The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 1 | Pages 53 - 57
1 Jan 1997
Noordeen MHH Lee J Gibbons CER Taylor BA Bentley G

We reviewed retrospectively the role of monitoring of somatosensory spinal evoked potentials (SSEP) in 99 patients with neuromuscular scoliosis who had had operative correction with Luque-Galveston rods and sublaminar wiring. Our findings showed that SSEP monitoring was useful and that a 50% decrease in the amplitude of the trace optimised both sensitivity and specificity. The detection of true-positive results was higher than in cases of idiopathic scoliosis, but the method was less sensitive and specific and there were more false-negative results. In contrast with the findings in idiopathic scoliosis, recovery of the trace was associated with a 50% to 60% risk of neurological impairment. Only one permanent injury occurred during the use of this technique, and any temporary impairment resolved within two months