Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1238 - 1244
1 Sep 2010
Nicoll D Rowley DI

This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003.

The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain.

Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 6 | Pages 819 - 824
1 Aug 2001
Prakash U Wigderowitz CA McGurty DW Rowley DI

Tibiofemoral alignment has a direct correlation with the survival of total knee arthroplasty. Traditionally, it has been measured using a goniometer on radiographs. We describe new software which measures this alignment on scanned radiographs by automatically detecting bones in the image.

Two surgeons used conventional methods and two clerical officers used the computerised routine to assess 58 radiographs of the knee on two occasions. There were no significant differences between any of the paired comparisons. The largest mean difference detected was 1.19°. Across all comparisons, the mean correlation was 0.755. A standardised routine for measuring tibiofemoral alignment was the greatest factor in reducing error in our study.

These results show that non-medical staff can reliably use the software to measure tibiofemoral alignment. It has the potential to measure all the parameters recommended by the Knee Society.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 2 | Pages 185 - 190
1 Mar 2001
Rowley DI McGurty DW

We describe a method of audit of a type of total knee replacement, including some details of the organisational difficulties of administering multicentre studies, and draw attention to how this can be done using industrial funding without prejudicing the study. This is a prospective record of 1439 patients who had an Insall-Burstein II (IBII) prosthesis implanted between 1990 and 1994. The data were collected using the American Knee Society scoring system. A method of storing radiographs digitally at low cost is also described.

The results emphasise the need for the long-term collection of data on commonly used devices implanted by a cross-section of surgeons. We conclude that for most patients the IBII cemented, posteriorly stabilised, cruciate-substituting prosthesis will relieve pain and give excellent functional results throughout the patients’ remaining years with a very small incidence of revision, except in cases of infection.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 217 - 221
1 Mar 2000
McNicholas MJ Rowley DI McGurty D Adalberth T Abdon P Lindstrand A Lohmander LS

We have carried out a prospective, longitudinal 30-year review of 95 adolescents who underwent total meniscectomy in one knee, and have compared the results with those observed 13 years earlier. All the medical records were scrutinised. Of the 63 patients reviewed clinically, 47 reported decreased sporting activity, although subjective satisfaction rose by 3% to 71%. The scores on the WOMAC osteoarthritis index differed significantly between patients grouped by subjective global assessment. Satisfactory function scores increased from 48% to 60%. In the 53 patients consenting to bilateral radiography of the knee, the incidence of narrowing of the articular cartilage in the operated knee increased significantly between the reviews (19% to 36%). Progression of degenerative change paralleled reduction in activity. Outcome measures were best after medial, intermediate after lateral and worst after double meniscectomy.