The purpose of this study was to determine whether
it would be feasible to use oblique lumbar interbody fixation for
patients with degenerative lumbar disease who required a fusion
but did not have a spondylolisthesis. A series of CT digital images from 60 patients with abdominal
disease were reconstructed in three dimensions (3D) using Mimics
v10.01: a digital cylinder was superimposed on the reconstructed
image to simulate the position of an interbody screw. The optimal
entry point of the screw and measurements of its trajectory were
recorded. Next, 26 cadaveric specimens were subjected to oblique
lumbar interbody fixation on the basis of the measurements derived
from the imaging studies. These were then compared with measurements
derived directly from the cadaveric vertebrae. Our study suggested that it is easy to insert the screws for
L1/2, L2/3 and L3/4 fixation: there was no significant difference
in measurements between those of the 3-D digital images and the
cadaveric specimens. For L4/5 fixation, part of L5 inferior articular
process had to be removed to achieve the optimal trajectory of the
screw. For L5/S1 fixation, the screw heads were blocked by iliac
bone: consequently, the interior oblique angle of the cadaveric specimens
was less than that seen in the 3D digital images. We suggest that CT scans should be carried out pre-operatively
if this procedure is to be adopted in clinical practice. This will
assist in determining the feasibility of the procedure and will
provide accurate information to assist introduction of the screws. Cite this article:
In patients with osteoporosis there is always
a strong possibility that pedicle screws will loosen. This makes
it difficult to select the appropriate osteoporotic patient for
a spinal fusion. The purpose of this study was to determine the
correlation between bone mineral density (BMD) and the magnitude
of torque required to insert a pedicle screw. To accomplish this,
181 patients with degenerative disease of the lumbar spine were
studied prospectively. Each underwent dual-energy x-ray absorptiometry
(DEXA) and intra-operative measurement of the torque required to
insert each pedicle screw. The levels of torque generated in patients
with osteoporosis and osteopenia were significantly lower than those
achieved in normal patients. Positive correlations were observed between
BMD and T-value at the instrumented lumbar vertebrae, mean BMD and
mean T-value of the lumbar vertebrae, and mean BMD and mean T-value
of the proximal femur. The predictive torque (Nm) generated during pedicle
screw insertion was [-0.127 + 1.62 × (BMD at the corresponding lumbar
vertebrae)], as measured by linear regression analysis. The positive
correlation between BMD and the maximum torque required to insert
a pedicle screw suggests that pre-operative assessment of BMD may
be useful in determining the ultimate strength of fixation of a
device, as well as the number of levels that need to be fixed with
pedicle screws in patients who are suspected of having osteoporosis.