Objectives. The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of
Objectives. Previous studies have suggested that
The peri-prosthetic tissue response to wear debris
is complex and influenced by various factors including the size, area
and number of particles. We hypothesised that the ‘biologically
active area’ of all metal wear particles may predict the type of
peri-prosthetic tissue response. . Peri-prosthetic tissue was sampled from 21 patients undergoing
revision of a small diameter
Aims. This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by
Aims. Many case reports and small studies have suggested that cobalt
ions are a potential cause of cardiac complications, specifically
cardiomyopathy, after
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul
Aims . The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and
Osteolysis is due to particulate wear debris and is responsible for the long-term failure of total hip replacements. It has stimulated the development of alternative joint surfaces such as
We have reviewed 42 patients who had revision of
Aims. There are many guidelines that help direct the management of
patients with
Bioengineering reasons for increased wear and failure of
We compared the incidence of pseudotumours after
large head
Aims. The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. . Methods. We analysed
In a double-blinded randomised controlled trial,
83 patients with primary osteoarthritis of the hip received either
a ceramic-on-metal (CoM) or
Patients from a randomised trial on resurfacing
hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24
to 65) comparing a conventional 28 mm
Aims. We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following
We report the outcome of total hip replacement in 29 failed
We evaluated the concentrations of chromium and cobalt ions in blood after
Objectives. The aims of this piece of work were to: 1) record the background
concentrations of blood chromium (Cr) and cobalt (Co) concentrations
in a large group of subjects; 2) to compare blood/serum Cr and Co
concentrations with retrieved
Arthroplasty registries are important for the
surveillance of joint replacements and the evaluation of outcome. Independent
validation of registry data ensures high quality. The ability for
orthopaedic implant retrieval centres to validate registry data
is not known. We analysed data from the National Joint Registry
for England, Wales and Northern Ireland (NJR) for primary metal-on-metal
hip arthroplasties performed between 2003 and 2013. Records were
linked to the London Implant Retrieval Centre (RC) for validation.
A total of 67 045 procedures on the NJR and 782 revised pairs of
components from the RC were included. We were able to link 476 procedures
(60.9%) recorded with the RC to the NJR successfully. However, 306
procedures (39.1%) could not be linked. The outcome recorded by the
NJR (as either revised, unrevised or death) for a primary procedure
was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval
linkage and correct assignment of outcome code improved over time.
The rates of error for component reference numbers on the NJR were
as follows: femoral head category number 14/229 (5.0%); femoral head
batch number 13/232 (5.3%); acetabular component category number
2/293 (0.7%) and acetabular component batch number 24/347 (6.5%). Registry-retrieval linkage provided a novel means for the validation
of data, particularly for component fields. This study suggests
that NJR reports may underestimate rates of revision for many types
of metal-on-metal hip replacement. This is topical given the increasing
scope for NJR data. We recommend a system for continuous independent
evaluation of the quality and validity of NJR data. Cite this article:
The long term biological effects of wear products
following total hip arthroplasty (THA) are unclear. However, the indications
for THA are expanding, with increasingly younger patients undergoing
the procedure. This prospective, randomised study compared two groups of patients
undergoing THA after being randomised to receive one of two different
bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal
(MoM) n = 23. We investigated the relationship between three variables:
bearing surface (MoP Our results demonstrated significantly higher mean cobalt and
chromium (Co and Cr) blood levels in the MoM group at all follow-up
points following surgery (p <
0.01), but there were no significant
differences in the chromosomal aberration indices between MoM and
MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p
= 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34)
for translocations, breaks, loss and gain of chromosomes respectively.
Regression analysis showed a strong linear relationship between
Cr levels and the total chromosomal aberration indices in the MoM
group (R2 = 0.90016), but this was not as strong for
Co (R2 = 0.68991). In the MoP group, the analysis revealed
a poor relationship between Cr levels and the total chromosomal
aberration indices (R2 = 0.23908) but a slightly stronger
relationship for Co (R2 = 0.64292). Across both groups,
Spearman’s correlation detected no overall association between Co and Cr
levels and each of the studied chromosomal aberrations. There remains
no clear indication which THA bearing couple is the most biocompatible,
especially in young active patients. While THA continues to be very
successful at alleviating pain and restoring function, the long-term
biological implications of the procedure still require further scrutiny. Cite this article:
We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine. We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of >
5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups. We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT.
We investigated the changes seen on serial metal
artefact reduction magnetic resonance imaging scans (MARS-MRI) of
metal-on-metal total hip arthroplasties (MoM THAs). In total 155
THAs, in 35 male and 100 female patients (mean age 70.4 years, 42
to 91), underwent at least two MRI scans at a mean interval of 14.6
months (2.6 to 57.1), at a mean of 48.2 months (3.5 to 93.3) after
primary hip surgery. Scans were graded using a modification of the
Oxford classification. Progression of disease was defined as an
increase in grade or a minimum 10% increase in fluid lesion volume
at second scan. A total of 16 hips (30%) initially classified as
‘normal’ developed an abnormality on the second scan. Of those with
‘isolated trochanteric fluid’ 9 (47%) underwent disease progression,
as did 7 (58%) of ‘effusions’. A total of 54 (77%) of hips initially
classified as showing adverse reactions to metal debris (ARMD) progressed,
with higher rates of progression in higher grades. Disease progression
was associated with high blood cobalt levels or an irregular pseudocapsule
lining at the initial scan. There was no association with changes
in functional scores. Adverse reactions to metal debris in MoM THAs
may not be as benign as previous reports have suggested. Close radiological
follow-up is recommended, particularly in high-risk groups. Cite this article:
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component. Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001). The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients). Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 ( We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.
The results of conventional hip replacement in young patients with osteoarthritis have not been encouraging even with improvements in the techniques of fixation and in the bearing surfaces. Modern metal-on-metal hip resurfacing was introduced as a less invasive method of joint reconstruction for this particular group. This is a series of 446 hip resurfacings (384 patients) performed by one of the authors (DJWM) using cemented femoral components and hydroxyapatite-coated uncemented acetabular components with a maximum follow-up of 8.2 years (mean 3.3). Their survival rate, Oxford hip scores and activity levels are reviewed. Six patients died due to unrelated causes. There was one revision (0.02%) out of 440 hips. The mean Oxford score of the surviving 439 hips is 13.5. None of the patients were told to change their activities at work or leisure; 31% of the men with unilateral resurfacings and 28% with bilateral resurfacings were involved in jobs that they considered heavy or moderately heavy; 92% of men with unilateral hip resurfacings and 87% of the whole group participate in leisure-time sporting activity. The extremely low rate of failure in spite of the resumption of high level occupational and leisure activities provides early evidence of the suitability of this procedure for young and active patients with arthritis.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
Aims. In
Aims. The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter
Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around
Aims. This study aims to assess the relationship between history of pseudotumour formation secondary to
Aims. Large-diameter
Aims. To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with
Aims. The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. Methods. A total of 81
Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings. We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring
Aims. Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of
Aims. The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. Methods. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083
Persistent groin pain after seemingly successful
total hip replacement (THR) appears to have become more common.
Recent studies have indicated a high incidence after metal-on-polyethylene
and
Objectives. Modern
Aims. The present study investigated the five-year interval changes in pseudotumours and measured serum metal ions at long-term follow-up of a previous report of 28 mm diameter
Aims. In previous studies, we identified multiple factors influencing the survivorship of hip resurfacing arthroplasties (HRAs), such as initial anatomical conditions and surgical technique. In addition, the University of California, Los Angeles (UCLA) activity score presents a ceiling effect, so a better quantification of activity is important to determine which activities may be advisable or detrimental to the recovered patient. We aimed to determine the effect of specific groups of sporting activities on the survivorship free of aseptic failure of a large series of HRA. Patients and Methods. A total of 661 patients (806 hips) representing 77% of a consecutive series of patients treated with
Aims. There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of
Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man between April 2003 and March 2017. All patients aged ≤ 20 years at the time of THA were included and the primary outcome was revision surgery. Descriptive statistics were used to summarize the data and Kaplan–Meier estimates calculated for the cumulative implant survival. Results. A total of 769 THAs were performed in 703 patients. The median follow-up was 5.1 years (interquartile range (IQR) 2.6 to 7.8). Eight patients died and 35 THAs were revised. The use of
Modern metal-on-metal bearings produce less wear debris and osteolysis, but have the potential adverse effect of release of ions. Improved ceramic-on-ceramic bearings have the lowest wear of all, but the corrosion process has not been analysed. Our aim was to measure the serum ion release (ng/ml) in 23 patients having stable hip prostheses with a ceramic-on-ceramic coupling (group A) and to compare it with the release in 42 patients with a
We report 17 patients (20 hips) in whom