We undertook a radiological analysis of 186 standard Souter implants to determine survivorship and to analyse the pattern of failure in those needing revision. The implants had been inserted as a primary procedure in patients with rheumatoid arthritis of the elbow at our hospital over the last 12 years. Taking revision as an endpoint, the survivorship after 12 years was 87%. If, however, revision and loosening, defined as the Hindex value equivalent to demarcation of 1 mm around the whole implant, are also included, the survivorship falls to 80%. Of the 24 implants revised, 18 (75%) were for problems with the
We used an inverted shoulder arthroplasty in 43 consecutive patients with a mean age of 78 years (65 to 97) who had sustained a three- or four-part fracture of the upper humerus. All except two were reviewed with a mean follow-up of 22 months (6 to 58). The clinical outcome was satisfactory with a mean active anterior elevation of 97° (35° to 160°) and a mean active external rotation in abduction of 30° (0° to 80°). The mean Constant and the mean modified Constant scores were respectively 44 (16 to 69) and 66% (25% to 97%). Complications included three patients with reflex sympathetic dystrophy, five with neurological complications, most of which resolved, and one with an anterior dislocation. Radiography showed peri-prosthetic calcification in 36 patients (90%), displacement of the tuberosities in 19 (53%) and a scapular notch in ten (25%). Compared with conventional hemiarthroplasty, satisfactory mobility was obtained despite frequent migration of the tuberosities. However, long-term results are required before reverse shoulder arthroplasty can be recommended as a routine procedure in complex fractures of the upper humerus in the elderly.