Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.
Between 1999 and 2005, 23 failed total ankle replacements were converted to arthrodeses. Three surgical techniques were used: tibiotalar arthrodesis with screw fixation, tibiotalocalcaneal arthrodesis with screw fixation, and tibiotalocalcaneal arthrodesis with an intramedullary nail. As experience was gained, the benefits and problems became apparent. Successful bony union was seen in 17 of the 23 ankles. The complication rate was higher in ankles where the loosening had caused extensive destruction of the body of the talus, usually in rheumatoid arthritis. In this situation we recommend tibiotalocalcaneal arthrodesis with an intramedullary nail. This technique can also be used when there is severe arthritic change in the subtalar joint. Arthrodesis of the tibiotalar joint alone using compression screws was generally possible in osteoarthritis because the destruction of the body of the talus was less extensive. Tibiotalocalcaneal arthrodesis fusion with compression screws has not been successful in our experience.