The objective of this study was to investigate bone healing after
internal fixation of displaced femoral neck fractures (FNFs) with
the Dynamic Locking Blade Plate (DLBP) in a young patient population
treated by various orthopaedic (trauma) surgeons. We present a multicentre prospective case series with a follow-up
of one year. All patients aged ≤ 60 years with a displaced FNF treated
with the DLBP between 1st August 2010 and December 2014 were included.
Patients with pathological fractures, concomitant fractures of the
lower limb, symptomatic arthritis, local infection or inflammation,
inadequate local tissue coverage, or any mental or neuromuscular
disorder were excluded. Primary outcome measure was failure in fracture
healing due to nonunion, avascular necrosis, or implant failure
requiring revision surgery.Aims
Patients and Methods
We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study.