The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
Total hip and knee arthroplasty (THA, TKA) are largely successful procedures; however, both have variable outcomes, resulting in some patients being dissatisfied with the outcome. Surgeons are turning to technologies such as robotic-assisted surgery in an attempt to improve outcomes. Robust studies are needed to find out if these innovations are really benefitting patients. The Robotic Arthroplasty Clinical and Cost Effectiveness Randomised Controlled Trials (RACER) trials are multicentre, patient-blinded randomized controlled trials. The patients have primary osteoarthritis of the hip or knee. The operation is Mako-assisted THA or TKA and the control groups have operations using conventional instruments. The primary clinical outcome is the Forgotten Joint Score at 12 months, and there is a built-in analysis of cost-effectiveness. Secondary outcomes include early pain, the alignment of the components, and medium- to long-term outcomes. This annotation outlines the need to assess these technologies and discusses the design and challenges when conducting such trials, including surgical workflows, isolating the effect of the operation, blinding, and assessing the learning curve. Finally, the future of robotic surgery is discussed, including the need to contemporaneously introduce and evaluate such technologies. Cite this article:
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF. Cite this article:
In this review, we discuss the evidence for patients returning to sport after hip arthroplasty. This includes the choices regarding level of sporting activity and revision or complications, the type of implant, fixation and techniques of implantation, and how these choices relate to health economics. It is apparent that despite its success over six decades, hip arthroplasty has now evolved to accommodate and support ever-increasing patient demands and may therefore face new challenges. Cite this article:
Periprosthetic joint infection (PJI) remains a challenging complication
following total hip arthroplasty (THA). It is associated with high
levels of morbidity, mortality and expense. Guidelines and protocols
exist for the management of culture-positive patients. Managing
culture-negative patients with a PJI poses a greater challenge to
surgeons and the wider multidisciplinary team as clear guidance
is lacking. We aimed to compare the outcomes of treatment for 50 consecutive
culture-negative and 50 consecutive culture-positive patients who
underwent two-stage revision THA for chronic infection with a minimum
follow-up of five years.Aims
Patients and Methods
Joint replacement of the hip and knee remain
very satisfactory operations. They are, however, expensive. The
actual manufacturing of the implant represents only 30% of the final
cost, while sales and marketing represent 40%. Recently, the patents
on many well established and successful implants have expired. Companies
have started producing and distributing implants that purport to
replicate existing implants with good long-term results. The aims of this paper are to assess the legality, the monitoring
and cost saving implications of such generic implants. We also assess
how this might affect the traditional orthopaedic implant companies. Cite this article:
The purpose of this article was to review the current literature
pertaining to the use of mobile compression devices (MCDs) for venous
thromboembolism (VTE) following total joint arthroplasty (TJA),
and to discuss the results of data from our institution. Previous studies have illustrated higher rates of post-operative
wound complications, re-operation and re-admission with the use
of more aggressive anticoagulation regimens, such as warfarin and
factor Xa inhibitors. This highlights the importance of the safety,
as well as efficacy, of the chemoprophylactic regimen.Aims
Patients and Methods