Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 367 - 370
1 Mar 2013
Bini SA Chen Y Khatod M Paxton EW

We evaluated the impact of pre-coating the tibial component with polymethylmethacrylate (PMMA) on implant survival in a cohort of 16 548 primary NexGen total knee replacements (TKRs) in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while in 2713 TKRs the non-pre-coated version of the same tray was used. All the TKRs were performed between 2001 and 2009 and were cemented. TKRs implanted with a pre-coated tibial component had a lower cumulative survival than those with a non-pre-coated tibial component (p = 0.01). After adjusting for diagnosis, age, gender, body mass index, American Society of Anesthesiologists grade, femoral coupling design, surgeon volume and hospital volume, pre-coating was an independent risk factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006). Revision for aseptic loosening was uncommon for both pre-coated and non-pre-coated trays (rates of 0.12% and 0%, respectively). Pre-coating with PMMA does not appear to be protective of revision for this tibial tray design at short-term follow-up.

Cite this article: Bone Joint J 2013;95-B:367–70.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 848 - 851
1 Jul 2019
Sautet P Parratte S Mékidèche T Abdel MP Flécher X Argenson J Ollivier M

Aims

The aims of this study were to compare the mean duration of antibiotic release and the mean zone of inhibition between vancomycin-loaded porous tantalum cylinders and antibiotic-loaded bone cement at intervals, and to evaluate potential intrinsic antimicrobial properties of tantalum in an in vitro medium environment against methicillin-sensitive Staphylococcus aureus (MSSA).

Materials and Methods

Ten porous tantalum cylinders and ten cylinders of cement were used. The tantalum cylinders were impregnated with vancomycin, which was also added during preparation of the cylinders of cement. The cylinders were then placed on agar plates inoculated with MSSA. The diameter of the inhibition zone was measured each day, and the cylinders were transferred to a new inoculated plate. Inhibition zones were measured with a Vernier caliper and using an automated computed evaluation, and the intra- and interobserver reproducibility were measured. The mean inhibition zones between the two groups were compared with Wilcoxon’s test.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 497 - 503
1 Apr 2012
Bercovy M Beldame J Lefebvre B Duron A

This prospective study compares the outcome of 157 hydroxyapatite (HA)-coated tibial components with 164 cemented components in the ROCC Rotating Platform total knee replacement in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There were two revisions for loosening: one for an HA-coated and one for a cemented tibial component. Radiological evaluation demonstrated no radiolucent lines with the HA-coated femoral components. A total of three HA-coated tibial components exhibited radiolucent lines at three months post-operatively and these disappeared after three further months of protected weight-bearing. With HA-coated components the operating time was shorter (p <  0.006) and the radiological assessment of the tibial interface was more stable (p < 0.01). Using revision for aseptic loosening of the tibial component as the end point, the survival rates at nine years was identical for both groups at 99.1%.

Our results suggest that HA-coated components perform at least as well as the same design with cemented components and compare favourably with those of series describing cemented or porous-coated knee replacements, suggesting that fixation of both components with hydroxyapatite is a reliable option in primary total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 49 - 53
1 Jan 2006
Tsumara N Yoshiya S Chin T Shiba R Kohso K Doita M

We undertook a prospective, randomised study in order to evaluate the efficacy of clamping the drains after intra-articular injection of saline with 1:500 000 adrenaline compared with post-operative blood salvage in reducing blood loss in 212 total knee arthroplasties. The mean post-operative drained blood volume after drain clamping was 352.1 ml compared to 662.3 ml after blood salvage (p < 0.0001). Allogenic blood transfusion was needed in one patient in the drain group and for three in the blood salvage group. Drain clamping with intra-articular injection of saline with adrenaline is more effective than post-operative autologous blood transfusion in reducing blood loss during total knee arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component.

All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant.

This pattern of subsidence is regarded as being beneficial for uncemented fixation.