The aims of this study were to assess the efficacy
of a newly designed radiological technique (the radial groove view)
for the detection of protrusion of screws in the groove for the
extensor pollicis longus tendon (EPL) during plating of distal radial
fractures. We also aimed to determine the optimum position of the
forearm to obtain this view. We initially analysed the anatomy of
the EPL groove by performing
Damage to the cartilage of the distal radioulnar
joint frequently leads to pain and limitation of movement, therefore repair
of this joint cartilage would be highly desirable. The purpose of
this study was to investigate the fixation of scaffold in cartilage
defects of this joint as part of matrix-assisted regenerative autologous
cartilage techniques. Two techniques of fixation of collagen scaffolds,
one involving fibrin glue alone and one with fibrin glue and sutures, were
compared in artificially created cartilage defects of the distal
radioulnar joint in a human cadaver. After being subjected to continuous
passive rotation, the methods of fixation were evaluated for cover
of the defect and pull out force. No statistically significant differences were found between the
two techniques for either cover of the defect or integrity of the
scaffold. However, a significantly increased mean pull out force
was found for the combined procedure, 0.665 N (0.150 to 1.160) This suggests that although successful fixation of a collagen
type I/III scaffold in a distal radioulnar joint cartilage defect
is feasible with both forms of fixation, fixation with glue and
sutures is preferable. Cite this article:
Penetration of the dorsal screw when treating
distal radius fractures with volar locking plates is an avoidable complication
that causes lesions of the extensor tendon in between 2% and 6%
of patients. We examined axial fluoroscopic views of the distal
end of the radius to observe small amounts of dorsal screw penetration,
and determined the ideal angle of inclination of the x-ray beam
to the forearm when making this radiological view. Six volar locking plates were inserted at the wrists of cadavers.
The actual screw length was measured under direct vision through
a dorsal approach to the distal radius. Axial radiographs were performed
for different angles of inclination of the forearm at the elbow. Comparing axial radiological measurements and real screw length,
a statistically significant correlation could be demonstrated at
an angle of inclination between 5° and 20°. The ideal angle of inclination
required to minimise the risk of implanting over-long screws in
a dorsal horizon radiological view is 15°. Cite this article: