Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154). Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications. We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis.
There are theoretical and practical advantages to modular rather than monoblock designs of prostheses for shoulder arthroplasty, but there are no reported studies which specifically compare the clinical and radiological results of their use. We have compared the results of unconstrained total shoulder arthroplasty for osteoarthritis using both types of implant. The monoblock design was used between 1992 and 1995 and the modular design after 1995. Both had cemented all-polyethylene glenoids, the monoblock with matched and the modular with mismatched radii of curvature. There were 34 consecutive shoulders in each group with a mean follow-up of 6.1 years in the first and 5.2 years in the second. There were no significant differences in improvement of pain scores, active elevation, external rotation, internal rotation, patient satisfaction, or the Neer ratings between the two groups. Two of 28 glenoid components in the first group and six of 30 in the second met the criteria for being radiologically at risk for loosening (p = 0.25). There were no significant differences in clinical outcome or radiological changes between the first- and second-generation designs of implant for shoulder arthroplasty.
In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis. Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p <
0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p <
0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them.
We report the long-term clinical and radiological outcomes of the Aequalis total shoulder replacement with a cemented all-polyethylene flat-back keeled glenoid component implanted for primary osteoarthritis between 1991 and 2003 in nine European centres. A total of 226 shoulders in 210 patients were retrospectively reviewed at a mean of 122.7 months (61 to 219) or at revision. Clinical outcome was assessed using the Constant score, patient satisfaction score and range of movement. Kaplan-Meier survivorship analysis was performed with glenoid revision for loosening and radiological glenoid loosening ( Younger patient age and the curettage technique for glenoid preparation correlated with loosening. The rate of glenoid revision and radiological loosening increased with duration of follow-up, but not until a follow-up of five years. Therefore, we recommend that future studies reporting radiological outcomes of new glenoid designs should report follow-up of at least five to ten years.