Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 252 - 257
1 Feb 2015
Wafa H Grimer RJ Reddy K Jeys L Abudu A Carter SR Tillman RM

We conducted a case-control study to examine the merit of silver-coated tumour prostheses. We reviewed 85 patients with Agluna-treated (silver-coated) tumour implants treated between 2006 and 2011 and matched them with 85 control patients treated between 2001 and 2011 with identical, but uncoated, tumour prostheses.

In all, 106 men and 64 women with a mean age of 42.2 years (18.4 to 90.4) were included in the study. There were 50 primary reconstructions (29.4%); 79 one-stage revisions (46.5%) and 41 two-stage revisions for infection (24.1%).

The overall post-operative infection rate of the silver-coated group was 11.8% compared with 22.4% for the control group (p = 0.033, chi-square test). A total of seven of the ten infected prostheses in the silver-coated group were treated successfully with debridement, antibiotics, and implant retention compared with only six of the 19 patients (31.6%) in the control group (p = 0.048, chi-square test). Three patients in the silver-coated group (3.5%) and 13 controls (15.3%) had chronic periprosthetic infection (p = 0.009, chi-square test).

The overall success rates in controlling infection by two-stage revision in the silver-coated group was 85% (17/20) compared with 57.1% (12/21) in the control group (p = 0.05, chi-square test). The Agluna-treated endoprostheses were associated with a lower rate of early periprosthetic infection. These silver-treated implants were particularly useful in two-stage revisions for infection and in those patients with incidental positive cultures at the time of implantation of the prosthesis.

Debridement with antibiotic treatment and retention of the implant appeared to be more successful with silver-coated implants.

Cite this article: Bone Joint J 2015;97-B:252–7.


Bone & Joint Research
Vol. 3, Issue 4 | Pages 101 - 107
1 Apr 2014
Edmondson MC Day R Wood D

Objectives

The most concerning infection of allografts and operative procedures is methicillin resistant Staphylococcus aureus (MRSA) and no current iontophoresed antibiotics effectively combat this microbe. It was initially hypothesised that iontophoresis of vancomycin through bone would not be effective due to its large molecular size and lack of charge. The aim of this study was to determine whether this was a viable procedure and to find the optimum conditions for its use.

Methods

An iontophoresis cell was set up with varying concentrations of Vancomycin within the medulla of a section of sheep tibia, sealed from an external saline solution. The cell was run for varying times, Vancomycin concentrations and voltages, to gain information on optimisation of conditions for impregnating the graft. Each graft was then sectioned and dust ground from the exposed surface. The dust was serially washed to extract the Vancomycin and concentrations measured and plotted for all variables tested.