Tissue responses to debris formed by abrasion of polymethylmethacrylate
(PMMA) spacers at two-stage revision arthroplasty for prosthetic
joint infection are not well described. We hypothesised that PMMA
debris induces immunomodulation in periprosthetic tissues. Samples of tissue were taken during 35 two-stage revision arthroplasties
(nine total hip and 26 total knee arthroplasties) in patients whose
mean age was 67 years (44 to 85). Fourier transform infrared microscopy
was used to confirm the presence of PMMA particles. Histomorphometry
was performed using Sudan Red and Haematoxylin-Eosin staining.
CD-68, CD-20, CD-11(c), CD-3 and IL-17 antibodies were used to immunophenotype
the inflammatory cells. All slides were scored semi-quantitatively
using the modified Willert scoring system.Aims
Patients and Methods
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article: