Aims. The aim of this study was to determine the clinical outcomes and factors contributing to failure of transposition osteotomy of the
Aims. This study aimed to determine clinical outcomes; relationships between postoperative anterior, lateral, and posterior acetabular coverage and joint survival; and prognostic factors for joint survival after transposition osteotomy of the
Aims. Although periacetabular osteotomies are widely used for the treatment of symptomatic dysplastic hips, long-term surgical outcomes and patient-reported outcome measures (PROMs) are still unclear. Accordingly, we assessed hip survival and PROMs at 20 years after transpositional osteotomy of the
Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and
Aims. The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the
We describe a little-known variety of hip dysplasia, termed ‘acetabular retroversion’, in which the alignment of the mouth of the
Aims. This study aimed to use intraoperative free electromyography to examine how the placement of a retractor at different positions along the anterior acetabular wall may affect the femoral nerve during total hip arthroplasty (THA) when undertaken using the direct anterior approach (THA-DAA). Methods. Intraoperative free electromyography was performed during primary THA-DAA in 82 patients (94 hips). The highest position of the anterior acetabular wall was defined as the “12 o’clock” position (middle position) when the patient was in supine position. After exposure of the
We have divided Severin group-V severely dysplastic hips with a false into three subtypes, based on the height and shape of the socket. We performed rotational acetabular osteotomy (RAO) in 19 hips in 17 young adults with a type-1 ‘low’ false
Our aim in this prospective study was to compare the bone mineral density (BMD) around cementless acetabular and femoral components which were identical in geometry and had the same alumina modular femoral head, but differed in regard to the material of the acetabular liners (alumina ceramic or polyethylene) in 50 patients (100 hips) who had undergone bilateral simultaneous primary total hip replacement. Dual energy X-ray absorptiometry scans of the pelvis and proximal femur were obtained at one week, at one year, and annually thereafter during the five-year period of the study. At the final follow-up, the mean BMD had increased significantly in each group in acetabular zone I of DeLee and Charnley (20% (15% to 26%), p = 0.003), but had decreased in acetabular zone II (24% (18% to 36%) in the alumina group and 25% (17% to 31%) in the polyethylene group, p = 0.001). There was an increase in the mean BMD in zone III of 2% (0.8% to 3.2%) in the alumina group and 1% (0.6% to 2.2%) in the polyethylene group (p = 0.315). There was a decrease in the mean BMD in the calcar region (femoral zone 7) of 15% (8% to 24%) in the alumina group and 14% (6% to 23%) in the polyethylene group (p <
0.001). The mean bone loss in femoral zone 1 of Gruen et al was 2% (1.1% to 3.1%) in the alumina group and 3% (1.3% to 4.3%) in the polyethylene group (p = 0.03), and in femoral zone 6, the mean bone loss was 15% (9% to 27%) in the alumina group and 14% (11% to 29%) in the polyethylene group compared with baseline values. There was an increase in the mean BMD on the final scans in femoral zones 2 (p = 0.04), 3 (p = 0.04), 4 (p = 0.12) and 5 (p = 0.049) in both groups. There was thus no significant difference in the bone remodelling of the
Large acetabular bone defects encountered in revision total hip arthroplasty (THA) are challenging to restore. Metal constructs for structural support are combined with bone graft materials for restoration. Autograft is restricted due to limited volume, and allogenic grafts have downsides including cost, availability, and operative processing. Bone graft substitutes (BGS) are an attractive alternative if they can demonstrate positive remodelling. One potential product is a biphasic injectable mixture (Cerament) that combines a fast-resorbing material (calcium sulphate) with the highly osteoconductive material hydroxyapatite. This study reviews the application of this biomaterial in large acetabular defects. We performed a retrospective review at a single institution of patients undergoing revision THA by a single surgeon. We identified 49 consecutive patients with large acetabular defects where the biphasic BGS was applied, with no other products added to the BGS. After placement of metallic acetabular implants, the BGS was injected into the remaining bone defects surrounding the new implants. Patients were followed and monitored for functional outcome scores, implant fixation, radiological graft site remodelling, and revision failures.Aims
Methods
We assessed the long-term outcomes of a large cohort of patients who have undergone a periacetabular osteotomy (PAO), and sought to validate a patient satisfaction questionnaire for use in a PAO cohort. All patients who had undergone a PAO from July 1998 to February 2013 were surveyed, with several patient-reported outcome measures (PROMs) and radiological measurements of preoperative acetabular dysplasia and postoperative correction also recorded. Patients were asked to rate their level of satisfaction with their operation in achieving pain relief, restoration of activities of daily living, ability to perform recreational activity, and their overall level of satisfaction with the procedure.Aims
Methods
A common situation presenting to the orthopaedic
surgeon today is a worn acetabular liner with substantial acetabular
and pelvic osteolysis. The surgeon has many options for dealing
with osteolytic defects. These include allograft, calcium based
substitutes, demineralised bone matrix, or combinations of these
options with or without addition of platelet rich plasma. To date
there are no clinical studies to determine the efficacy of using
bone-stimulating materials in osteolytic defects at the time of
revision surgery and there are surprisingly few studies demonstrating
the clinical efficacy of these treatment options. Even when radiographs
appear to demonstrate incorporation of graft material CT studies
have shown that incorporation is incomplete. The surgeon, in choosing
a graft material for a surgical procedure must take into account
the efficacy, safety, cost and convenience of that material. Cite this article:
The aim of this study was to analyze the effect of a lateral rim mesh on the survival of primary total hip arthroplasty (THA) in young patients, aged 50 years or younger. We compared a study group of 235 patients (257 hips) who received a primary THA with the use of impaction bone grafting (IBG) with an additional lateral rim mesh with a group of 306 patients (343 hips) who received IBG in the absence of a lateral rim mesh during the same period from 1988 to 2015. In the mesh group, there were 74 male and 183 female patients, with a mean age of 35 years (13 to 50). In the no-mesh group, there were 173 male and 170 female patients, with a mean age of 38 years (12.6 to 50). Cox regression analyses were performed to study the effect of a lateral rim mesh on acetabular component survival. Kaplan–Meier analyses with 95% confidence intervals (CIs) were performed to estimate the survival of the acetabular implant.Aims
Patients and Methods
The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising. Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft.
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at five years. All hips which failed did so because of aseptic loosening with progressive osteolysis or radiolucencies. We therefore recommend early and regular review of all patients with this combination of implants and early revision surgery in order to avoid massive bone loss.
Accurate placement of the acetabular component during total hip
arthroplasty (THA) is an important factor in the success of the
procedure. However, the reported accuracy varies greatly and is
dependent upon whether free hand or navigated techniques are used.
The aim of this study was to assess the accuracy of an instrument
system that incorporates 3D printed, patient-specific guides designed
to optimise the placement of the acetabular component. A total of 100 consecutive patients were prospectively enrolled
and the accuracy of placement of the acetabular component was measured
using post-operative CT scans.Aims
Patients and Methods
We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients.Aims
Patients and Methods
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of failure and modes of failure. We subsequently investigated
the factors that may affect the potential for the restoration of bone
stock and the durability of the re-revision reconstruction using
multivariate analysis. At the time of re-revision, there were ten host acetabula with
no significant defects, 14 with contained defects, nine with minor-column,
seven with major-column defects and ten with pelvic discontinuity.
When bone defects at re-revision were compared with those at the
previous revision, there was restoration of bone stock in 31 hips, deterioration
of bone stock in nine and remained unchanged in ten. This was a
significant improvement (p <
0.001). Morselised allografting
at the index revision was not associated with the restoration of
bone stock. In 17 hips (34%), re-revision was possible using a simple acetabular
component without allograft, augments, rings or cages. There were
47 patients with a mean follow-up of 70 months (6 to 146) available
for survival analysis. Within this group, the successful cases had
a minimum follow-up of two years after re-revision. There were 22 clinical
or radiological failures (46.7%), 18 of which were due to aseptic
loosening. The five and ten year Kaplan–Meier survival rate was
75% (95% CI, 60 to 86) and 56% (95% CI, 40 to 70) respectively with
aseptic loosening as the endpoint. The rate of aseptic loosening
was higher for hips with pelvic discontinuity (p = 0.049) and less
when the allograft had been in place for longer periods (p = 0.040). The use of a cage or ring over structural allograft bone for
massive uncontained defects in acetabular revision can restore host
bone stock and facilitate subsequent re-revision surgery to a certain
extent. Cite this article:
There are few reports describing the technique
of managing acetabular protrusio in primary total hip replacement. Most
are small series with different methods of addressing the challenges
of significant medial and proximal migration of the joint centre,
deficient medial bone and reduced peripheral bony support to the
acetabular component. We describe our technique and the clinical
and radiological outcome of using impacted morsellised autograft
with a porous-coated cementless cup in 30 primary THRs with mild
(n = 8), moderate (n = 10) and severe (n = 12) grades of acetabular
protrusio. The mean Harris hip score had improved from 52 pre-operatively
to 85 at a mean follow-up of 4.2 years (2 to 10). At final follow-up,
27 hips (90%) had a good or excellent result, two (7%) had a fair
result and one (3%) had a poor result. All bone grafts had united
by the sixth post-operative month and none of the hips showed any
radiological evidence of recurrence of protrusio, osteolysis or
loosening. By using impacted morsellised autograft and cementless
acetabular components it was possible to achieve restoration of
hip mechanics, provide a biological solution to bone deficiency
and ensure long-term fixation without recurrence in arthritic hips
with protrusio undergoing THR. Cite this article:
It is accepted that resurfacing hip replacement
preserves the bone mineral density (BMD) of the femur better than total
hip replacement (THR). However, no studies have investigated any
possible difference on the acetabular side. Between April 2007 and March 2009, 39 patients were randomised
into two groups to receive either a resurfacing or a THR and were
followed for two years. One patient’s resurfacing subsequently failed,
leaving 19 patients in each group. Resurfaced replacements maintained proximal femoral BMD and,
compared with THR, had an increased bone mineral density in Gruen
zones 2, 3, 6, and particularly zone 7, with a gain of 7.5% (95%
confidence interval (CI) 2.6 to 12.5) compared with a loss of 14.6%
(95% CI 7.6 to 21.6). Resurfacing replacements maintained the BMD
of the medial femoral neck and increased that in the lateral zones
between 12.8% (95% CI 4.3 to 21.4) and 25.9% (95% CI 7.1 to 44.6). On the acetabular side, BMD was similar in every zone at each
point in time. The mean BMD of all acetabular regions in the resurfaced
group was reduced to 96.2% (95% CI 93.7 to 98.6) and for the total
hip replacement group to 97.6% (95% CI 93.7 to 101.5) (p = 0.4863).
A mean total loss of 3.7% (95% CI 1.0 to 6.5) and 4.9% (95% CI 0.8
to 9.0) of BMD was found above the acetabular component in W1 and
10.2% (95% CI 0.9 to 19.4) and 9.1% (95% CI 3.8 to 14.4) medial
to the implant in W2 for resurfaced replacements and THRs respectively.
Resurfacing resulted in a mean loss of BMD of 6.7% (95% CI 0.7 to
12.7) in W3 but the BMD inferior to the acetabular component was
maintained in both groups. These results suggest that the ability of a resurfacing hip replacement
to preserve BMD only applies to the femoral side.
The technique of femoral cement-in-cement revision
is well established, but there are no previous series reporting its
use on the acetabular side at the time of revision total hip replacement.
We describe the technique and report the outcome of 60 consecutive
acetabular cement-in-cement revisions in 59 patients at a mean follow-up
of 8.5 years (5 to 12). All had a radiologically and clinically
well-fixed acetabular cement mantle at the time of revision. During
the follow-up 29 patients died, but no hips were lost to follow-up.
The two most common indications for acetabular revision were recurrent
dislocation (46, 77%) and to complement femoral revision (12, 20%). Of the 60 hips, there were two cases of aseptic loosening of
the acetabular component (3.3%) requiring re-revision. No other
hip was clinically or radiologically loose (96.7%) at the latest
follow-up. One hip was re-revised for infection, four for recurrent
dislocation and one for disarticulation of a constrained component.
At five years the Kaplan-Meier survival rate was 100% for aseptic
loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause
as the endpoint. These results support the use of cement-in-cement revision on
the acetabular side in appropriate cases. Theoretical advantages
include preservation of bone stock, reduced operating time, reduced
risk of complications and durable fixation.
The purpose of this study was to compare the
amount of acetabular bone removed during hip resurfacing (HR) and cementless
total hip replacement (THR), after controlling for the diameter
of the patient’s native femoral head. Based on a power analysis,
64 consecutive patients (68 hips) undergoing HR or THR were prospectively
enrolled in the study. The following data were recorded intra-operatively:
the diameter of the native femoral head, the largest reamer used,
the final size of the acetabular component, the size of the prosthetic
femoral head and whether a decision was made to increase the size
of the acetabular component in order to accommodate a larger prosthetic femoral
head. Results were compared using two-sided, independent samples
Student’s
Aims. To clarify the mid-term results of transposition osteotomy of the
Aims. Rotational acetabular osteotomy (RAO) has been reported to be effective in improving symptoms and preventing osteoarthritis (OA) progression in patients with mild to severe develomental dysplasia of the hip (DDH). However, some patients develop secondary OA even when the preoperative joint space is normal; determining who will progress to OA is difficult. We evaluated whether the preoperative cartilage condition may predict OA progression following surgery using T2 mapping MRI. Methods. We reviewed 61 hips with early-stage OA in 61 patients who underwent RAO for DDH. They underwent preoperative and five-year postoperative radiological analysis of the hip. Those with a joint space narrowing of more than 1 mm were considered to have 'OA progression'. Preoperative assessment of articular cartilage was also performed using 3T MRI with the T2 mapping technique. The region of interest was defined as the weightbearing portion of the
Aims. This study was performed to investigate the association between the acetabular morphology and the joint space narrowing rate (JSNR) in the non-arthritic hip. Methods. We retrospectively reviewed standing whole-leg radiographs of patients who underwent knee arthroplasty from February 2012 to March 2020 at our institute. Patients with a history of hip surgery, Kellgren-Lawrence grade ≥ II hip osteoarthritis, or rheumatoid arthritis were excluded. The hip JSNR was measured, and the normalized JSNR (nJSNR) was calculated by calibrating the joint space width with the size of the femoral head in 395 patients (790 hips) with a mean age of 73.7 years (SD 8.6). The effects of the lateral centre-edge angle (CEA) and acetabular roof obliquity (ARO) in the standing and supine positions were examined using a multivariate regression model. Results. The mean JSNR and nJSNR were 0.115 mm/year (SD 0.181) and 2.451 mm/year (SD 3.956), respectively. Multivariate regressions showed that older age was associated with a larger nJSNR (p = 0.010, standardized coefficient (SC) 0.096). The quadratic curve approximation showed that the joint space narrowing was smallest when the CEA was approximately 31.9°. This optimal CEA was the same in the standing and supine positions. Multivariate regressions were separately performed for joints with a CEA of < 31.9° and > 31.9°. When the CEA was < 31.9°, a smaller CEA was associated with a larger nJSNR (p < 0.001, SC 0.282). When the CEA was > 31.9°, a larger CEA was associated with a larger nJSNR (p = 0.012, SC 0.152). The ARO was not associated with the nJSNR. Conclusion. Both insufficient coverage and over-coverage of the
Aims. The aim of this study is to evaluate whether acetabular retroversion (AR) represents a structural anatomical abnormality of the pelvis or is a functional phenomenon of pelvic positioning in the sagittal plane, and to what extent the changes that result from patient-specific functional position affect the extent of AR. Methods. A comparative radiological study of 19 patients (38 hips) with AR were compared with a control group of 30 asymptomatic patients (60 hips). CT scans were corrected for rotation in the axial and coronal planes, and the sagittal plane was then aligned to the anterior pelvic plane. External rotation of the hemipelvis was assessed using the superior iliac wing and inferior iliac wing angles as well as quadrilateral plate angles, and correlated with cranial and central acetabular version. Sagittal anatomical parameters were also measured and correlated to version measurements. In 12 AR patients (24 hips), the axial measurements were repeated after matching sagittal pelvic rotation with standing and supine anteroposterior radiographs. Results. Acetabular version was significantly lower and measurements of external rotation of the hemipelvis were significantly increased in the AR group compared to the control group. The AR group also had increased evidence of anterior projection of the iliac wing in the sagittal plane. The acetabular orientation angles were more retroverted in the supine compared to standing position, and the change in acetabular version correlated with the change in sagittal pelvic tilt. An anterior pelvic tilt of 1° correlated with 1.02° of increased cranial retroversion and 0.76° of increased central retroversion. Conclusion. This study has demonstrated that patients with symptomatic AR have both an externally rotated hemipelvis and increased anterior projection of the iliac wing compared to a control group of asymptomatic patients. Functional sagittal pelvic positioning was also found to affect AR in symptomatic patients: the
Aims. There has been limited literature regarding outcomes of acetabular rim syndrome (ARS) with persistent acetabular os in the setting of acetabular dysplasia. The purpose of this study was to characterize a cohort of adolescent and young adult patients with ARS with persistent os and compare their radiological and clinical outcomes to patients with acetabular dysplasia without an os. Methods. We reviewed a prospective database of patients undergoing periacetabular osteotomy (PAO) for symptomatic acetabular dysplasia between January 1999 and December 2021 to identify hips with preoperative os acetabuli, defined as a closed triradiate cartilage but persistence of a superolateral os
Aims. Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the
Aims. In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Methods. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the
Aims. The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. Methods. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous
Aims. Hip arthroscopy (HA) has become the treatment of choice for femoroacetabular impingement (FAI). However, less favourable outcomes following arthroscopic surgery are expected in patients with severe chondral lesions. The aim of this study was to assess the outcomes of HA in patients with FAI and associated chondral lesions, classified according to the Outerbridge system. Methods. A systematic search was performed on four databases. Studies which involved HA as the primary management of FAI and reported on chondral lesions as classified according to the Outerbridge classification were included. The study was registered on PROSPERO. Demographic data, patient-reported outcome measures (PROMs), complications, and rates of conversion to total hip arthroplasty (THA) were collected. Results. A total of 24 studies were included with a total of 3,198 patients (3,233 hips). Patients had significantly less improvement in PROMs if they had Outerbridge grade III and IV lesions (p = 0.012). Compared with microfracture, autologous matrix-induced chondrogenesis (AMIC) resulted in significantly reduced rates of conversion to THA (p = 0.042) and of revision arthroscopy (p = 0.038). Chondral repair procedures in these patients also did not significantly reduce the rates of conversion to THA (p = 0.931), or of revision arthroscopy (p = 0.218). However, compared with microfracture, AMIC significantly reduced the rates of conversion to THA (p = 0.001) and of revision arthroscopy (p = 0.011) in these patients. Those with Outerbridge grade III and IV lesions also had significantly increased rates of conversion to THA (p = 0.029) and of revision arthroscopy (p = 0.023) if they had associated lesions of the
Aims. The aims of this study were to validate the Forgotten Joint Score-12 (FJS-12) in the postoperative evaluation of periacetabular osteotomy (PAO), identify factors associated with joint awareness after PAO, and determine the FJS-12 threshold for patient-acceptable symptom state (PASS). Methods. Data from 686 patients (882 hips) with hip dysplasia who underwent transposition osteotomy of the
Aims. Periacetabular osteotomy (PAO) is widely recognized as a demanding surgical procedure for acetabular reorientation. Reports about the learning curve have primarily focused on complication rates during the initial learning phase. Therefore, our aim was to assess the PAO learning curve from an analytical perspective by determining the number of PAOs required for the duration of surgery to plateau and the accuracy to improve. Methods. The study included 118 consecutive PAOs in 106 patients. Of these, 28 were male (23.7%) and 90 were female (76.3%). The primary endpoint was surgical time. Secondary outcome measures included radiological parameters. Cumulative summation analysis was used to determine changes in surgical duration. A multivariate linear regression model was used to identify independent factors influencing surgical time. Results. The learning curve in this series was 26 PAOs in a period of six months. After 26 PAO procedures, a significant drop in surgical time was observed and a plateau was also achieved. The mean duration of surgery during the learning curve was 103.8 minutes (SD 33.2), and 69.7 minutes (SD 18.6) thereafter (p < 0.001). Radiological correction of acetabular retroversion showed a significant improvement after having performed a total of 93 PAOs, including anteverting PAOs on 35 hips with a retroverted acetabular morphology (p = 0.005). Several factors were identified as independent variables influencing duration of surgery, including patient weight (β = 0.5 (95% confidence interval (CI) 0.2 to 0.7); p < 0.001), learning curve procedure phase of 26 procedures (β = 34.0 (95% CI 24.3 to 43.8); p < 0.001), and the degree of lateral correction expressed as the change in the lateral centre-edge angle (β = 0.7 (95% CI 0.001 to 1.3); p = 0.048). Conclusion. The learning curve for PAO surgery requires extensive surgical training at a high-volume centre, with a minimum of 50 PAOs per surgeon per year. This study defined a cut-off value of 26 PAO procedures, after which a significant drop in surgical duration occurred. Furthermore, it was observed that a retroverted morphology of the
Aims. Developmental dysplasia of the hip (DDH) describes a pathological relationship between the femoral head and
Aims. When the present study was initiated, we changed the treatment for late-detected developmental dislocation of the hip (DDH) from several weeks of skin traction to markedly shorter traction time. The aim of this prospective study was to evaluate this change, with special emphasis on the rate of stable closed reduction according to patient age, the development of the
Aims. In the native hip, the hip capsular ligaments tighten at the limits of range of hip motion and may provide a passive stabilizing force to protect the hip against edge loading. In this study we quantified the stabilizing force vectors generated by capsular ligaments at extreme range of motion (ROM), and examined their ability to prevent edge loading. Methods. Torque-rotation curves were obtained from nine cadaveric hips to define the rotational restraint contributions of the capsular ligaments in 36 positions. A ligament model was developed to determine the line-of-action and effective moment arms of the medial/lateral iliofemoral, ischiofemoral, and pubofemoral ligaments in all positions. The functioning ligament forces and stiffness were determined at 5 Nm rotational restraint. In each position, the contribution of engaged capsular ligaments to the joint reaction force was used to evaluate the net force vector generated by the capsule. Results. The medial and lateral arms of the iliofemoral ligament generated the highest inbound force vector in positions combining extension and adduction providing anterior stability. The ischiofemoral ligament generated the highest inbound force in flexion with adduction and internal rotation (FADIR), reducing the risk of posterior dislocation. In this position the hip joint reaction force moved 0.8° inbound per Nm of internal capsular restraint, preventing edge loading. Conclusion. The capsular ligaments contribute to keep the joint force vector inbound from the edge of the
Aims. The aim of this study was to understand the experience of mature patients who undergo a periacetabular osteotomy (PAO), a major hip-preserving surgical procedure that treats symptomatic hip dysplasia by realigning the
Aims. The primary aim of our study was to assess the influence of age on hip-specific outcome following total hip arthroplasty (THA). Secondary aims were to assess health-related quality of life (HRQoL) and level of activity according to age. Methods. A prospective cohort study was conducted. All patients were fitted with an Exeter stem with a 32 mm head on highly cross-linked polyethylene (X3RimFit) cemented
Aims. It may not be possible to undertake revision total hip arthroplasty
(THA) in the presence of massive loss of acetabular bone stock using
standard cementless hemispherical acetabular components and metal
augments, as satisfactory stability cannot always be achieved. We
aimed to study the outcome using a reconstruction cage and a porous
metal augment in these patients. Patients and Methods. A total of 22 acetabular revisions in 19 patients were performed
using a combination of a reconstruction cage and porous metal augments.
The augments were used in place of structural allografts. The mean
age of the patients at the time of surgery was 70 years (27 to 85)
and the mean follow-up was 39 months (27 to 58). The mean number
of previous THAs was 1.9 (1 to 3). All patients had segmental defects
involving more than 50% of the
Aims. Cross-table lateral (CTL) radiographs are commonly used to measure acetabular component anteversion after total hip arthroplasty (THA). The CTL measurements may differ by > 10° from CT scan measurements but the reasons for this discrepancy are poorly understood. Anteversion measurements from CTL radiographs and CT scans are compared to identify spinopelvic parameters predictive of inaccuracy. Methods. THA patients (n = 47; 27 males, 20 females; mean age 62.9 years (SD 6.95)) with preoperative spinopelvic mobility, radiological analysis, and postoperative CT scans were retrospectively reviewed. Acetabular component anteversion was measured on postoperative CTL radiographs and CT scans using 3D reconstructions of the pelvis. Two cohorts were identified based on a CTL-CT error of ≥ 10° (n = 11) or < 10° (n = 36). Spinopelvic mobility parameters were compared using independent-samples t-tests. Correlation between error and mobility parameters were assessed with Pearson’s coefficient. Results. Patients with CTL error > 10° (10° to 14°) had stiffer lumbar spines with less mean lumbar flexion (38.9°(SD 11.6°) vs 47.4° (SD 13.1°); p = 0.030), different sagittal balance measured by pelvic incidence-lumbar lordosis mismatch (5.9° (SD 18.8°) vs -1.7° (SD 9.8°); p = 0.042), more pelvic extension when seated (pelvic tilt -9.7° (SD 14.1°) vs -2.2° (SD 13.2°); p = 0.050), and greater change in pelvic tilt between supine and seated positions (12.6° (SD 12.1°) vs 4.7° (SD 12.5°); p = 0.036). The CTL measurement error showed a positive correlation with increased CTL anteversion (r = 0.5; p = 0.001), standing lordosis (r = 0.23; p = 0.050), seated lordosis (r = 0.4; p = 0.009), and pelvic tilt change between supine and step-up positions (r = 0.34; p = 0.010). Conclusion. Differences in spinopelvic mobility may explain the variability of acetabular anteversion measurements made on CTL radiographs. Patients with stiff spines and increased compensatory pelvic movement have less accurate measurements on CTL radiographs. Flexion of the contralateral hip is required to obtain clear CTL radiographs. In patients with lumbar stiffness, this movement may extend the pelvis and increase anteversion of the
Aims. One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the
In this study we used subject-specific finite
element analysis to investigate the mechanical effects of rotational acetabular
osteotomy (RAO) on the hip joint and analysed the correlation between
various radiological measurements and mechanical stress in the hip
joint. We evaluated 13 hips in 12 patients (two men and ten women, mean
age at surgery 32.0 years; 19 to 46) with developmental dysplasia
of the hip (DDH) who were treated by RAO. Subject-specific finite element models were constructed from
CT data. The centre–edge (CE) angle, acetabular head index (AHI),
acetabular angle and acetabular roof angle (ARA) were measured on
anteroposterior pelvic radiographs taken before and after RAO. The
relationship between equivalent stress in the hip joint and radiological measurements
was analysed. The equivalent stress in the
Three-dimensional surface models of the normal hemipelvis derived from volumetric CT data on 42 patients were used to determine the radius, depth and orientation of the native
Aims. We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability. Patients and Methods. A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true
Eccentric reductions may become concentric through femoral head ‘docking’ (FHD) following closed reduction (CR) for developmental dysplasia of the hip (DDH). However, changes regarding position and morphology through FHD are not well understood. We aimed to assess these changes using serial MRI. We reviewed 103 patients with DDH successfully treated by CR and spica casting in a single institution between January 2016 and December 2020. MRI was routinely performed immediately after CR and at the end of each cast. Using MRI, we described the labrum-acetabular cartilage complex (LACC) morphology, and measured the femoral head to triradiate cartilage distance (FTD) on the midcoronal section. A total of 13 hips with initial complete reduction (i.e. FTD < 1 mm) and ten hips with incomplete MRI follow-up were excluded. A total of 86 patients (92 hips) with a FTD > 1 mm were included in the analysis.Aims
Methods
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted
Sagittal lumbar pelvic alignment alters with posterior pelvic tilt (PT) following total hip arthroplasty (THA) for developmental dysplasia of the hip (DDH). The individual value of pelvic sagittal inclination (PSI) following rebalancing of lumbar-pelvic alignment is unknown. In different populations, PT regresses in a linear relationship with pelvic incidence (PI). PSI and PT have a direct relationship to each other via a fixed individual angle ∠γ. This study aimed to investigate whether the new PI created by acetabular component positioning during THA also has a linear regression relationship with PT/PSI when lumbar-pelvic alignment rebalances postoperatively in patients with Crowe type III/IV DDH. Using SPINEPARA software, we measured the pelvic sagittal parameters including PI, PT, and PSI in 61 patients with Crowe III/IV DDH. Both PSI and PT represent the pelvic tilt state, and the difference between their values is ∠γ (PT = PSI + ∠γ). The regression equation between PI and PT at one year after THA was established. By substituting ∠γ, the relationship between PI and PSI was also established. The Bland-Altman method was used to evaluate the consistency between the PSI calculated by the linear regression equation (ePSI) and the actual PSI (aPSI) measured one year postoperatively.Aims
Methods
Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies. Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment.Aims
Methods
Hyaline cartilage has a low capacity for regeneration. Untreated osteochondral lesions of the femoral head can lead to progressive and symptomatic osteoarthritis of the hip. The purpose of this study is to analyze the clinical and radiological long-term outcome of patients treated with osteochondral autograft transfer. To our knowledge, this study represents a series of osteochondral autograft transfer of the hip with the longest follow-up. We retrospectively evaluated 11 hips in 11 patients who underwent osteochondral autograft transfer in our institution between 1996 and 2012. The mean age at the time of surgery was 28.6 years (8 to 45). Outcome measurement included standardized scores and conventional radiographs. Kaplan-Meier survival curve was used to determine the failure of the procedures, with conversion to total hip arthroplasty (THA) defined as the endpoint.Aims