Aims. Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or
We performed a prospective study of 54 patients (76 knees) who underwent Osteonics series 3000 cruciate-retaining cementless total knee arthroplasty between December 1990 and June 1993. Five patients (seven knees) were lost to follow-up (90.7% completion). One patient required revision at 10.5 years after operation. The rate of survival was 100% at ten years and 96.7% at 13 years. The mean ten-year knee and function scores were 79 and 59 respectively. Both were significantly better than the pre-operative scores. The range of movement also improved. Although a radiolucent line around the tibial component enlarged in six knees (20.7%) at ten years, the clinical outcome was generally good. In a patient who died after 5.5 years, post-mortem examination of the knee showed no bony ingrowth into the tibial component. Despite poor bony ingrowth,
The use of platelet-rich plasma (PRP) as an adjuvant
to tissue repair is gaining favour in orthopaedic surgery. Tunnel widening
after anterior cruciate ligament (ACL) reconstruction is a recognised
phenomenon that could compromise revision surgery. The purpose of
this study was to determine whether PRP might prevent tunnel widening
in ACL reconstruction. Patients undergoing ACL reconstruction using a hamstring graft
were randomly allocated either to have PRP introduced into the tunnels
peri-operatively or not. CT scanning of the knees was carried out
on the day after surgery and at three months post-operatively and
the width of the tunnels was measured. Patients were also evaluated
clinically at three months, when laxity was also measured. Each group comprised 25 patients, and at three months post-operatively
all were pain-free with stable knees, a negative Lachman test and
a good range of movement. Arthrometric results had improved significantly
in both groups (p <
0.001). Despite slightly less tunnel widening
in the PRP group, there was no significant difference between the
groups at the femoral opening or the mid-tunnel (p = 0.370 and p
= 0.363, respectively) nor at the tibial opening or mid-tunnel (p
= 0.333 and p = 0.177, respectively). We conclude that PRP has no significant effect in preventing
tunnel widening after ACL reconstruction. Cite this article:
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article:
This review considers the surgical treatment
of displaced fractures involving the knee in elderly, osteoporotic patients.
The goals of treatment include pain control, early mobilisation,
avoidance of complications and minimising the need for further surgery.
Open reduction and internal fixation (ORIF) frequently results in
loss of reduction, which can result in post-traumatic arthritis
and the occasional conversion to total knee replacement (TKR). TKR
after failed internal fixation is challenging, with modest functional
outcomes and high complication rates. TKR undertaken as treatment
of the initial fracture has better results to late TKR, but does
not match the outcome of primary TKR without complications. Given
the relatively infrequent need for late TKR following failed fixation,
ORIF is the preferred management for most cases. Early TKR can be
considered for those patients with pre-existing arthritis, bicondylar
femoral fractures, those who would be unable to comply with weight-bearing restrictions,
or where a single definitive procedure is required.
We compared peri-prosthetic bone mineral density between identical cemented and cementless LCS rotating platform total knee arthroplasties. Two matched cohorts had dual energy x-ray absorptiometry scans two years post-operatively using a modified validated densitometric analysis protocol, to assess peri-prosthetic bone mineral density. The knee that was not operated on was also scanned to enable the calculation of a relative bone mineral density difference. Oxford Knee and American Knee Society scores were comparable in the two cohorts. Statistical analysis revealed no significant difference in absolute, or relative peri-prosthetic bone mineral density with respect to the method of fixation. However, the femoral peri-prosthetic bone mineral density and relative bone mineral density difference were significantly decreased, irrespective of the method of fixation, particularly in the anterior distal portion of the femur, with a mean reduction in relative bone mineral density difference of 27%. There was no difference in clinical outcome between the cemented and cementless LCS total knee arthroplasty. However, both produce stress-shielding around the femoral implants. This leads us to question the use of more expensive cementless total knee components.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.