Aims. Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented
Aims. This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. Patients and Methods. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation. Results. Of the 964 patients enrolled, 482 died or were lost to follow-up
(50%). No significant differences were noted in EQ-5D between groups,
with a mean difference at four months of 0.037 in favour of the
Exeter/Unitrax implant (95% confidence interval (CI) 0.014 to 0.087,
p = 0.156), rising to 0.045 (95% CI 0.007 to 0.098, p = 0.09) when
patients who died were excluded. The minimum clinically important
difference for EQ-5D-5L used in this study is 0.08, therefore any
benefit between implants is unlikely to be noticeable to the patient.
There was no difference in mortality or mobility score. Conclusion. Allowing for the high rate of loss to follow-up, the use of the
traditional
Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an ODEP rating despite over 50 years of clinical use, likely due to the paucity of implant survival data. Nationally, adherence to these guidelines is varied as there is debate as to which prosthesis optimises patient outcomes. This study design is a multi-centre, multi-surgeon, parallel, two arm, standard-of-care pragmatic randomised controlled trial. It will be embedded within the WHiTE Comprehensive Cohort Study (ISRCTN63982700). The main analysis is a two-way equivalence comparison between Hemi-Thompson and Hemi-Exeter polished taper with Unitrax head. Secondary outcomes will include radiological leg length discrepancy measured as per Bidwai and Willett, mortality, re-operation rate and indication for re-operation, length of index hospital stay and revision at four months. This study will be supplemented by the NHFD (National Hip Fracture Database) dataset.Background
Design
We undertook a prospective randomised controlled trial involving 400 patients with a displaced intracapsular fracture of the hip to determine whether there was any difference in outcome between treatment with a cemented
We performed a prospective, randomised trial comparing three treatments for displaced intracapsular fractures of the hip in 280 patients aged 65 to 79 years. The mean patient survival was significantly higher in the group undergoing reduction and internal fixation (79 months) compared with that with a cemented
Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses. A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented polished tapered stem hemiarthroplasty or an uncemented Furlong hydroxyapatite-coated hemiarthroplasty. Follow-up was conducted by a nurse blinded to the implant at set intervals for up to one year from surgery.Aims
Methods
This study describes and compares the operative management and outcomes in a consecutive case series of patients with dislocated hemiarthroplasties of the hip, and compares outcomes with those of patients not sustaining a dislocation. Of 3326 consecutive patients treated with hemiarthroplasty for fractured neck of femur, 46 (1.4%) sustained dislocations. Of the 46 dislocations, there were 37 female patients (80.4%) and nine male patients (19.6%) with a mean age of 83.8 years (66 to 100). Operative intervention for each, and subsequent dislocations, were recorded. The following outcome measures were recorded: dislocation; mortality up to one-year post-injury; additional surgery; residential status; mobility; and pain score at one year.Aims
Patients and Methods
Periprosthetic femoral fractures (PFF) following total hip arthroplasty
(THA) are devastating complications that are associated with functional
limitations and increased overall mortality. Although cementless
implants have been associated with an increased risk of PFF, the
precise contribution of implant geometry and design on the risk
of both intra-operative and post-operative PFF remains poorly investigated.
A systematic review was performed to aggregate all of the PFF literature
with specific attention to the femoral implant used. A systematic search strategy of several journal databases and
recent proceedings from the American Academy of Orthopaedic Surgeons
was performed. Clinical articles were included for analysis if sufficient
implant description was provided. All articles were reviewed by
two reviewers. A review of fundamental investigations of implant
load-to-failure was performed, with the intent of identifying similar
conclusions from the clinical and fundamental literature.Aims
Patients and Methods
National Institute of Clinical Excellence guidelines
state that cemented stems with an Orthopaedic Data Evaluation Panel
(ODEP) rating of >
3B should be used for hemiarthroplasty when treating
an intracapsular fracture of the femoral neck. These recommendations
are based on studies in which most, if not all stems, did not hold
such a rating. This case-control study compared the outcome of hemiarthroplasty
using a cemented (Exeter) or uncemented (Corail) femoral stem. These
are the two prostheses most commonly used in hip arthroplasty in
the UK. Data were obtained from two centres; most patients had undergone
hemiarthroplasty using a cemented Exeter stem (n = 292/412). Patients
were matched for all factors that have been shown to influence mortality
after an intracapsular fracture of the neck of the femur. Outcome
measures included: complications, re-operations and mortality rates
at two, seven, 30 and 365 days post-operatively. Comparable outcomes
for the two stems were seen. There were more intra-operative complications in the uncemented
group (13% This study therefore supports the use of both cemented and uncemented
stems of proven design, with an ODEP rating of 10A, in patients
with an intracapsular fracture of the neck of the femur. Cite this article:
We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p <
0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p <
0.01). Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.
Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck. We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years. Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D. Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years. The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.
Prospective data on hip fracture from 3686 patients at a United Kingdom teaching hospital were analysed to investigate the risk factors, financial costs and outcomes associated with deep or superficial wound infections after hip fracture surgery. In 1.2% (41) of patients a deep wound infection developed, and 1.1% (39) had a superficial wound infection. A total of 57 of 80 infections (71.3%) were due to No statistically significant pre-operative risk factors were detected. Length of stay, cost of treatment and pre-discharge mortality all significantly increased with deep wound infection. The one-year mortality was 30%, and this increased to 50% in those who developed an infection (p <
0.001). A deep infection resulted in doubled operative costs, tripled investigation costs and quadrupled ward costs. MRSA infection increased costs, length of stay, and pre-discharge mortality compared with non-MRSA infection.
We reviewed 59 bone graft substitutes marketed
by 17 companies currently available for implantation in the United Kingdom,
with the aim of assessing the peer-reviewed literature to facilitate
informed decision-making regarding their use in clinical practice.
After critical analysis of the literature, only 22 products (37%)
had any clinical data. Norian SRS (Synthes), Vitoss (Orthovita),
Cortoss (Orthovita) and Alpha-BSM (Etex) had Level I evidence. We question
the need for so many different products, especially with limited
published clinical evidence for their efficacy, and conclude that
there is a considerable need for further prospective randomised
trials to facilitate informed decision-making with regard to the
use of current and future bone graft substitutes in clinical practice. Cite this article: