This study investigates the use of porous biphasic ceramics as graft extenders in impaction grafting of the femur during revision hip surgery. Impaction grafting of the femur was performed in four groups of sheep. Group one received pure allograft, group two 50% allograft and 50% BoneSave, group three 50% allograft and 50% BoneSave type 2 and group four 10% allograft and 90% BoneSave as the graft material. Function was assessed using an index of pre- and post-operative peak vertical ground reaction force ratios. Changes in bone mineral density were measured by dual energy X ray absorptiometry (DEXA) scanning. Loosening and subsidence were assessed radiographically and by histological examination of the explanted specimens. There was no statistically significant difference between the four groups after 18 months of unrestricted functional loading for all outcome measures.
Axial forces were measured during limb lengthening in a series of ten patients with varying pathologies in order to assess the mechanical characteristics of the distracted tissues and the levels of axial force to which soft tissues are subjected during leg lengthening. The pattern of force was found to vary according to the underlying pathology. For post-traumatic shortening in adults both the peak and the resting forces rose steadily during lengthening reaching maximum forces of the order of 300 N. Patients with congenitally short limbs developed very high peak forces (in some cases over 1000 N) and also showed large amounts of force relaxation (typically 400 to 500 N). When very high levels of force were recorded, there was a higher complication rate. In particular, there was a high instance of angular deformity. This occurred because the loads encountered resulted in failure of some of the external fixation frames.