The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of acetabular inclination,
neck-shaft and stem-shaft angles, and functional outcome scores
including the Harris hip score, the Western Ontario and McMaster
Universities Osteoarthritis Index and the University of California
at Los Angeles activity scale. In total, 17 patients were lost to follow-up leaving 103 patients
at two years. There were no revisions in the cementless group and
three revisions (5%) in the cemented group (two because of hip pain
and one for pseudotumour). The total BMD was significantly higher in the cementless group
at six months (p <
0.001) and one year (p = 0.01) than in the
cemented group, although there was a loss of statistical significance
in the difference at two years (p = 0.155). All patient outcomes improved significantly: there were no significant
differences between the two groups. The results show better preservation of femoral neck BMD with
a cementless femoral component after two years of follow-up. Further
investigation is needed to establish whether this translates into
improved survivorship. Cite this article:
The introduction of a trabecular tantalum rod
has been proposed for the management of early-stage osteonecrosis of
the femoral head but serves as a single-point of support of the
necrotic lesion. We describe a technique using two or three 4.2
mm (or later 4.7 mm) tantalum pegs for the prevention of collapse
of the necrotic lesion. We prospectively studied 21 patients (26
hips) with non-traumatic osteonecrosis of the femoral head treated
in this manner. Of these, 21 patients (24 hips) were available for
radiological and clinical evaluation at a mean follow-up of 46 months
(18 to 67). Radiological assessment showed that only eight hips
deteriorated according to the Association Research Circulation Osseous
classification, and four hips according to the Classification of
the Japanese Investigation Committee of Health and Welfare. Functional
improvement was obtained with an improvement in the mean Harris
hip score from 65.2 (33.67 to 95) to 88.1 (51.72 to 100), the mean
Merle D’Aubigné-Postel score from 13 (6 to 18) to 16 (11 to 18),
a mean visual analogue score for pain from 5.2 (0 to 9.5) to 2.6
(0 to 7), and the mean Short-Form 36 score from 80.4 (56.8 to 107.1)
to 92.4 (67.5 to 115.7). Of these 24 hips followed for a minimum
of 18 months, three were considered as failures at the final follow-up,
having required total hip replacement. One of the hips without full
follow-up was also considered to be a failure. In more than two-thirds
of the surviving hips a satisfactory clinical outcome was achieved
with promising radiological findings. The estimated mean implant
survival was 60 months (95% confidence interval 53.7 to 66.3).