Results 1 - 20 of 38
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1013 - 1018
1 Aug 2008
Goto K Kawanabe K Akiyama H Morimoto T Nakamura T

We reviewed 44 consecutive revision hip replacements in 38 patients performed using the cement-in-cement technique. All were performed for acetabular loosening in the presence of a well-fixed femoral component. The mean follow-up was 5.1 years (2 to 10.1). Radiological analysis at final follow-up indicated no loosening of the femoral component, except for one case with a continuous radiolucent line in all zones and peri-prosthetic fracture which required further revision. Peri-operative complications included nine proximal femoral fractures (20.4%) and perforation of the proximal femur in one hip. In five hips wiring or fixation with a braided suture was undertaken but no additional augmentation was required. There was an improvement in the mean Japanese Orthopaedic Association score from 55.5 (28 to 81) pre-operatively to 77.8 (40 to 95) at final follow-up (p < 0.001). Revision using a cement-in-cement technique allows increased exposure for acetabular revision and is effective in the medium term. Further follow-up is required to assess the long-term results in the light of in vitro studies which have questioned the quality of the cement-in-cement bond


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1450 - 1457
1 Nov 2017
vanWinterswijk PJTS Whitehouse SL Timperley AJ Hubble MJW Howell JR Wilson MJ

Aims. We report the incidence of radiolucent lines (RLLs) using two flanged acetabular components at total hip arthroplasty (THA) and the effect of the Rim Cutter. Patients and Methods. We performed a retrospective review of 300 hips in 292 patients who underwent primary cemented THA. A contemporary flanged acetabular component was used with (group 1) and without (group 2) the use of the Rim Cutter and the Rimfit acetabular component was used with the Rim Cutter (group 3). RLLs and clinical outcomes were evaluated immediately post-operatively and at five years post-operatively. Results. There was no significant difference in the incidence of RLLs on the immediate post-operative radiographs (p = 0.241) or at five years post-operatively (p = 0.463). RLLs were seen on the immediate post-operative radiograph in 2% of hips in group 1, in 5% in group 2 and in 7% in group 3. Five years post-operatively, there were RLLs in 42% of hips in group 1, 41% in group 2 and in 49% in group 3. In the vast majority of hips, in each group, the RLL was present in DeLee and Charnley zone 1 only (86%, 83%, 67% respectively). Oxford and Harris Hip scores improved significantly in all groups. There was no significant difference in these scores or in the change in scores between the groups, with follow-up. Conclusion. Despite the Rim Cutter showing promising results in early laboratory and clinical studies, this analysis of the radiological and clinical outcome five years post-operatively does not show any advantage over and above modern cementing techniques in combination with a well performing cemented acetabular component. For this reason, we no longer use the Rim Cutter in routine primary THA. . Cite this article: Bone Joint J 2017;99-B:1450–7


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1623 - 1627
1 Dec 2015
Mounsey EJ Williams DH Howell JR Hubble MJ

Revision of a cemented hemiarthroplasty of the hip may be a hazardous procedure with high rates of intra-operative complications. Removing well-fixed cement is time consuming and risks damaging already weak bone or perforating the femoral shaft. The cement-in-cement method avoids removal of intact cement and has shown good results when used for revision total hip arthroplasty (THA). The use of this technique for the revision of a hemiarthroplasty to THA has not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients) were revised to a THA using an Exeter stem and the cement-in-cement technique. There were four men and 24 women; their mean age was 80 years (35 to 93). Clinical and radiographic data, as well as operative notes, were collected prospectively and no patient was lost to follow-up. Four patients died within two years of surgery. The mean follow up of the remainder was 70 months (25 to 124). Intra-operatively there was one proximal perforation, one crack of the femoral calcar and one acetabular fracture. No femoral components have required subsequent revision for aseptic loosening or are radiologically loose. . Four patients with late complications (14%) have since undergone surgery (two for a peri-prosthetic fracture, and one each for deep infection and recurrent dislocation) resulting in an overall major rate of complication of 35.7%. The cement-in-cement technique provides reliable femoral fixation in this elderly population and may reduce operating time and rates of complication. Cite this article: Bone Joint J 2015;97-B:1623–7


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 577 - 582
1 May 2009
Duncan WW Hubble MJW Howell JR Whitehouse SL Timperley AJ Gie GA

The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility.

We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening.

A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1482 - 1486
1 Nov 2012
Brogan K Charity J Sheeraz A Whitehouse SL Timperley AJ Howell JR Hubble MJW

The technique of femoral cement-in-cement revision is well established, but there are no previous series reporting its use on the acetabular side at the time of revision total hip replacement. We describe the technique and report the outcome of 60 consecutive acetabular cement-in-cement revisions in 59 patients at a mean follow-up of 8.5 years (5 to 12). All had a radiologically and clinically well-fixed acetabular cement mantle at the time of revision. During the follow-up 29 patients died, but no hips were lost to follow-up. The two most common indications for acetabular revision were recurrent dislocation (46, 77%) and to complement femoral revision (12, 20%).

Of the 60 hips, there were two cases of aseptic loosening of the acetabular component (3.3%) requiring re-revision. No other hip was clinically or radiologically loose (96.7%) at the latest follow-up. One hip was re-revised for infection, four for recurrent dislocation and one for disarticulation of a constrained component. At five years the Kaplan-Meier survival rate was 100% for aseptic loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause as the endpoint.

These results support the use of cement-in-cement revision on the acetabular side in appropriate cases. Theoretical advantages include preservation of bone stock, reduced operating time, reduced risk of complications and durable fixation.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 212 - 220
1 Feb 2022
Fishley WG Selvaratnam V Whitehouse SL Kassam AM Petheram TG

Aims. Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique. Methods. We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases. Results. Seven patients (7.9%) required further revision for infection. Ten patients died of causes unrelated to their infection before their two-year review (mean 5.9 months; 0.9 to 18.6). One patient was lost to follow-up at five months after surgery, and two patients died of causes unrelated to their hip shortly after their two-year review was due without attending. Of the remaining patients, 69 remained infection-free at final review. Radiological review confirms the mechanical success of the procedure as previously described in aseptic revision, and postoperative Oxford Hip Scores suggest satisfactory functional outcomes. Conclusion. In conclusion, we found that retaining a well-fixed femoral cement mantle in the presence of infection and undertaking a cement-in-cement revision was successful in 82 of the patients (92.1%) in our series of 89, both in terms of eradication of infection and component fixation. These results are comparable to other more invasive techniques and offer significant potential benefits to the patient. Cite this article: Bone Joint J 2022;104-B(2):212–220


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1002 - 1009
1 Aug 2018
Westerman RW Whitehouse SL Hubble MJW Timperley AJ Howell JR Wilson MJ

Aims. The aim of this study was to report the initial results of the Exeter V40 stem, which became available in 2000. Patients and Methods. A total of 540 total hip arthroplasties (THAs) were performed in our unit using this stem between December 2000 and May 2002. Our routine protocol is to review patients postoperatively and at one, five, and ten years following surgery. Results. A total of 145 patients (26.9%) died before ten years and of the remaining 395 stems, 374 (94.7%) remain in situ. A total of 21 well-fixed stems (5.3%) were revised. Ten were exchanged using a cement-in-cement technique to facilitate acetabular revision. Three were revised for infection, one for instability, one for fracture of the stem, and six following a periprosthetic fracture. An additional 16 acetabular components (4.1%) were revised; five for aseptic loosening and 11 for instability. There were no revisions for aseptic loosening of the stem, and no evidence of aseptic loosening in any hip. The fate of every stem is known and all patients remain under review. Survivorship, with revision of the stem for aseptic loosening as the endpoint, was 100%. At 13.5 years, the Kaplan–Meier survival rate for all-cause revision of the stem was 96.8% (95% confidence interval (CI) 94.8 to 98.8) and all-cause revision (including acetabular revision, infection, and instability) was 91.2% (95% CI 88.3 to 94.1). Conclusion. No stem was revised for aseptic loosening in this series. The contemporary Exeter V40 stem continues to perform well, and survival has remained comparable with that of the Exeter Universal stem. Cite this article: Bone Joint J 2018;100-B:1002–9


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 27 - 32
1 Apr 2017
Cnudde PHJ Kärrholm J Rolfson O Timperley AJ Mohaddes M

Aims. Compared with primary total hip arthroplasty (THA), revision surgery can be challenging. The cement-in-cement femoral revision technique involves removing a femoral component from a well-fixed femoral cement mantle and cementing a new stem into the original mantle. This technique is widely used and when carried out for the correct indications, is fast, relatively inexpensive and carries a reduced short-term risk for the patient compared with the alternative of removing well-fixed cement. We report the outcomes of this procedure when two commonly used femoral stems are used. Patients and Methods. We identified 1179 cement-in-cement stem revisions involving an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty Register (SHAR) between January 1999 and December 2015. Kaplan-Meier survival analysis was performed. Results. Survivorship is reported up to six years and was better in the Exeter group (91% standard deviation (. sd). 2.8% versus 85% . sd. 5.0%) (p = 0.02). There was, however, no significant difference in the survival of the stem and risk of re-revision for any reason (p = 0.58) and for aseptic loosening (p = 0.97), between revisions in which the Exeter stem (94% . sd. 2.2%; 98% . sd. 1.6%) was used compared with those in which the Lubinus stem (95% . sd. 3.2%; 98% . sd.  2.2%) was used. The database did not allow identification of whether a further revision was indicated for loosening of the acetabular or femoral component or both. Conclusion. The cement-in-cement technique for revision of the femoral component gave promising results using both designs of stem, six years post-operatively. Cite this article: Bone Joint J 2017;99-B(4 Supple B):27–32


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 19 - 25
1 Jan 2006
Scheerlinck T de Mey J Deklerck R Noble PC

Using a modern cementing technique, we implanted 22 stereolithographic polymeric replicas of the Charnley-Kerboul stem in 11 pairs of human cadaver femora. On one side, the replicas were cemented line-to-line with the largest broach. On the other, one-size undersized replicas were used (radial difference, 0.89 mm . sd. 0.13). CT analysis showed that the line-to-line stems without distal centralisers were at least as well aligned and centered as undersized stems with a centraliser, but were surrounded by less cement and presented more areas of thin (< 2 mm) or deficient (< 1 mm) cement. These areas were located predominantly at the corners and in the middle and distal thirds of the stem. Nevertheless, in line-to-line stems, penetration of cement into cancellous bone resulted in a mean thickness of cement of 3.1 mm (. sd. 0.6) and only 6.2% of deficient and 26.4% of thin cement. In over 90% of these areas, the cement was directly supported by cortical bone or cortical bone with less than 1 mm of cancellous bone interposed. When Charnley-Kerboul stems are cemented line-to-line, good clinical results are observed because cement-deficient areas are limited and are frequently supported by cortical bone


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives. Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds. Methods. A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens. Results. The mean shear strength for OSCAR-prepared specimens (33.6 MPa) was significantly lower than for the control (46.3 MPa) and burr (45.8 MPa) groups (p < 0.001; one-way analysis of variance (ANOVA) with Tukey’s post hoc analysis). There was no significant difference in shear strengths between control and burr groups (p = 0.57). Scanning electron microscopy of OSCAR specimens revealed evidence of porosity undiscovered in previous studies. Conclusion. Results show that the cement removal technique impacts on final cement-in-cement bonds. This in vitro study demonstrates significantly weaker bonds when using OSCAR prior to recementation into an old cement mantle compared with cement prepared with a burr or no treatment. This infers that care must be taken in surgical decision-making regarding cement removal techniques used during cement-in-cement revision arthroplasty, suggesting that the risks and benefits of ultrasonic cement removal need consideration. Cite this article: A. Liddle, M. Webb, N. Clement, S. Green, J. Liddle, M. German, J. Holland. Ultrasonic cement removal in cement-in-cement revision total hip arthroplasty: What is the effect on the final cement-in-cement bond? Bone Joint Res 2019;8:246–252. DOI: 10.1302/2046-3758.86.BJR-2018-0313.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1515 - 1521
1 Nov 2010
Clauss M Ilchmann T Zimmermann P Ochsner PE

The aim of this study was to obtain detailed long-term data on the cement-bone interface in patients with cemented stems, implanted using the constrained fixation technique. A total of eight stems were removed together with adjacent bone during post-mortem examinations of patients with well-functioning prostheses. Specimens were cut at four defined levels, contact radiographs were obtained for each level, and slices were prepared for histological analysis. Clinical data, clinical radiographs, contact radiographs and histological samples were examined for signs of loosening and remodelling. The mean radiological follow-up was 9.6 years and all stems were well-fixed, based on clinical and radiological criteria. Contact radiographs revealed an incomplete cement mantle but a complete filling of the medullary canal for all implants. Various amounts of polyethylene particles were evident at the cement-bone interface of seven stems, with no accompanying inflammatory reaction. Cortical atrophy and the formation of an ‘inner cortex’ were confirmed in six of eight stems by contact radiographs and histology, but were only visible on two clinical radiographs. Our results confirm that a complete cement mantle is not essential for the survival of Müller straight stems into the mid term, and support our hypothesis that no benefit to long-term survival can be expected from modern cementing techniques


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 928 - 933
1 Jul 2005
Berli BJ Schäfer D Morscher EW

The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant in situ. Nine could only be interviewed by telephone. We included 85 patients with 88 hips in the clinical and radiological follow-up examinations. None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed. The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy IW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims

Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components.

Methods

From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.


Bone & Joint Open
Vol. 2, Issue 5 | Pages 278 - 292
3 May 2021
Miyamoto S Iida S Suzuki C Nakatani T Kawarai Y Nakamura J Orita S Ohtori S

Aims

The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA.

Methods

A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 71 - 78
1 Jan 2021
Maggs JL Swanton E Whitehouse SL Howell JR Timperley AJ Hubble MJW Wilson MJ

Aims

Periprosthetic fractures (PPFs) around cemented taper-slip femoral prostheses often result in a femoral component that is loose at the prosthesis-cement interface, but where the cement-bone interface remains well-fixed and bone stock is good. We aim to understand how best to classify and manage these fractures by using a modification of the Vancouver classification.

Methods

We reviewed 87 PPFs. Each was a first episode of fracture around a cemented femoral component, where surgical management consisted of revision surgery. Data regarding initial injury, intraoperative findings, and management were prospectively collected. Patient records and serial radiographs were reviewed to determine fracture classification, whether the bone cement was well fixed (B2W) or loose (B2L), and time to fracture union following treatment.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 199 - 203
1 Feb 2017
Sandiford NA Jameson SS Wilson MJ Hubble MJW Timperley AJ Howell JR

Aims

We present the clinical and radiological results at a minimum follow-up of five years for patients who have undergone multiple cement-in-cement revisions of their femoral component at revision total hip arthroplasty (THA).

Patients and Methods

We reviewed the outcome on a consecutive series of 24 patients (10 men, 14 women) (51 procedures) who underwent more than one cement-in-cement revision of the same femoral component. The mean age of the patients was 67.5 years (36 to 92) at final follow-up.

Function was assessed using the original Harris hip score (HHS), Oxford Hip Score (OHS) and the Merle D’Aubigné Postel score (MDP).


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1441 - 1449
1 Nov 2016
Petheram TG Whitehouse SL Kazi HA Hubble MJW Timperley AJ Wilson MJ Howell JR

Aims

We present a minimum 20-year follow-up study of 382 cemented Exeter Universal total hip arthroplasties (350 patients) operated on at a mean age of 66.3 years (17 to 94).

Patients and Methods

All patients received the same design of femoral component, regardless of the original diagnosis. Previous surgery had been undertaken for 33 hips (8.6%). During the study period 218 patients with 236 hips (62%) died, 42 hips (11%) were revised and 110 hips (29%) in 96 patients were available for review. The acetabular components were varied and some designs are now obsolete, however they were all cemented.