Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components. Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints. There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001). We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 656 - 660
1 May 2010
Ikävalko M Tiihonen R Skyttä ET Belt EA

Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1460 - 1463
1 Nov 2006
Landor I Vavrik P Jahoda D Guttler K Sosna A

We assessed the long-term results of 58 Souter-Strathclyde total elbow replacements in 49 patients with rheumatoid arthritis. The mean length of follow-up was 9.5 years (0.7 to 16.7). The mean pre-operative Mayo Elbow Performance Score was 30 (15 to 80) and at final follow-up was 82 (60 to 95). A total of 13 elbows (22.4%) were revised, ten (17.2%) for aseptic loosening, one (1.7%) for instability, one (1.7%) for secondary loosening after fracture, and one elbow (1.7%) was removed because of deep infection. The Kaplan-Meier survival rate was 70% and 53% at ten and 16 years, respectively. Failure of the ulnar component was found to be the main problem in relation to the loosening. Anterior transposition of the ulnar nerve had no influence on ulnar nerve paresthaesiae in these patients


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 78 - 83
1 Jan 2006
Sneftrup SB Jensen SL Johannsen HV Søjbjerg JO

We studied retrospectively the results of revision arthroplasty of the elbow using a linked Coonrad-Morrey implant in 23 patients (24 elbows) after a mean follow-up period of 55 months. According to the Mayo Elbow Performance Score, 19 elbows were satisfactory, nine were excellent and ten good. The median total score had improved from 35 points (20 to 75) before the primary arthroplasty to 85 points (40 to 100) at the latest follow-up. There was a marked relief of pain, but the range of movement showed no overall improvement. Two patients had a second revision because of infection and two for aseptic loosening. The estimated five-year survival rate of the prosthesis was 83.1% (95% confidence interval 61.1 to 93.3). Revision elbow arthroplasty using the Coonrad-Morrey implant provided satisfactory results but with complications occurring in 13 cases


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 1 | Pages 121 - 125
1 Jan 1998
Wachtl SW Guggenheim PR Sennwald GR

We performed 88 primary ball-and-socket arthroplasties of the trapeziometacarpal joint in 84 patients (69 women and 15 men) with a mean age of 61 years (37 to 81). Cemented de la Caffinière prostheses were implanted in 43 joints from 1988 to 1991 and 45 cementless Ledoux implants were used between 1992 and 1994. Of the 61 surviving prostheses still in situ, 51 were reviewed clinically and radiologically with a mean follow-up of 25 months for the Ledoux and 63 months for the de la Caffinière implant. The survival rate for the Ledoux prosthesis was 58.9% at 16 months and for the de la Caffinière implant 66.4% at 68 months. Loosening occurred in 15% of the Ledoux stems, in 46% of the Ledoux cups, in 24% of the de la Caffinière stems and in 28% of the de la Caffinière cups. Both prostheses behaved similarly, and it is clear that a constrained ball-and-socket prosthesis is not suitable for the trapeziometacarpal joint


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 671 - 677
1 May 2012
Raiss P Sowa B Bruckner T Eck S Woerz S Rohr K Rickert M Kasten P

The aim of this study was to compare a third-generation cementing procedure for glenoid components with a new technique for cement pressurisation. In 20 pairs of scapulae, 20 keeled and 20 pegged glenoid components were implanted using either a third-generation cementing technique (group 1) or a new pressuriser (group 2). Cement penetration was measured by three-dimensional (3D) analysis of micro-CT scans. The mean 3D depth of penetration of the cement was significantly greater in group 2 (p < 0.001). The mean thickness of the cement mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and 5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm (0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement mantle < 2 mm was detected less frequently in group 2 (p < 0.001). Using the cement pressuriser the proportion of cement mantles < 2 mm was significantly reduced compared with the third-generation cementing technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1540 - 1545
1 Nov 2012
Reigstad O Lütken T Grimsgaard C Bolstad B Thorkildsen R Røkkum M

The Motec cementless modular metal-on-metal ball-and-socket wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate advanced collapse (SLAC) in 30 patients (20 men) with severe (grades 3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of the patients was 52 years (31 to 71). All prostheses integrated well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1) no luxation or implant breakage occurred. Two wrists were converted to an arthrodesis for persistent pain. Loosening occurred in one further wrist at five years post-operatively. The remainder demonstrated close bone–implant contact. The clinical results were good, with markedly decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain scores, and increased movement and grip strength. No patient used analgesics and most had returned to work.

Good short-term function was achieved using this wrist arthroplasty in a high-demand group of patients with post-traumatic osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 75 - 81
1 Jan 2009
Cil A Veillette CJH Sanchez-Sotelo J Sperling JW Schleck C Cofield RH

Between 1976 and 2004, 38 revision arthroplasties (35 patients) were performed for aseptic loosening of the humeral component. The mean interval from primary arthroplasty to revision was 7.1 years (0.4 to 16.6). A total of 35 shoulders (32 patients) were available for review at a mean follow-up of seven years (2 to 19.3).

Pre-operatively, 34 patients (97%) had moderate or severe pain; at final follow-up, 29 (83%) had no or only mild pain (p < 0.0001). The mean active abduction improved from 88° to 107° (p < 0.01); and the mean external rotation from 37° to 46° (p = 0.27). Excellent or satisfactory results were achieved in 25 patients (71%) according to the modified Neer rating system. Humeral components were cemented in 29, with ingrowth implants used in nine cases. There were 19 of standard length and 17 were longer (two were custom replacements and are not included). Bone grafting was required for defects in 11 humeri. Only two glenoid components were left unrevised. Intra-operative complications included cement extrusion in eight cases, fracture of the shaft of the humerus is two and of the tuberosity in four. There were four re-operations, one for recurrent humeral loosening, with 89% survival free of re-operations at ten years.

Revision surgery for aseptic loosening of the humeral component provides reliable pain relief and modest improvement of movement, although there is a substantial risk of intra-operative complications. Revision to a total shoulder replacement gives better results than to a hemiarthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1416 - 1421
1 Oct 2010
Qureshi F Draviaraj KP Stanley D

Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation.

Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component.

With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91).

Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result.

Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1382 - 1389
1 Oct 2012
Sewell MD Kang SN Al-Hadithy N Higgs DS Bayley I Falworth M Lambert SM

There is little information about the management of peri-prosthetic fracture of the humerus after total shoulder replacement (TSR). This is a retrospective review of 22 patients who underwent a revision of their original shoulder replacement for peri-prosthetic fracture of the humerus with bone loss and/or loose components. There were 20 women and two men with a mean age of 75 years (61 to 90) and a mean follow-up 42 months (12 to 91): 16 of these had undergone a previous revision TSR. Of the 22 patients, 12 were treated with a long-stemmed humeral component that bypassed the fracture. All their fractures united after a mean of 27 weeks (13 to 94). Eight patients underwent resection of the proximal humerus with endoprosthetic replacement to the level of the fracture. Two patients were managed with a clam-shell prosthesis that retained the original components. The mean Oxford shoulder score (OSS) of the original TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival using re-intervention for any reason as the endpoint was 91% (95% confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at one and five years, respectively.

There were two revisions for dislocation of the humeral head, one open reduction for modular humeral component dissociation, one internal fixation for nonunion, one trimming of a prominent screw and one re-cementation for aseptic loosening complicated by infection, ultimately requiring excision arthroplasty. Two patients sustained nerve palsies.

Revision TSR after a peri-prosthetic humeral fracture associated with bone loss and/or loose components is a salvage procedure that can provide a stable platform for elbow and hand function. Good rates of union can be achieved using a stem that bypasses the fracture. There is a high rate of complications and function is not as good as with the original replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 235 - 241
1 Feb 2010
van der Lugt JCT Valstar ER Witvoet-Braam SW Nelissen RGHH

Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs.

All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (sd 1.0) and in the non-revised 0.7 mm (sd 0.5, p = 0.01). Most humeral components migrated into external rotation resulting in an anterior and varus tilt. The ulnar components remained stable.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 751 - 756
1 Jun 2008
Terrier A Reist A Merlini F Farron A

Reversed shoulder prostheses are increasingly being used for the treatment of glenohumeral arthropathy associated with a deficient rotator cuff. These non-anatomical implants attempt to balance the joint forces by means of a semi-constrained articular surface and a medialised centre of rotation. A finite element model was used to compare a reversed prosthesis with an anatomical implant. Active abduction was simulated from 0° to 150° of elevation. With the anatomical prosthesis, the joint force almost reached the equivalence of body weight. The joint force was half this for the reversed prosthesis. The direction of force was much more vertically aligned for the reverse prosthesis, in the first 90° of abduction. With the reversed prosthesis, abduction was possible without rotator cuff muscles and required 20% less deltoid force to achieve it.

This force analysis confirms the potential mechanical advantage of reversed prostheses when rotator cuff muscles are deficient.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1341 - 1346
1 Oct 2006
Gille J Ince A González O Katzer A Loehr JF

This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection.

Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was Staphylococcus aureus. Four of the six elbows had a developed radiolucency around one component or the other. Successful single-stage exchange arthroplasty was carried out with antibiotic-loaded cement in five of the six cases. In one, the revision prosthesis had to be removed following recurrence of the infection. The functional result was good in three elbows, fair in one, poor in one and fair in the resection arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 345 - 350
1 Mar 2006
Loew M Heitkemper S Parsch D Schneider S Rickert M

We reviewed 39 patients with displaced three- and four-part fractures of the humerus. In 21 patients (group A) we had used an anatomical prosthesis for the humeral head and in 18 (group B) an implant designed for fractures.

When followed up at a mean of 29.3 months after surgery the overall Constant score was 51.9 points; in group A it was 51.5 and in group B 52.4 points. The subjective satisfaction of the patients was assessed using a numerical rating scale and was similar in both groups. In group A complete healing of the tuberosities was found in 29% and 50% in group B. Partial integration was seen in 29% of group A and in only one patient in group B, while resorption was noted in 43% of group A and 44% of group B. The functional outcome was significantly better in patients with complete or partial healing of the tuberosities (p = 0.022). The specific trauma prosthesis did not lead to better healing of the tuberosities. The difference in clinical outcome obtained by the two designs did not reach statistical significance.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 4 | Pages 496 - 501
1 Apr 2006
Haines JF Trail IA Nuttall D Birch A Barrow A

We have undertaken a prospective clinical and radiological analysis of 124 shoulder arthroplasties (113 patients) carried out for osteoarthritis. The clinical results showed improvement in the absolute Constant score and the American Shoulder and Elbow Surgeons score of 22 and 43, respectively. Both were statistically significant (p < 0.001). There was no significant difference in the scores after hemiarthroplasty and total arthroplasty in those patients with an intact rotator cuff.

When revision was used as the end-point for survival at ten years, survival of 86%, or 90% if glenoid components made of Hylamer sterilised in air were omitted, was obtained in primary osteoarthritis. The most common cause for revision in the hemiarthroplasty group was glenoid pain at a mean of 1.5 years; in the total arthroplasty group it was loosening of the glenoid at a mean of 4.5 years. Analysis of pre-operative factors showed that the risk of gross loosening of the glenoid increased threefold when there was evidence of erosion of the glenoid at operation. Shoulder arthroplasty should not be delayed once symptomatic osteoarthritis has been established and should be undertaken before failure of the cuff or erosion of the glenoid are present.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 950 - 954
1 Jul 2005
Khatri M Stirrat AN

We present the outcome of 47 Souter-Strathclyde replacements of the elbow with a mean follow-up of 82 months (12 to 129). The clinical results were assessed using a condition-specific outcome measure. The mean total score (maximum 100) before the operation was 47.21 and improved to 79.92 (p < 0.001). The mean pain score (maximum 50) improved from 21.41 to 46.70 (p < 0.001) and the mean functional component of the score (maximum 30) from 11.19 to 18.65 (p < 0.001). There was negligible change in the score for the range of movement although a significant improvement in mean flexion from 124° to 136° was noted (p < 0.001).

Revision surgery was required in four patients, for dislocation, wound dehiscence and early infection in one, late infection in two and aseptic loosening in one. The cumulative survival was 75% at nine years for all causes of failure and 97% at ten years for aseptic loosening alone. Our study demonstrates the value of the Souter-Strathclyde total elbow arthroplasty in providing relief from pain and functional improvement in rheumatoid patients.