Aims. The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde
Aims. The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device. Methods. This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral
Lengthening of the humerus is now an established
technique. We compared the complications of humeral lengthening
with those of
As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30).Aims
Methods
Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.Aims
Methods
Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. The medical records and radiographs of all patients who underwent
CHAOS for femoral deformity at our institution between 2005 and
2011 were retrospectively reviewed. Records were available for all
55 consecutive procedures undertaken in 49 patients with a mean
age of 35.6 years (10.9 to 75.3) at the time of surgery.Aims
Patients and Methods
Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest. The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion.
External fixation is widely used in orthopaedic
and trauma surgery. Infections around pin or wire sites, which are usually
localised, non-invasive, and are easily managed, are common. Occasionally,
more serious invasive complications such as necrotising fasciitis
(NF) and toxic shock syndrome (TSS) may occur. We retrospectively reviewed all patients who underwent external
fixation between 1997 and 2012 in our limb lengthening and reconstruction
programme. A total of eight patients (seven female and one male)
with a mean age of 20 years (5 to 45) in which pin/wire track infections
became limb- or life-threatening were identified. Of these, four
were due to TSS and four to NF. Their management is described. A
satisfactory outcome was obtained with early diagnosis and aggressive
medical and surgical treatment. Clinicians caring for patients who have external fixation and
in whom infection has developed should be aware of the possibility
of these more serious complications. Early diagnosis and aggressive
treatment are required in order to obtain a satisfactory outcome. Cite this article:
Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications. Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult.
We present a retrospective review of a single-surgeon series of 30 consecutive lengthenings in 27 patients with congenital short femur using the Ilizarov technique performed between 1994 and 2005. The mean increase in length was 5.8 cm/18.65% (3.3 to 10.4, 9.7% to 48.8%), with a mean time in the frame of 223 days (75 to 363). By changing from a distal to a proximal osteotomy for lengthening, the mean range of knee movement was significantly increased from 98.1° to 124.2° (p = 0.041) and there was a trend towards a reduced requirement for quadricepsplasty, although this was not statistically significant (p = 0.07). The overall incidence of regenerate deformation or fracture requiring open reduction and internal fixation was similar in the distal and proximal osteotomy groups (56.7% and 53.8%, respectively). However, in the proximal osteotomy group, pre-placement of a Rush nail reduced this rate from 100% without a nail to 0% with a nail (p <
0.001). When comparing a distal osteotomy with a proximal one over a Rush nail for lengthening, there was a significant decrease in fracture rate from 58.8% to 0% (p = 0.043). We recommend that in this group of patients lengthening of the femur with an Ilizarov construct be carried out through a proximal osteotomy over a Rush nail. Lengthening should also be limited to a maximum of 6 cm during one treatment, or 20% of the original length of the femur, in order to reduce the risk of complications.