Charcot neuro-osteoarthropathy (CN) of the midfoot
presents a major reconstructive challenge for the foot and ankle
surgeon. The Synthes 6 mm Midfoot Fusion Bolt is both designed and
recommended for patients who have a deformity of the medial column
of the foot due to CN. We present the results from the first nine
patients (ten feet) on which we attempted to perform fusion of the
medial column using this bolt. Six feet had concurrent hindfoot fusion
using a retrograde nail. Satisfactory correction of deformity of
the medial column was achieved in all patients. The mean correction
of calcaneal pitch was from 6°
(-15° to +18°) pre-operatively to 16° (7° to 23°) post-operatively;
the mean Meary angle from 26° (3° to 46°) to 1° (1° to 2°); and
the mean talometatarsal angle on dorsoplantar radiographs from 27°
(1° to 48°) to 1° (1° to 3°). However, in all but two feet, at least one joint failed to fuse.
The bolt migrated in six feet, all of which showed progressive radiographic
osteolysis, which was considered to indicate loosening. Four of
these feet have undergone a revision procedure, with good radiological
evidence of fusion. The medial column bolt provided satisfactory correction
of the deformity but failed to provide adequate fixation for fusion
in CN deformities in the foot. In its present form, we cannot recommend the routine use of this
bolt. Cite this article: