Conventional cemented acetabular components are
reported to have a high rate of failure when implanted into previously
irradiated bone. We recommend the use of a cemented reconstruction
with the addition of an acetabular reinforcement cross to improve
fixation. We reviewed a cohort of 45 patients (49 hips) who had undergone
irradiation of the pelvis and a cemented total hip arthroplasty
(THA) with an acetabular reinforcement cross. All hips had received
a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour
or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3:
45 to 60; mean: 49.57, 32 to 72). The mean follow-up after THA was 51 months (17 to 137). The cumulative
probability of revision of the acetabular component for a mechanical
reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60
months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One
hip was revised for mechanical failure and three for infection. Cemented acetabular components with a reinforcement cross provide
good medium-term fixation after pelvic irradiation. These patients
are at a higher risk of developing infection of their THA. Cite this article:
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.