The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation. The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.
We studied 185 total hip replacements and related the identification of radiolucent lines (RLLs) at two years to the later development of lytic lesions and loosening. Linear polyethylene wear was also measured. RLLs appeared in 34 hips at a mean of 2.0 years after operation, and lytic lesions in ten hips at 5.7 years. Of 151 THRs without RLLs there was neither rapid migration nor loosening and only one developed a possible lytic lesion. Of 23 hips with non-progressive RLLs there was neither rapid migration nor loosening, but six developed a lytic lesion. By contrast, 11 THRs with progressive RLLs migrated rapidly and seven developed a lytic lesion. Six THRs with progressive RLLs failed. The wear rates were the same in all groups, although limited numbers were available for study. If the surgeon achieves secure initial fixation as shown by slow or no migration and no RLLs during the first two years, it is likely that no lytic lesions will develop by five years or aseptic loosening by ten years. If an imperfect, but adequate, interface is achieved, as shown by slow migration and non-progressive RLLs lytic lesions adjacent to the RLLs may develop by five years, but aseptic loosening will be unlikely at ten. Insecure initial fixation, as shown by more rapid migration and progressive RLLs at two years, is likely to lead to the formation of lytic lesions at five years and loosening at ten. The outcome after THR is therefore determined at the initial operation and may be predicted at two years. The presence of lytic lesions reflects soft tissue at the interface as shown by the RLLs which accompany and promote loosening but, in our study, did not cause it.
We reviewed a consecutive series of 527 uninfected hip replacements in patients resident in the UK which had been implanted from 1981 to 1993. All had the same basic design of femoral prosthesis, but four fixation techniques had been used: two press-fit, one HA-coated and one cemented. Review and radiography were planned prospectively. For assessment the components were retrospectively placed into two groups: those which had failed from two years onwards by aseptic femoral loosening and those in which the femoral component had survived without revision or recommendation for revision. All available radiographs in both groups were measured to determine vertical migration and examined by two observers to agree the presence of radiolucent lines (RLLs), lytic lesions, resorption of the neck, proximal osteopenia and distal intramedullary and distal subperiosteal formation of new bone. We then related the presence or absence of these features and the rate of migration at two years to the outcome with regard to aseptic loosening and determined the predictive value of each of these variables. Migration of ≥2 mm at two years, the presence of an RLL of 2 mm occupying one-third of any one zone, and subperiosteal formation of new bone at the tip of the stem were predictors of aseptic loosening after two years. There were too few lytic lesions to assess at two years, but at five years a lytic lesion ≥2 mm also predicted failure. We discuss the use of these variables as predictors of femoral aseptic loosening for groups of hips and for individual hips. We conclude that if a group of about 50 total hip replacements, perhaps with a new design of femoral stem, were studied in this way at two years, a mean migration of <
0.4 mm and an incidence of <
10% of RLLs of 2 mm in any one zone would predict 95% survival at ten years. For an individual prosthesis, migration of <
2 mm and the absence of an RLL of ≤2 mm at two years predict a 6% chance of revision over approximately ten years. If either 2 mm of migration Our findings suggest that replacements using a limited number of any new design of femoral prosthesis should be screened radiologically at two years before they are generally introduced. We also suggest that radiographs of individual patients at two years and perhaps at five years should be studied to help to decide whether or not the patient should remain under close review or be discharged from specialist follow-up.
We report the radiological and clinical outcome of a press-fit (SLF) acetabular component at two to three years in two groups of patients having primary total hip replacement. In 69 the implant was coated with hydroxyapatite (HA) and in 40 it was uncoated. The stability of the cup was assessed by measurement of proximal migration and change in the angle of inclination. The clinical results in the two groups did not differ significantly, and the mean proximal linear wear was similar in both. Fewer radiolucent lines (RLLs) were seen on the radiographs of cups coated with HA. The mean proximal migration was studied by calculating regression lines for each patient using migration measurements: for the SLF+HA group the mean slope was 0.06 mm/year and for the SLF–HA group 0.20 mm/year (p = 0.22). The change in the angle of inclination during follow-up was also consistently smaller in HA–coated cups. Using regression methods the SLF+HA group had a mean slope of 0.08°/year and the SLF-HA group 0.44°/year (p = 0.023). Partial HA coating appeared to have no effect on the clinical outcome or on the rate of wear of polyethylene, but there was a trend towards a reduced rate of proximal migration, and a significant reduction in rotational migration and the number of radiolucent lines. This suggests that HA coating enhances the stability of this acetabular component.