Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
Bone & Joint Research
Vol. 5, Issue 5 | Pages 191 - 197
1 May 2016
Kienast B Kowald B Seide K Aljudaibi M Faschingbauer M Juergens C Gille J

Objectives. The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator. Materials and Methods. An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur. Results. A total of 39 cases of clinical application were reviewed for this study. In total, four different types of healing curves were observed: fast healing; slow healing; plateau followed by healing; and non-healing. Conclusion. The electronically instrumented internal fixator proved to be valuable for the assessment of bone healing in difficult healing situations. Cost-effective manufacturing is possible because the used electronic components are derived from large-scale production. The incorporation of microelectronics into orthopaedic implants will be an important innovation in future clinical care. Cite this article: B. Kienast, B. Kowald, K. Seide, M. Aljudaibi, M. Faschingbauer, C. Juergens, J. Gille. An electronically instrumented internal fixator for the assessment of bone healing. Bone Joint Res 2016;5:191–197. DOI: 10.1302/2046-3758.55.2000611


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 398 - 404
1 Mar 2012
Seide K Aljudaibi M Weinrich N Kowald B Jürgens C Müller J Faschingbauer M

In an interdisciplinary project involving electronic engineers and clinicians, a telemetric system was developed to measure the bending load in a titanium internal femoral fixator. As this was a new device, the main question posed was: what clinically relevant information could be drawn from its application? As a first clinical investigation, 27 patients (24 men, three women) with a mean age of 38.4 years (19 to 66) with femoral nonunions were treated using the system. The mean duration of the nonunion was 15.4 months (5 to 69). The elasticity of the plate-callus system was measured telemetrically until union. Conventional radiographs and a CT scan at 12 weeks were performed routinely, and healing was staged according to the CT scans. All nonunions healed at a mean of 21.5 weeks (13 to 37). Well before any radiological signs of healing could be detected, a substantial decrease in elasticity was recorded. The relative elasticity decreased to 50% at a mean of 7.8 weeks (3.5 to 13) and to 10% at a mean of 19.3 weeks (4.5 to 37). At 12 weeks the mean relative elasticity was 28.1% (0% to 56%). The relative elasticity was significantly different between the different healing stages as determined by the CT scans. Incorporating load measuring electronics into implants is a promising option for the assessment of bone healing. Future application might lead to a reduction in the need for exposure to ionising radiation to monitor fracture healing


Bone & Joint Open
Vol. 3, Issue 2 | Pages 165 - 172
21 Feb 2022
Kuwahara Y Takegami Y Tokutake K Yamada Y Komaki K Ichikawa T Imagama S

Aims

Postoperative malalignment of the femur is one of the main complications in distal femur fractures. Few papers have investigated the impact of intraoperative malalignment on postoperative function and bone healing outcomes. The aim of this study was to investigate how intraoperative fracture malalignment affects postoperative bone healing and functional outcomes.

Methods

In total, 140 patients were retrospectively identified from data obtained from a database of hospitals participating in a trauma research group. We divided them into two groups according to coronal plane malalignment of more than 5°: 108 had satisfactory fracture alignment (< 5°, group S), and 32 had unsatisfactory alignment (> 5°, group U). Patient characteristics and injury-related factors were recorded. We compared the rates of nonunion, implant failure, and reoperation as healing outcomes and Knee Society Score (KSS) at three, six, and 12 months as functional outcomes. We also performed a sub-analysis to assess the effect of fracture malalignment by plates and nails on postoperative outcomes.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 787 - 797
1 Jun 2018
Shuler MS Roskosky M Kinsey T Glaser D Reisman W Ogburn C Yeoman C Wanderman NR Freedman B

Aims

The aim of this study was to evaluate near-infrared spectroscopy (NIRS) as a continuous, non-invasive monitor for acute compartment syndrome (ACS).

Patients and Methods

NIRS sensors were placed on 86 patients with, and 23 without (controls), severe leg injury. NIRS values were recorded for up to 48 hours. Longitudinal data were analyzed using summary and graphical methods, bivariate comparisons, and multivariable multilevel modelling.