National Institute of Clinical Excellence guidelines
state that cemented stems with an Orthopaedic Data Evaluation Panel
(ODEP) rating of >
3B should be used for hemiarthroplasty when treating
an intracapsular fracture of the femoral neck. These recommendations
are based on studies in which most, if not all stems, did not hold
such a rating. . This case-control study compared the outcome of hemiarthroplasty
using a cemented (Exeter) or uncemented (Corail) femoral stem. These
are the two prostheses most commonly used in hip arthroplasty in
the UK. Data were obtained from two centres; most patients had undergone
hemiarthroplasty using a cemented Exeter stem (n = 292/412). Patients
were matched for all factors that have been shown to influence mortality
after an intracapsular fracture of the neck of the femur. Outcome
measures included: complications, re-operations and mortality rates
at two, seven, 30 and 365 days post-operatively. Comparable outcomes
for the two stems were seen. . There were more intra-operative complications in the uncemented
group (13% vs 0%), but the cemented group had a
greater mortality in the early post-operative period (n = 6). There
was no overall difference in the rate of re-operation (5%) or death
(365 days: 26%) between the two groups at any time post-operatively. This study therefore supports the use of both cemented and uncemented
stems of proven design, with an
Background. Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an
Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes. Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.Aims
Methods
This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation.Aims
Patients and Methods
A total of 56 male patients with a displaced
intracapsular fracture of the hip and a mean age of 81 years (62
to 94), were randomised to be treated with either a cemented hemiarthroplasty
(the Exeter Trauma Stem) or reduction and internal fixation using
the Targon Femoral Plate. All surviving patients were reviewed one
year after the injury, at which time restoration of function and
pain in the hip was assessed. There was no statistically significant
difference in mortality between the two groups (7/26; 26.9% for
hemiarthroplasty These results indicate that cemented hemiarthroplasty gives better
results than internal fixation in elderly men with a displaced intracapsular
fracture of the hip. Cite this article:
The National Institute for Health and Clinical
Excellence (NICE) guidelines from 2011 recommend the use of cemented
hemi-arthroplasty for appropriate patients with an intracapsular
hip fracture. In our institution all patients who were admitted
with an intracapsular hip fracture and were suitable for a hemi-arthroplasty
between April 2010 and July 2012 received an uncemented prosthesis
according to our established departmental routine practice. A retrospective
analysis of outcome was performed to establish whether the continued
use of an uncemented stem was justified. Patient, surgical and outcome
data were collected on the National Hip Fracture database. A total
of 306 patients received a Cathcart modular head on a Corail uncemented
stem as a hemi-arthroplasty. The mean age of the patients was 83.3
years ( Cite this article: