highly dependent upon the position of the leg, which was adjusted during surgery to provide exposure to the acetabulum and femoral head. Gross external rotation of the hip gave a significant decrease in oxygenation of the femoral head. Straightening the limb led to recovery in oxygen concentration, indicating that the blood supply was maintained. The oxygen concentration at the end of the procedure was not significantly different from that at the start.

The anterolateral approach appears to produce less disruption to the blood flow in the femoral head-neck junction than the posterior approach for patients undergoing hip resurfacing. This may be reflected subsequently in a lower incidence of fracture of the femoral neck and avascular necrosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 16 - 20
1 Jan 2007
Gill HS Campbell PA Murray DW De Smet KA

Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty.

The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014).

We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1677 - 1677
1 Dec 2006
MURRAY DW GILL HS