Aims. Tissue responses to debris formed by abrasion of polymethylmethacrylate
(PMMA) spacers at two-stage revision arthroplasty for prosthetic
joint infection are not well described. We hypothesised that PMMA
debris induces immunomodulation in periprosthetic tissues. Patients and Methods. Samples of tissue were taken during 35 two-stage revision arthroplasties
(nine total hip and 26 total knee arthroplasties) in patients whose
mean age was 67 years (44 to 85). Fourier transform infrared microscopy
was used to confirm the presence of
Fungal periprosthetic joint infections (fPJIs) are rare complications, constituting only 1% of all PJIs. Neither a uniform definition for fPJI has been established, nor a standardized treatment regimen. Compared to bacterial PJI, there is little evidence for fPJI in the literature with divergent results. Hence, we implemented a novel treatment algorithm based on three-stage revision arthroplasty, with local and systemic antifungal therapy to optimize treatment for fPJI. From 2015 to 2018, a total of 18 patients with fPJI were included in a prospective, single-centre study (DKRS-ID 00020409). The diagnosis of PJI is based on the European Bone and Joint Infection Society definition of periprosthetic joint infections. The baseline parameters (age, sex, and BMI) and additional data (previous surgeries, pathogen spectrum, and Charlson Comorbidity Index) were recorded. A therapy protocol with three-stage revision, including a scheduled spacer exchange, was implemented. Systemic antifungal medication was administered throughout the entire treatment period and continued for six months after reimplantation. A minimum follow-up of 24 months was defined.Aims
Methods
We describe the development and early clinical application of a ported, proximally-cemented titanium stem for cemented total hip arthroplasty.
Vancomycin is commonly added to acrylic bone cement during revision
arthroplasty surgery. Proprietary cement preparations containing
vancomycin are available, but are significantly more expensive.
We investigated whether the elution of antibiotic from ‘home-made’
cement containing vancomycin was comparable with more expensive
commercially available vancomycin impregnated cement. A total of 18 cement discs containing either proprietary CopalG+V;
or ‘home-made’ CopalR+G with vancomycin added by hand, were made.
Each disc contained the same amount of antibiotic (0.5 g gentamycin,
2 g vancomycin) and was immersed in ammonium acetate buffer in a
sealed container. Fluid from each container was sampled at eight
time points over a two-week period. The concentrations of gentamicin
and vancomycin in the fluid were analysed using high performance
liquid chromatography mass spectrometry.Aims
Materials and Methods
Periprosthetic joint infection (PJI) is a devastating
complication for patients and results in greatly increased costs
of care for both healthcare providers and patients. More than 15
500 revision hip and knee procedures were recorded in England, Wales
and Northern Ireland in 2013, with infection accounting for 13%
of revision hip and 23% of revision knee procedures. We report our experience of using antibiotic eluting absorbable
calcium sulphate beads in 15 patients (eight men and seven women
with a mean age of 64.8 years; 41 to 83) as part of a treatment
protocol for PJI in revision arthroplasty. The mean follow-up was 16 months (12 to 22). We report the outcomes
and complications, highlighting the risk of hypercalcaemia which
occurred in three patients. We recommend that serum levels of calcium be routinely sought
following the implantation of absorbable calcium sulphate beads
in orthopaedic surgery. Cite this article:
The aim of this study was to identify the incidence of positive
cultures during the second stage of a two-stage revision arthroplasty
and to analyse the association between positive cultures and an
infection-free outcome. This single-centre retrospective review of prospectively collected
data included patients with a periprosthetic joint infection (PJI)
of either the hip or the knee between 2013 and 2015, who were treated
using a standardised diagnostic and therapeutic algorithm with two-stage
exchange. Failure of treatment was assessed according to a definition
determined by a Delphi-based consensus. Logistic regression analysis
was performed to assess the predictors of positive culture and risk
factors for failure. The mean follow-up was 33 months (24 to 48).Aims
Patients and Methods
Bone cements produced by different manufacturers vary in their mechanical properties and antibiotic elution characteristics. Small changes in the formulation of a bone cement, which may not be apparent to surgeons, can also affect these properties. The supplier of Palacos bone cement with added gentamicin changed in 2005. We carried out a study to examine the mechanical characteristics and antibiotic elution of Schering-Plough Palacos, Heraeus Palacos and Depuy CMW Smartset bone cements. Both Heraeus Palacos and Smartset bone cements performed significantly better than Schering-Plough Palacos in terms of mechanical characteristics, with and without additional vancomycin (p <
0.001). All cements show a deterioration in flexural strength with increasing addition of vancomycin, albeit staying above ISO minimum levels. Both Heraeus Palacos and Smartset elute significantly more gentamicin cumulatively than Schering-Plough Palacos. Smartset elutes significantly more vancomycin cumulatively than Heraeus Palacos. The improved antibiotic elution characteristics of Smartset and Heraeus Palacos are not associated with a deterioration in mechanical properties. Although marketed as the ‘original’ Palacos, Heraeus Palacos has significantly altered mechanical and antibiotic elution characteristics compared with the most commonly-used previous version.