In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any
Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined. Results. There were 158,677 individuals (42.4%) with an implant containing CoCr. There were 47,963 deaths, 27,332 heart outcomes, 35,720 cancers, and 22,025 neurodegenerative disorders. There was no evidence of an association between patients with CoCr implants and higher rates of any of the outcomes. Conclusion. CoCr-containing THAs did not have an increased risk of all-cause mortality, or clinically meaningful heart outcomes, cancer, or neurodegenerative disorders into the second decade post-implantation. Our findings will help reassure clinicians and the increasing number of patients receiving primary THA worldwide that the use of CoCr-containing implants is not associated with significant
Infection of a total hip replacement (THR) requires component removal and thorough local debridement. Usually, long-term antibiotic treatment in conjunction with a two-stage revision is required. This may take several months. One-stage revision using antibiotic-loaded cement has not gained widespread use, although the clinical and economic advantages are obvious. Allograft bone may be impregnated with high levels of antibiotics, and in revision of infected THR, act as a carrier providing a sustained high local concentration. We performed 37 one-stage revision of infected THRs, without the use of cement. There were three hips which required further revision because of recurrent infection, the remaining 34 hips (92%) stayed free from infection and stable at a mean follow-up of 4.4 years (2 to 8). No
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface. Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer. The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no
We have investigated the effect of using tranexamic
acid (TXA) during peri-acetabular osteotomy (PAO) on peri-operative
blood loss and blood transfusion requirements. In addition we analysed
whether the use of TXA was associated with an increased risk of
venous thromboembolism (VTE) following this procedure. A consecutive series of 96 PAOs, performed by a single surgeon,
were reviewed. A total of 48 patients received TXA and 48 did not.
The TXA group received a continuous infusion of TXA at a rate of
10 mg/kg/h. The primary outcome measure was the requirement for
blood transfusion. Secondary outcomes included total blood loss,
the decrease in the level of haemoglobin in the blood, the length
of hospital stay, and the complications of this treatment. The mean rate of transfusion was significantly lower in the TXA
group (62.5% vs 12.5%, p <
0.001). The mean blood
loss was also significantly reduced in the TXA group (1.9 L (standard
deviation (. sd. ) 0.9) vs 1.5 L (. sd. 0.7),
p <
0.01). No post-operative episodes of VTE were identified
in either group. . The use of TXA reduced the blood loss and the rate of transfusion
after PAO significantly, without
Perthes’ disease (PD) is a childhood hip disorder that can affect the quality of life in adulthood due to femoral head deformity and osteoarthritis. There is very little data on how PD patients function as adults, especially from the patients’ perspective. The purpose of this study was to collect treatment history, demographic details, the University of California, Los Angeles activity score (UCLA), the 36-Item Short Form survey (SF-36) score, and the Hip disability and Osteoarthritis Outcome score (HOOS) of adults who had PD using a web-based survey method and to compare their outcomes to the outcomes from an age- and sex-matched normative population. The English REDCap-based survey was made available on a PD study group website. The survey included childhood and adult PD history, UCLA, SF-36, and HOOS. Of the 1,182 participants who completed the survey, the 921 participants who did not have a total hip arthroplasty are the focus of this study. The mean age at survey was 38 years (SD 12) and the mean duration from age at PD onset to survey participation was 30.8 years (SD 12.6).Aims
Methods
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
Metal-on-metal bearings are being increasingly used in young patients. The potential
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane. Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.Aims
Methods
To evaluate the effect of ultrasound-targeted simvastatin-loaded microbubble destruction (UTMD In vitro, OA chondrocytes were treated with ultrasound (US), US-targeted microbubble destruction (UTMD), simvastatin (SV), and UTMDAims
Methods
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Postoperative delirium (POD) and postoperative cognitive decline (POCD) are common surgical complications. In the UK, the Best Practice Tariff incentivizes the screening of delirium in patients with hip fracture. Further, a National Hip Fracture Database (NHFD) performance indicator is the reduction in the incidence of POD. To aid in its recognition, we sought to determine factors associated with POD and POCD in patients with hip fractures. We interrogated the NHFD data on patients presenting with hip fractures to our institution from 2016 to 2018. POD was determined using the 4AT score, as recommended by the NHFD and UK Department of Health. POCD was defined as a decline in Abbreviated Mental Test Score (AMTS) of two or greater. Using logistic regression, we adjusted for covariates to identify factors associated with POD and POCD.Aims
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic surgery, aiming to provide efficient, high-quality care. However, there are concerns about the safety of this practice. The purpose of this study was to examine the safety and efficacy of a model of partially overlapping surgery that we termed ‘swing room’ in the practice of primary total hip (THA) and knee arthroplasty (TKA). A retrospective review of prospectively collected data was carried out on patients who underwent primary THA and TKA between 2006 and 2017 in two academic centres. Cases were stratified as partially overlapping (swing room), in which the surgeon is in one operating room (OR) while the next patient is being prepared in another, or nonoverlapping surgery. The demographic details of the patients which were collected included operating time, length of stay (LOS), postoperative complications within six weeks of the procedure, unplanned hospital readmissions, and unplanned reoperations. Fisher's exact, Wilcoxon rank-sum tests, chi-squared tests, and logistic regression analysis were used for statistical analysis.Aims
Methods
In previous studies, we identified multiple factors influencing the survivorship of hip resurfacing arthroplasties (HRAs), such as initial anatomical conditions and surgical technique. In addition, the University of California, Los Angeles (UCLA) activity score presents a ceiling effect, so a better quantification of activity is important to determine which activities may be advisable or detrimental to the recovered patient. We aimed to determine the effect of specific groups of sporting activities on the survivorship free of aseptic failure of a large series of HRA. A total of 661 patients (806 hips) representing 77% of a consecutive series of patients treated with metal-on-metal hybrid HRA answered a survey to determine the types and amounts of sporting activities they regularly participated in. There were 462 male patients (70%) and 199 female patients (30%). Their mean age at the time of surgery was 51.9 years (14 to 78). Their mean body mass index (BMI) was 26.5 kg/m2 (16.7 to 46.5). Activities were regrouped into 17 categories based on general analogies between these activities. Scores for typical frequency and duration of the sessions were used to quantify the patients’ overall time spent engaging in sporting activities. Impact and cycle scores were computed. Multivariable models were used.Aims
Patients and Methods
Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (Aims
Patients and Methods
The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction. Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.Aims
Patients and Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
The aim of this study was to compare the rate of perioperative
complications following aseptic revision total hip arthroplasty
(THA) in patients aged ≥ 80 years with that in those aged <
80
years, and to identify risk factors for the incidence of serious
adverse events in those aged ≥ 80 years using a large validated
national database. Patients who underwent aseptic revision THA were identified in
the 2005 to 2015 National Surgical Quality Improvement Program (NSQIP)
database and stratified into two age groups: those aged <
80
years and those aged ≥ 80 years. Preoperative and procedural characteristics
were compared. Multivariate regression analysis was used to compare
the risk of postoperative complications and readmission. Risk factors
for the development of a serious adverse event in those aged ≥ 80
years were characterized.Aims
Patients and Methods
The World Health Organization (WHO) and the Centre
for Disease Control and Prevention (CDC) recently published guidelines
for the prevention of surgical site infection. The WHO guidelines,
if implemented worldwide, could have an immense impact on our practices
and those of the CDC have implications for healthcare policy in
the United States. Our aim was to review the strategies for prevention of periprosthetic
joint infection in light of these and other recent guidelines. Cite this article:
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
The acetabular labrum is a soft-tissue structure
which lines the acetabular rim of the hip joint. Its role in hip
joint biomechanics and joint health has been of particular interest
over the past decade. In normal hip joint biomechanics, the labrum
is crucial in retaining a layer of pressurised intra-articular fluid
for joint lubrication and load support/distribution. Its seal around
the femoral head is further regarded as a contributing to hip stability through
its suction effect. The labrum itself is also important in increasing
contact area thereby reducing contact stress. Given the labrum’s
role in normal hip joint biomechanics, surgical techniques for managing
labral damage are continuously evolving as our understanding of
its anatomy and function continue to progress. The current paper
aims to review the anatomy and biomechanical function of the labrum
and how they are affected by differing surgical techniques. Take home message: The acetabular labrum plays a critical role
in hip function and maintaining and restoring its function during
surgical intervention remain an essential goal. Cite this article:
The accurate reconstruction of hip anatomy and
biomechanics is thought to be important in achieveing good clinical
outcomes following total hip arthroplasty (THA). To this end some
newer hip designs have introduced further modularity into the design
of the femoral component such that neckshaft angle and anteversion,
which can be adjusted intra-operatively. The clinical effect of
this increased modularity is unknown. We have investigated the changes
in these anatomical parameters following conventional THA with a
prosthesis of predetermined neck–shaft angle and assessed the effect
of changes in the hip anatomy on clinical outcomes. In total, 44 patients (mean age 65.3 years (standard deviation
( The mean pre-operative neck–shaft angle was significantly increased
by 2.8° from 128° ( Cite this article:
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
The aim of this retrospective cohort study was
to identify any difference in femoral offset as measured on pre-operative
anteroposterior (AP) radiographs of the pelvis, AP radiographs of
the hip and corresponding CT scans in a consecutive series of 100
patients with primary end-stage osteoarthritis of the hip (43 men
and 57 women with a mean age of 61 years (45 to 74) and a mean body
mass index of 28 kg/m2 (20 to 45)). Patients were positioned according to a standardised protocol
to achieve reproducible projection and all images were calibrated.
Inter- and intra-observer reliability was evaluated and agreement
between methods was assessed using Bland-Altman plots. In the entire cohort, the mean femoral offset was 39.0 mm (95%
confidence interval (CI) 37.4 to 40.6) on radiographs of the pelvis,
44.0 mm (95% CI 42.4 to 45.6) on radiographs of the hip and 44.7
mm (95% CI 43.5 to 45.9) on CT scans. AP radiographs of the pelvis
underestimated femoral offset by 13% when compared with CT (p <
0.001).
No difference in mean femoral offset was seen between AP radiographs
of the hip and CT (p = 0.191). Our results suggest that femoral offset is significantly underestimated
on AP radiographs of the pelvis but can be reliably and accurately
assessed on AP radiographs of the hip in patients with primary end-stage
hip osteoarthritis. We, therefore, recommend that additional AP radiographs of the
hip are obtained routinely for the pre-operative assessment of femoral
offset when templating before total hip replacement.
Two-stage exchange remains the gold standard
for treatment of peri-prosthetic joint infection after total hip replacement
(THR). In the first stage, all components and associated cement
if present are removed, an aggressive debridement is undertaken
including a complete synovectomy, and an antibiotic-loaded cement
spacer is put in place. Patients are then treated with six weeks
of parenteral antibiotics, followed by an ‘antibiotic free period’
to help ensure the infection has been eradicated. If the clinical
evaluation and serum inflammatory markers suggest the infection
has resolved, then the second stage can be completed, which involves
removal of the cement spacer, repeat debridement, and placement
of a new THR. Cite this article:
We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature. The clinical significance of a raised metal ion level remains under investigation.
In this prospective study we studied the effect
of the inclination angle of the acetabular component on polyethylene wear
and component migration in cemented acetabular sockets using radiostereometric
analysis. A total of 120 patients received either a cemented Reflection
All-Poly ultra-high-molecular-weight polyethylene or a cemented
Reflection All-Poly highly cross-linked polyethylene acetabular
component, combined with either cobalt–chrome or Oxinium femoral
heads. Femoral head penetration and migration of the acetabular
component were assessed with repeated radiostereometric analysis
for two years. The inclination angle was measured on a standard
post-operative anteroposterior pelvic radiograph. Linear regression
analysis was used to determine the relationship between the inclination
angle and femoral head penetration and migration of the acetabular component. We found no relationship between the inclination angle and penetration
of the femoral head at two years’ follow-up (p = 0.9). Similarly,
our data failed to reveal any statistically significant correlation
between inclination angle and migration of these cemented acetabular
components (p = 0.07 to p = 0.9).
We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham hip resurfacing procedures (130 patients) at a minimum of ten years. There were ten revisions during this time. Although no patients were lost to follow-up some did not complete the scoring assessment or undergo radiological assessment at ten years. The ten-year survival for male patients was 98.0% (95% confidence interval 95.2 to 100). The ten-year survival for the total cohort with aseptic revision as the endpoint was 95.5% (95% confidence interval 91.8 to 99.0) and including revisions for sepsis was 93.5% (95% confidence interval 89.2 to 97.6). The median modified Oxford hip score at ten years was 4.2% (interquartile range 0 to 19) and the median University of California, Los Angeles score was 7.0 (interquartile range 5.0 to 8.0). This study confirms the midterm reports that metal-on-metal hip resurfacing using the Birmingham Hip provides a durable alternative to total hip replacement, particularly in younger male patients wishing to maintain a high level of function, with low risk of revision for at least ten years.
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.
We measured the plasma 25-hydroxyvitamin D3 (25(OH)D3) levels in 62 consecutive Caucasian patients undergoing total hip replacement for osteoarthritis. The patients were divided into two groups based on whether they were vitamin D sufficient or deficient. The groups were matched for age, gender and the American Society of Anaesthesiologists (ASA) grade. The prevalence of vitamin D deficiency in our patients was comparable with recent population-based studies performed in the United Kingdom. Patients with vitamin D deficiency had lower pre-operative Harris hip scores (Mann-Whitney test, p = 0.018) and were significantly less likely to attain an excellent outcome from total hip replacement (chi-squared test, p = 0.038). Vitamin D levels were found to positively correlate with both pre- and post-operative Harris hip scores. These results warrant further study of vitamin D deficiency in patients undergoing joint replacement as it is a risk factor for a suboptimal outcome which is relatively simple and cheap to correct.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
The design of the Charnley total hip replacement follows the principle of low frictional torque. It is based on the largest possible difference between the radius of the femoral head and that of the outer aspect of the acetabular component. The aim is to protect the bone-cement interface by movement taking place at the smaller radius, the articulation. This is achieved in clinical practice by a 22.225 mm diameter head articulating with a 40 mm or 43 mm diameter acetabular component of ultra-high molecular weight polyethylene. We compared the incidence of aseptic loosening of acetabular components with an outer diameter of 40 mm and 43 mm at comparable depths of penetration with a mean follow-up of 17 years (1 to 40). In cases with no measurable wear none of the acetabular components were loose. With increasing acetabular penetration there was an increased incidence of aseptic loosening which reflected the difference in the external radii, with 1.5% at 1 mm, 8.8% at 2 mm, 9.7% at 3 mm and 9.6% at 4 mm of penetration in favour of the larger 43 mm acetabular component. Our findings support the Charnley principle of low frictional torque. The level of the benefit is in keeping with the predicted values.
Total hip replacement for high dislocation of the hip joint remains technically difficult in terms of preparation of the true acetabulum and restoration of leg length. We describe our experience of cementless total hip replacement combined with a subtrochanteric femoral shortening osteotomy in 20 hips with Crowe grade IV dislocation with a mean follow-up of 8.1 years (4 to 11.5). There was one man and 17 women with a mean age of 55 years (44 to 69) at the time of the operation. After placment of the acetabular component at the site of the natural acetabulum, a cementless porous-coated cylindrical femoral component was implanted following a subtrochanteric femoral shortening osteotomy. The mean Japanese Orthopedic Association hip score improved from a mean of 38 (22 to 62) to a mean of 83 points (55 to 98) at the final follow-up. The mean lengthening of the leg was 14.8 mm (−9 to 34) in patients with iliofemoral osteoarthritis and 35.3 mm (15 to 51) in patients with no arthritic changes. No nerve palsy was observed. Total hip replacement combined with subtrochanteric shortening femoral osteotomy in this situation is beneficial in avoiding nerve injury and still permits valuable improvement in inequality of leg length.
The use of plate-and-cable constructs to treat periprosthetic fractures around a well-fixed femoral component in total hip replacements has been reported to have high rates of failure. Our aim was to evaluate the results of a surgical treatment algorithm to use these lateral constructs reliably in Vancouver type-B1 and type-C fractures. The joint was dislocated and the stability of the femoral component was meticulously evaluated in 45 type-B1 fractures. This led to the identification of nine (20%) unstable components. The fracture was considered to be suitable for single plate-and-cable fixation by a direct reduction technique if the integrity of the medial cortex could be restored. Union was achieved in 29 of 30 fractures (97%) at a mean of 6.4 months (3 to 30) in 29 type-B1 and five type-C fractures. Three patients developed an infection and one construct failed. Using this algorithm plate-and-cable constructs can be used safely, but indirect reduction with minimal soft-tissue damage could lead to shorter times to union and lower rates of complications.
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component. Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001). The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
Conventional treatment of mild slipped capital femoral epiphysis consists of fixation Between October 1984 and December 1995 we treated 16 patients for unilateral mild slipped capital femoral epiphysis by fixation While our clinical data favours conventional treatment, our radiological findings are in support of restoring the anatomy of the proximal femur to avoid or delay the development of femoroacetabular impingement following mild slipped capital femoral epiphysis.
Sciatic nerve palsy is a recognised complication of primary total hip replacement. In our unit this complication was rare with an incidence of <
0.2% in the past ten years. We describe six cases of sciatic nerve palsy occurring in 355 consecutive primary total hip replacements (incidence 1.69%). Each of these palsies was caused by post-operative haematoma in the region of the sciatic nerve. Cases, which were recognised early and surgically-evacuated promptly, showed earlier and more complete recovery. Those patients for whom the diagnosis was delayed, and who were therefore managed expectantly, showed little or no recovery. Unexpected pain and significant swelling in the buttock, as well as signs of sciatic nerve irritation, suggest the presence of haematoma in the region of the sciatic nerve. It is, therefore, of prime importance to be vigilant for the features of a sciatic nerve palsy in the early post-operative period as, when recognised and treated early, the injury to the sciatic nerve may be reversed.
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p <
0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.
We report a retrospective review of the incidence of venous thromboembolism in 463 consecutive patients who underwent primary total hip arthroplasty (487 procedures). Treatment included both total hip replacement and hip resurfacing, and the patients were managed without anticoagulants. The thromboprophylaxis regimen included an antiplatelet agent, generally aspirin, hypotensive epidural anaesthesia, elastic compression stockings and early mobilisation. In 258 of these procedures (244 patients) performed in 2005 (cohort A) mechanical compression devices were not used, whereas in 229 (219 patients) performed during 2006 (cohort B) bilateral intermittent pneumatic calf compression was used. All operations were performed through a posterior mini-incision approach. Patients who required anticoagulation for pre-existing medical problems and those undergoing revision arthroplasty were excluded. Doppler ultrasonographic screening for deep-vein thrombosis was performed in all patients between the fourth and sixth post-operative days. All patients were reviewed at a follow-up clinic six to ten weeks after the operation. In addition, reponse to a questionnaire was obtained at the end of 12 weeks post-operatively. No symptomatic calf or above-knee deep-vein thrombosis or pulmonary embolism occurred. In 25 patients in cohort A (10.2%) and in ten patients in cohort B (4.6%) asymptomatic calf deep-vein thromboses were detected ultrasonographically. This difference was statistically significant (p = 0.03). The regimen followed by cohort B offers the prospect of a low incidence of venous thromboembolism without subjecting patients to the higher risk of bleeding associated with anticoagulant use.
Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.
A reduced femoral offset in total hip replacement has been thought to be disadvantageous. We reviewed the results of 54 consecutive primary total hip replacements in 49 patients (mean age of 68 years) performed between August 1990 and December 1994, with a mean follow-up of 8.8 years ( At their latest follow-up, surviving patients had a significant improvement in the performance of their hip. Three had undergone revision, one each for deep infection, recurrent dislocation and late pain with subluxation. No hips had been revised, or were at risk of revision, for aseptic loosening. The mean annual linear rate of wear was 0.2 mm (