Results 1 - 20 of 118
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 76 - 81
1 Jan 2005
Pajarinen J Lindahl J Michelsson O Savolainen V Hirvensalo E

We treated 108 patients with a pertrochanteric femoral fracture using either the dynamic hip screw or the proximal femoral nail in this prospective, randomised series. We compared walking ability before fracture, intra-operative variables and return to their residence. Patients treated with the proximal femoral nail (n = 42) had regained their pre-operative walking ability significantly (p = 0.04) more often by the four-month review than those treated with the dynamic hip screw (n = 41). Peri-operative or immediate post-operative measures of outcome did not differ between the groups, with the exception of operation time. The dynamic hip screw allowed a significantly greater compression of the fracture during the four-month follow-up, but consolidation of the fracture was comparable between the two groups. Two major losses of reduction were observed in each group, resulting in a total of four revision operations. Our results suggest that the use of the proximal femoral nail may allow a faster postoperative restoration of walking ability, when compared with the dynamic hip screw


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 616 - 621
1 May 2011
O’Neill F Condon F McGloughlin T Lenehan B Coffey JC Walsh M

We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study


The Journal of Bone & Joint Surgery British Volume
Vol. 68-B, Issue 4 | Pages 557 - 560
1 Aug 1986
Esser M Kassab J Jones D

In a randomised prospective trial 98 elderly women with trochanteric fractures of the femur were treated with either a 135 degrees Jewett nail-plate or a 135 degrees Dynamic hip screw. The results at six weeks, three months and six months were statistically analysed. There were no significant differences in the two groups with regard to pain, length of hospital stay, morbidity or mortality. Although operative difficulties and open reduction were more common with the Dynamic hip screw, at the end of six months more patients in this group were mobile and there was significant radiological evidence of better compression without loss of fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 5 | Pages 794 - 796
1 Sep 1990
Richards R Evans G Egan J Shearer

We compared, under laboratory conditions, the resistance to cutting out of the AO dynamic hip screw and the Pugh sliding nail. The mean load at cut out, adjusted for bone strength, was 70% greater for the Pugh sliding nail. The reasons for this difference are discussed


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1073 - 1078
1 Aug 2008
Little NJ Verma V Fernando C Elliott DS Khaleel A

We compared the outcome of patients treated for an intertrochanteric fracture of the femoral neck with a locked, long intramedullary nail with those treated with a dynamic hip screw (DHS) in a prospective randomised study. Each patient who presented with an extra-capsular hip fracture was randomised to operative stabilisation with either a long intramedullary Holland nail or a DHS. We treated 92 patients with a Holland nail and 98 with a DHS. Pre-operative variables included the Mini Mental test score, patient mobility, fracture pattern and American Society of Anesthesiologists grading. Peri-operative variables were anaesthetic time, operating time, radiation time and blood loss. Post-operative variables were time to mobilising with a frame, wound infection, time to discharge, time to fracture union, and mortality. We found no significant difference in the pre-operative variables. The mean anaesthetic and operation times were shorter in the DHS group than in the Holland nail group (29.7 vs 40.4 minutes, p < 0.001; and 40.3 vs 54 minutes, p < 0.001, respectively). There was an increased mean blood loss within the DHS group versus the Holland nail group (160 ml vs 78 ml, respectively, p < 0.001). The mean time to mobilisation with a frame was shorter in the Holland nail group (DHS 4.3 days, Holland nail 3.6 days, p = 0.012). More patients needed a post-operative blood transfusion in the DHS group (23 vs seven, p = 0.003) and the mean radiation time was shorter in this group (DHS 0.9 minutes vs Holland nail 1.56 minutes, p < 0.001). The screw of the DHS cut out in two patients, one of whom underwent revision to a Holland nail. There were no revisions in the Holland nail group. All fractures in both groups were united when followed up after one year. We conclude that the DHS can be implanted more quickly and with less exposure to radiation than the Holland nail. However, the resultant blood loss and need for transfusion is greater. The Holland nail allows patients to mobilise faster and to a greater extent. We have therefore adopted the Holland nail as our preferred method of treating intertrochanteric fractures of the hip


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 5 | Pages 789 - 793
1 Sep 1993
Radford P Needoff M Webb J

We made a randomised prospective comparison of the Dynamic Hip Screw and the Gamma locking nail for the internal fixation of 200 petrochanteric femoral fractures in elderly patients. There was less intraoperative blood loss and a lower rate of wound complications in the patients treated by the Gamma nail. They had, however, a high incidence of femoral shaft fracture which we relate in part to implant design. We do not recommend the use of the Gamma nail for these fractures


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 2 | Pages 330 - 334
1 Mar 1991
Bridle S Patel A Bircher M Calvert P

We have prospectively compared the fixation of 100 intertrochanteric fractures of the proximal femur in elderly patients with random use of either a Dynamic Hip Screw (DHS) or a new intramedullary device, the Gamma nail. We found no difference in operating time, blood loss, wound complications, stay in hospital, place of eventual discharge, or the patients' mobility at final review. There was no difference in failure of proximal fixation: cut-out occurred in three cases with the DHS, and twice with the Gamma nail. However, in four cases fracture of the femur occurred close to the Gamma nail, requiring further major surgery. In the absence of these complications, union was seen by six months in both groups


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 3 | Pages 345 - 351
1 May 1992
Leung K So W Shen W Hui P

The Gamma nail was introduced for the treatment of peritrochanteric fractures with the theoretical advantage of a load-sharing femoral component which could be implanted by a closed procedure. We report a randomised prospective study of 186 fractures treated by either the Gamma nail or a dynamic hip screw. Gamma nails were implanted with significantly shorter screening times, smaller incisions, and less intraoperative bleeding. The Gamma nail group had a shorter convalescence and earlier full weight-bearing, but there was no significant difference in mortality within six months, postoperative mobility, or hip function at review. More intra-operative complications were recorded in the Gamma nail group, mainly due to the mismatching of the femoral component of the nail to the small femurs of Chinese people. Use of a smaller modified nail reduced these complications. We conclude that with careful surgical technique and the modified femoral component, the Gamma nail is an advance in the treatment of peritrochanteric fractures


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 4 | Pages 625 - 625
1 Jul 1992
Parker M


Bone & Joint Research
Vol. 1, Issue 6 | Pages 104 - 110
1 Jun 2012
Swinteck BJ Phan DL Jani J Owen JR Wayne JS Mounasamy V

Objectives. The use of two implants to manage concomitant ipsilateral femoral shaft and proximal femoral fractures has been indicated, but no studies address the relationship of dynamic hip screw (DHS) side plate screws and the intramedullary nail where failure might occur after union. This study compares different implant configurations in order to investigate bridging the gap between the distal DHS and tip of the intramedullary nail. Methods. A total of 29 left synthetic femora were tested in three groups: 1) gapped short nail (GSN); 2) unicortical short nail (USN), differing from GSN by the use of two unicortical bridging screws; and 3) bicortical long nail (BLN), with two angled bicortical and one unicortical bridging screws. With these findings, five matched-pairs of cadaveric femora were tested in two groups: 1) unicortical long nail (ULN), with a longer nail than USN and three bridging unicortical screws; and 2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally rotated 90°/sec until failure. Results. For synthetic femora, a difference was detected between GSN and BLN in energy to failure (p = 0.04) and torque at failure (p = 0.02), but not between USN and other groups for energy to failure (vs GSN, p = 0.71; vs BLN, p = 0.19) and torque at failure (vs GSN, p = 0.55; vs BLN, p = 0.15). For cadaveric femora, ULN and BLN performed similarly because of the improvement provided by the bridging screws. Conclusions. Our study shows that bicortical angled screws in the DHS side plate are superior to no screws at all in this model and loading scenario, and suggests that adding unicortical screws to a gapped construct is probably beneficial


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1134 - 1138
1 Aug 2013
Hsu C Shih C Wang C Huang K

Although the importance of lateral femoral wall integrity is increasingly being recognised in the treatment of intertrochanteric fracture, little attention has been put on the development of a secondary post-operative fracture of the lateral wall. Patients with post-operative fractures of the lateral wall were reported to have high rates of re-operation and complication. To date, no predictors of post-operative lateral wall fracture have been reported. In this study, we investigated the reliability of lateral wall thickness as a predictor of lateral wall fracture after dynamic hip screw (DHS) implantation.

A total of 208 patients with AO/OTA 31-A1 and -A2 classified intertrochanteric fractures who received internal fixation with a DHS between January 2003 and May 2012 were reviewed. There were 103 men and 150 women with a mean age at operation of 78 years (33 to 94). The mean follow-up was 23 months (6 to 83). The right side was affected in 97 patients and the left side in 111. Clinical information including age, gender, side, fracture classification, tip–apex distance, follow-up time, lateral wall thickness and outcome were recorded and used in the statistical analysis.

Fracture classification and lateral wall thickness significantly contributed to post-operative lateral wall fracture (both p < 0.001). The lateral wall thickness threshold value for risk of developing a secondary lateral wall fracture was found to be 20.5 mm.

To our knowledge, this is the first study to investigate the risk factors of post-operative lateral wall fracture in intertrochanteric fracture. We found that lateral wall thickness was a reliable predictor of post-operative lateral wall fracture and conclude that intertrochanteric fractures with a lateral wall thickness < 20.5 mm should not be treated with DHS alone.

Cite this article: Bone Joint J 2013;95-B:1134–8.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 602 - 611
21 Aug 2023
James HK Pattison GTR Griffin J Fisher JD Griffin DR

Aims. To evaluate if, for orthopaedic trainees, additional cadaveric simulation training or standard training alone yields superior radiological and clinical outcomes in patients undergoing dynamic hip screw (DHS) fixation or hemiarthroplasty for hip fracture. Methods. This was a preliminary, pragmatic, multicentre, parallel group randomized controlled trial in nine secondary and tertiary NHS hospitals in England. Researchers were blinded to group allocation. Overall, 40 trainees in the West Midlands were eligible: 33 agreed to take part and were randomized, five withdrew after randomization, 13 were allocated cadaveric training, and 15 were allocated standard training. The intervention was an additional two-day cadaveric simulation course. The control group received standard on-the-job training. Primary outcome was implant position on the postoperative radiograph: tip-apex distance (mm) (DHS) and leg length discrepancy (mm) (hemiarthroplasty). Secondary clinical outcomes were procedure time, length of hospital stay, acute postoperative complication rate, and 12-month mortality. Procedure-specific secondary outcomes were intraoperative radiation dose (for DHS) and postoperative blood transfusion requirement (hemiarthroplasty). Results. Eight female (29%) and 20 male trainees (71%), mean age 29.4 years, performed 317 DHS operations and 243 hemiarthroplasties during ten months of follow-up. Primary analysis was a random effect model with surgeon-level fixed effects of patient condition, patient age, and surgeon experience, with a random intercept for surgeon. Under the intention-to-treat principle, for hemiarthroplasty there was better implant position in favour of cadaveric training, measured by leg length discrepancy ≤ 10 mm (odds ratio (OR) 4.08 (95% confidence interval (CI) 1.17 to 14.22); p = 0.027). There were significantly fewer postoperative blood transfusions required in patients undergoing hemiarthroplasty by cadaveric-trained compared to standard-trained surgeons (OR 6.00 (95% CI 1.83 to 19.69); p = 0.003). For DHS, there was no significant between-group difference in implant position as measured by tip-apex distance ≤ 25 mm (OR 6.47 (95% CI 0.97 to 43.05); p = 0.053). No between-group differences were observed for any secondary clinical outcomes. Conclusion. Trainees randomized to additional cadaveric training performed hip fracture fixation with better implant positioning and fewer postoperative blood transfusions in hemiarthroplasty. This effect, which was previously unknown, may be a consequence of the intervention. Further study is required. Cite this article: Bone Jt Open 2023;4(8):602–611


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 391 - 397
1 Mar 2015
van Embden D Stollenwerck GANL Koster LA Kaptein BL Nelissen RGHH Schipper IB

The aim of this study was to quantify the stability of fracture-implant complex in fractures after fixation. A total of 15 patients with an undisplaced fracture of the femoral neck, treated with either a dynamic hip screw or three cannulated hip screws, and 16 patients with an AO31-A2 trochanteric fracture treated with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric analysis was used at six weeks, four months and 12 months post-operatively to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of the femoral neck and seven with a trochanteric fracture. By four months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1) had occurred in the fracture of the femoral neck group and 5.0 mm (-0.13 to 12.9) in the trochanteric fracture group. A wide range of rotation occurred in both types of fracture. Right-sided trochanteric fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site occurs continuously during the first four post-operative months, after which stabilisation occurs. This information may allow the early recognition of patients at risk of failure of fixation. Cite this article: Bone Joint J 2015;97-B:391–7


Bone & Joint Open
Vol. 1, Issue 9 | Pages 594 - 604
24 Sep 2020
James HK Pattison GTR Griffin J Fisher JD Griffin DR

Aims. To develop a core outcome set of measurements from postoperative radiographs that can be used to assess technical skill in performing dynamic hip screw (DHS) and hemiarthroplasty, and to validate these against Van der Vleuten’s criteria for effective assessment. Methods. A Delphi exercise was undertaken at a regional major trauma centre to identify candidate measurement items. The feasibility of taking these measurements was tested by two of the authors (HKJ, GTRP). Validity and reliability were examined using the radiographs of operations performed by orthopaedic resident participants (n = 28) of a multicentre randomized controlled educational trial (ISRCTN20431944). Trainees were divided into novice and intermediate groups, defined as having performed < ten or ≥ ten cases each for DHS and hemiarthroplasty at baseline. The procedure-based assessment (PBA) global rating score was assumed as the gold standard assessment for the purposes of concurrent validity. Intra- and inter-rater reliability testing were performed on a random subset of 25 cases. Results. In total, 327 DHS and 248 hemiarthroplasty procedures were performed by 28 postgraduate year (PGY) 3 to 5 orthopaedic trainees during the 2014 to 2015 surgical training year at nine NHS hospitals in the West Midlands, UK. Overall, 109 PBAs were completed for DHS and 80 for hemiarthroplasty. Expert consensus identified four ‘final product analysis’ (FPA) radiological parameters of technical success for DHS: tip-apex distance (TAD); lag screw position in the femoral head; flushness of the plate against the lateral femoral cortex; and eight-cortex hold of the plate screws. Three parameters were identified for hemiarthroplasty: leg length discrepancy; femoral stem alignment; and femoral offset. Face validity, content validity, and feasibility were excellent. For all measurements, performance was better in the intermediate compared with the novice group, and this was statistically significant for TAD (p < 0.001) and femoral stem alignment (p = 0.023). Concurrent validity was poor when measured against global PBA score. This may be explained by the fact that they are measuring difference facets of competence. Intra-and inter-rater reliability were excellent for TAD, moderate for lag screw position (DHS), and moderate for leg length discrepancy (hemiarthroplasty). Use of a large multicentre dataset suggests good generalizability of the results to other settings. Assessment using FPA was time- and cost-effective compared with PBA. Conclusion. Final product analysis using post-implantation radiographs to measure technical skill in hip fracture surgery is feasible, valid, reliable, and cost-effective. It can complement traditional workplace-based assessment for measuring performance in the real-world operating room . It may have particular utility in competency-based training frameworks and for assessing skill transfer from the simulated to live operating theatre. Cite this article: Bone Joint Open 2020;1-9:594–604


Bone & Joint Open
Vol. 1, Issue 8 | Pages 500 - 507
18 Aug 2020
Cheruvu MS Bhachu DS Mulrain J Resool S Cool P Ford DJ Singh RA

Aims. Our rural orthopaedic service has undergone service restructure during the COVID-19 pandemic in order to sustain hip fracture care. All adult trauma care has been centralised to the Royal Shrewsbury Hospital for assessment and medical input, before transferring those requiring operative intervention to the Robert Jones and Agnes Hunt Orthopaedic Hospital. We aim to review the impact of COVID-19 on hip fracture workload and service changes upon management of hip fractures. Methods. We reviewed our prospectively maintained trust database and National Hip Fracture Database records for the months of March and April between the years 2016 and 2020. Our assessment included fracture pattern (intrascapular vs extracapsular hip fracture), treatment intervention, length of stay and mortality. Results. We treated 288 patients during March and April between 2016 and 2020, with a breakdown of 55, 58, 53, 68, and 54 from 2016 to 2020 respectively. Fracture pattern distribution in the pre-COVID-19 years of 2016 to 2019 was 58% intracapsular and 42% extracapsular. In 2020 (COVID-19 period) the fracture patterns were 65% intracapsular and 35% extracapsular. Our mean length of stay was 13.1 days (SD 8.2) between 2016 to 2019, and 5.0 days (6.3) days in 2020 (p < 0.001). Between 2016 and 2019 we had three deaths in hip fracture patients, and one death in 2020. Hemiarthroplasty and dynamic hip screw fixation have been the mainstay of operative intervention across the five years and this has continued in the COVID-19 period. We have experienced a rise in conservatively managed patients; ten in 2020 compared to 14 over the previous four years. Conclusion. There has not been a reduction in the number of hip fractures during COVID-19 period compared to the same time period over previous years. In our experience, there has been an increase in conservative treatment and decreased length of stay during the COVID -19 period. Cite this article: Bone Joint Open 2020;1-8:500–507


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 162 - 169
1 Feb 2020
Hoellwarth JS Tetsworth K Kendrew J Kang NV van Waes O Al-Maawi Q Roberts C Al Muderis M

Aims. Osseointegrated prosthetic limbs allow better mobility than socket-mounted prosthetics for lower limb amputees. Fractures, however, can occur in the residual limb, but they have rarely been reported. Approximately 2% to 3% of amputees with socket-mounted prostheses may fracture within five years. This is the first study which directly addresses the risks and management of periprosthetic osseointegration fractures in amputees. Methods. A retrospective review identified 518 osseointegration procedures which were undertaken in 458 patients between 2010 and 2018 for whom complete medical records were available. Potential risk factors including time since amputation, age at osseointegration, bone density, weight, uni/bilateral implantation and sex were evaluated with multiple logistic regression. The mechanism of injury, technique and implant that was used for fixation of the fracture, pre-osseointegration and post fracture mobility (assessed using the K-level) and the time that the prosthesis was worn for in hours/day were also assessed. Results. There were 22 periprosthetic fractures; they occurred exclusively in the femur: two in the femoral neck, 14 intertrochanteric and six subtrochanteric, representing 4.2% of 518 osseointegration operations and 6.3% of 347 femoral implants. The vast majority (19/22, 86.4%) occurred within 2 cm of the proximal tip of the implant and after a fall. No fractures occurred spontaneously. Fixation most commonly involved dynamic hip screws (10) and reconstruction plates (9). No osseointegration implants required removal, the K-level was not reduced after fixation of the fracture in any patient, and all retained a K-level of ≥ 2. All fractures united, 21 out of 22 patients (95.5%) wear their osseointegration-mounted prosthetic limb longer daily than when using a socket, with 18 out of 22 (81.8%) reporting using it for ≥ 16 hours daily. Regression analysis identified a 3.89-fold increased risk of fracture for females (p = 0.007) and a 1.02-fold increased risk of fracture per kg above a mean of 80.4 kg (p = 0.046). No increased risk was identified for bilateral implants (p = 0.083), time from amputation to osseointegration (p = 0.974), age at osseointegration (p = 0.331), or bone density (g/cm2, p = 0.560; T-score, p = 0.247; Z-score, p = 0.312). Conclusion. The risks and sequelae of periprosthetic fracture after press-fit osseointegration for amputation should not deter patients or clinicians from considering this procedure. Females and heavier patients are likely to have an increased risk of fracture. Age, years since amputation, and bone density do not appear influential. Cite this article: Bone Joint J 2020;102-B(2):162–169


Bone & Joint Research
Vol. 6, Issue 5 | Pages 270 - 276
1 May 2017
Gosiewski JD Holsgrove TP Gill HS

Objectives. Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Methods. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm. 3. , 0.16 g/cm. 3. and 0.24 g/cm. 3. ). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°. Results. The deployable crucifix design (X-Bolt), was more torsionally stable, compared to both the dynamic hip screw (DHS, p = 0.008) and helical blade (DHS Blade, p= 0.008) designs in bone substitute material representative of osteoporotic bone (0.16 g/cm. 3. polyurethane foam). In 0.08 g/cm. 3. density substrate, the crucifix design (X-Bolt) had a higher resistance to torsion than the screw (DHS, p = 0.008). There were no significant differences (p = 0.101) between the implants in 0.24 g/cm. 3. density bone substitute. Conclusions. Our findings indicate that the clinical standard proximal fracture fixator design, the screw (DHS), was the least effective at resisting torsional load, and a novel crucifix design (X-Bolt), was the most effective design in resisting torsional load in bone substitute material with density representative of osteoporotic bone. At other densities the torsional stability was also higher for the X-Bolt, although not consistently significant by statistical analysis. Cite this article: J. D. Gosiewski, T. P. Holsgrove, H. S. Gill. The efficacy of rotational control designs in promoting torsional stability of hip fracture fixation. Bone Joint Res 2017;6:270–276. DOI: 10.1302/2046-3758.65.BJR-2017-0287.R1


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1198 - 1201
1 Sep 2014
Magu NK Singla R Rohilla R Gogna P Mukhopadhyay R Singh A

We report the outcome of 39 patients who underwent a modified Pauwels’ intertrochanteric osteotomy for nonunion of a femoral neck fracture following failed osteosynthesis. There were 31 men and eight women with a mean age of 47.2 years (34 to 59). By Pauwels’ classification, there were 11 Type II fractures and 28 Type III fractures. The mean follow-up was 7.9 years (2 to 19). In the 11 patients whose initial treatment had been osteotomy, union was achieved in nine (81.8%). In 28 patients whose initial treatment had been with a lag screw or a dynamic hip screw, union was achieved in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients (87.5%) with pre-operative shortening. The mean neck-shaft angle improved significantly from 100.5° (80° to 120°) to 131.6° (120° to 155°) (p = 0.004). The mean modified Harris hip score was 85.6 points (70 to 97) and the mean modified Merle d’Aubigné score was 14.3 (11 to 18). Good to excellent functional outcomes were achieved in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy is a reliable method of treating ununited fractures of the femoral neck following failed osteosynthesis: coxa vara and shortening can also simultaneously be addressed. Cite this article: Bone Joint J 2014;96-B:1198–1201


Bone & Joint Research
Vol. 1, Issue 4 | Pages 50 - 55
1 Apr 2012
O’Neill F Condon F McGloughlin T Lenehan B Coffey C Walsh M

Introduction. The objective of this study was to determine if a synthetic bone substitute would provide results similar to bone from osteoporotic femoral heads during in vitro testing with orthopaedic implants. If the synthetic material could produce results similar to those of the osteoporotic bone, it could reduce or eliminate the need for testing of implants on bone. Methods. Pushout studies were performed with the dynamic hip screw (DHS) and the DHS Blade in both cadaveric femoral heads and artificial bone substitutes in the form of polyurethane foam blocks of different density. The pushout studies were performed as a means of comparing the force displacement curves produced by each implant within each material. Results. The results demonstrated that test material with a density of 0.16 g/cm. 3. (block A) produced qualitatively similar force displacement curves for the DHS and qualitatively and quantitatively similar force displacement curves for the DHS Blade, whereas the test material with a density of 0.08 g/cm. 3. (block B) did not produce results that were predictive of those recorded within the osteoporotic cadaveric femoral heads. Conclusion. This study demonstrates that synthetic material with a density of 0.16 g/cm. 3. can provide a good substitute for cadaveric osteoporotic femoral heads in the testing of implants. However we do recognise that no synthetic material can be considered as a definitive substitute for bone, therefore studies performed with artificial bone substrates may need to be validated by further testing with a small bone sample in order to produce conclusive results


Bone & Joint Open
Vol. 3, Issue 11 | Pages 907 - 912
23 Nov 2022
Hurley RJ McCabe FJ Turley L Maguire D Lucey J Hurson CJ

Aims

The use of fluoroscopy in orthopaedic surgery creates risk of radiation exposure to surgeons. Appropriate personal protective equipment (PPE) can help mitigate this. The primary aim of this study was to assess if current radiation protection in orthopaedic trauma is safe. The secondary aims were to describe normative data of radiation exposure during common orthopaedic procedures, evaluate ways to improve any deficits in protection, and validate the use of electronic personal dosimeters (EPDs) in assessing radiation dose in orthopaedic surgery.

Methods

Radiation exposure to surgeons during common orthopaedic trauma operations was prospectively assessed using EPDs and thermoluminescent dosimeters (TLDs). Normative data for each operation type were calculated and compared to recommended guidelines.