Limited access surgery is thought to reduce post-operative morbidity and provide faster recovery of function. The percutaneous compression plate (PCCP) is a recently introduced device for the fixation of intertrochanteric fractures with minimal exposure. It has several potential mechanical advantages over the conventional compression hip screw (CHS). Our aim in this prospective, randomised, controlled study was to compare the outcome of patients operated on using these two devices. We randomised 104 patients with intertrochanteric fractures (AO/OTA 31.A1–A2) to surgical treatment with either the PCCP or CHS and followed them for one year postoperatively. The mean operating blood loss was 161.0 ml (8 to 450) in the PCCP group and 374.0 ml (11 to 980) in the CHS group (Student’s The PCCP device was associated with reduced intra-operative blood loss, less postoperative pain and a reduced incidence of collapse of the fracture.