We analysed 42 weight-bearing lateral radiographs of the ankle, 20 of which were from patients with a clinical and plain radiological diagnosis of talocalcaneal coalition (TCC) who subsequently had CT. The remainder were from 22 healthy volunteers with no clinical findings suggestive of hindfoot pathology. Four observers, blinded to the CT findings, independently evaluated the radiographs on two separate occasions. With the 95% confidence interval and using the CT findings as the comparison we calculated the sensitivity, specificity, accuracy, and positive and negative predictive values for the C-sign, and for other signs known to be associated with TCC. Similarly, we also calculated the interobserver and intraobserver
Injury to the dorsomedial cutaneous nerve in the foot may occur after operations for hallux valgus. Pressure neuropathy before operation is also described but remains largely unexplored. We have investigated the incidence of sensory deficit in the great toe before operating for hallux valgus and examined to what extent any deficit was related to the degree of angulation of the joint. Forty-three patients with a total of 61 great toes with hallux valgus presenting for consideration of surgical correction had their sensation tested in pre-designated zones using a five-filament set of Semmes-Weinstein monofilaments. These allowed good inter-observer
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).