We determined the short-term clinical outcome
and migration within the bone of the humeral cementless component
of the Instrumented Bone Preserving (IBP) total elbow replacement
in a series of 16 patients. There were four men and 12 women with
a mean age at operation of 63 years (40 to 81). Migration was calculated
using radiostereometric analysis. There were no intra-operative
complications and no revisions. At two-year follow-up, all patients
showed a significant reduction in pain and functional improvement
of the elbow (both p <
0.001). Although ten components (63%)
showed movement or micromovement during the first six weeks, 14
(88%) were stable at one year post-operatively. Translation was
primarily found in the proximal direction (median 0.3 mm (interquartile range
(IQR) -0.09 to 0.8); the major rotational movement was an anterior
tilt (median 0.7° (IQR 0.4° to 1.6°)). One malaligned component
continued to migrate during the second year, and one component could
not be followed beyond three months because migration had caused
the markers to break off the prosthesis. This study shows promising early results for the cementless humeral
component of the IBP total elbow replacement. All patients had a
good clinical outcome, and most components stabilised within six
months of the operation. Cite this article:
Valgus high tibial osteotomy for osteoarthritis of the medial compartment of the knee can be performed using medial opening- and lateral closing-wedge techniques. The latter have been thought to offer greater initial stability. We measured and compared the stability of opening- and closing-wedge osteotomies fixed by TomoFix plates using radiostereometry in a series of 42 patients in a prospective, randomised clinical trial. There were no differences between the opening- and closing-wedge groups in the time to regain knee function and full weight-bearing. Pain and knee function were significantly improved in both groups without any differences between them. All the osteotomies united within one year. Radiostereometry showed no clinically relevant movement of bone or differences between either group. Medial opening-wedge high tibial osteotomy secured by a TomoFix plate offers equal stability to a lateral closing-wedge technique. Both give excellent initial stability and provide significantly improved knee function and reduction in pain, although the opening-wedge technique was more likely to produce the intended correction.