Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 235 - 241
1 Feb 2010
van der Lugt JCT Valstar ER Witvoet-Braam SW Nelissen RGHH

Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs.

All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (sd 1.0) and in the non-revised 0.7 mm (sd 0.5, p = 0.01). Most humeral components migrated into external rotation resulting in an anterior and varus tilt. The ulnar components remained stable.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 6 | Pages 1005 - 1012
1 Nov 1999
Gschwend N Scheier NH Baehler AR

Between 1978 and 1986, 59 patients received a GSB III elbow prosthesis, six of them in both elbows. Rheumatoid arthritis (RA) was the underlying cause in 51 of the patients and post-traumatic osteoarthritis (PTOA) in eight. Of these, 24 patients (28 prostheses) have since died; two, both operated on bilaterally, had had their implants for more than ten years and had already been assessed for inclusion in the long-term follow-up. Two patients, each with one elbow prosthesis, have been lost to follow-up and three males who are still living (two with PTOA, one with juvenile RA) had their prosthesis removed before ten years had elapsed. The remaining 32 patients (28 RA, 4 PTOA) with 36 GSB III elbows were examined clinically and radiologically after a mean period of 13.5 years. Pain was considerably reduced in 91.6%. Mobility was increased by 37° in those with RA and by 67° in those with PTOA. There were three cases of aseptic loosening and three of deep infection. The main complication was disassembly of the prosthetic component in nine elbows (13.8%). This last group included two patients with postoperative fractures unrelated to the operative technique and one with neuropathic arthritis. Ulnar neuritis occurred in two patients. Since 87.7% of all the GSB III prostheses implanted in this period remained in situ, our results are comparable with those for hip and knee arthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1129 - 1134
1 Nov 2000
Rozing P

We have reviewed 66 consecutive Souter-Strathclyde arthroplasties of the elbow implanted in 59 patients between 1982 and 1993. Thirteen patients (15 elbows) (19.6%) died. Sixteen elbows (24.2%) were revised, six for aseptic loosening (9%), four (6%) because of fracture or loosening after a fracture, three (4.5%) for infection and three (4.5%) for dislocation. Four patients refused to attend for review. In 33 elbows with a follow-up of 93 months (60 to 167) complete relief of pain was achieved in 22 (67%) when seen at one year. After ten years or more 36% of the elbows were painfree and 64% had occasional slight pain especially under loading or stress. The mean gain in the arc of movement was 16°, but a mean flexion contracture of 33° remained. The main early complications were intraoperative fractures of the epicondyles (9%), postoperative dislocation (4.5%) and ulnar neuropathy. The incidence of ulnar neuropathy before operation was 19%. After operation 20 patients (33%) had an ulnar neuropathy, in seven of which it had been present before operation, and of these weakness of the hypothenar muscle occurred in two. The probability of survival of the Souter-Strathclyde elbow prosthesis based on the Kaplan-Meier calculation is 69% at ten years


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1460 - 1463
1 Nov 2006
Landor I Vavrik P Jahoda D Guttler K Sosna A

We assessed the long-term results of 58 Souter-Strathclyde total elbow replacements in 49 patients with rheumatoid arthritis. The mean length of follow-up was 9.5 years (0.7 to 16.7). The mean pre-operative Mayo Elbow Performance Score was 30 (15 to 80) and at final follow-up was 82 (60 to 95). A total of 13 elbows (22.4%) were revised, ten (17.2%) for aseptic loosening, one (1.7%) for instability, one (1.7%) for secondary loosening after fracture, and one elbow (1.7%) was removed because of deep infection. The Kaplan-Meier survival rate was 70% and 53% at ten and 16 years, respectively. Failure of the ulnar component was found to be the main problem in relation to the loosening. Anterior transposition of the ulnar nerve had no influence on ulnar nerve paresthaesiae in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 191 - 195
1 Feb 2005
Mighell MA Dunham RC Rommel EA Frankle MA

We present six patients with chronic dislocation of the elbow who were treated by primary semiconstrained total elbow arthroplasty. All were women with a mean age of 65 years (51 to 76), the mean interval between dislocation and surgery was 17 weeks (5 to 52) and the mean follow-up 58 months (24 to 123).

The most dramatic improvement was in function. The mean American Shoulder and Elbow Surgeon score was 5.2 times better (p < 0.001) and the mean total range of movement increased from 33° to 121° (p < 0.001) after operation. Three patients developed wear of polyethylene. One required revision for a periprosthetic fracture, and another required a bushing exchange.

Primary semiconstrained elbow arthroplasty provides significant, predictable functional improvement. Potential solutions for wear of polyethylene include a different operative technique or design of implant. Despite the high incidence of such wear, total elbow arthroplasty should be considered as a viable treatment option for chronic dislocation of the elbow in elderly patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 661 - 667
1 May 2010
van Riet RP Sanchez-Sotelo J Morrey BF

There is little information available at present regarding the mechanisms of failure of modern metallic radial head implants. Between 1998 and 2008, 44 consecutive patients (47 elbows) underwent removal of a failed metallic radial head replacement. In 13 patients (13 elbows) the initial operation had been undertaken within one week of a fracture of the radial head, at one to six weeks in seven patients (seven elbows) and more than six weeks (mean of 2.5 years (2 to 65 months)) in 22 patients (25 elbows). In the remaining two elbows the replacement was inserted for non-traumatic reasons. The most common indication for further surgery was painful loosening (31 elbows). Revision was undertaken for stiffness in 18 elbows, instability in nine, and deep infection in two. There were signs of over-lengthening of the radius in 11 elbows. Degenerative changes were found in all but one. Only three loose implants had been fixed with cement. Instability was not identified in any of the bipolar implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 486 - 489
1 Apr 2007
Bassi RS Simmons D Ali F Nuttall D Birch A Trail IA Stanley JK

The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function.

No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 62 - 65
1 Jan 2007
Ito H Matsumoto T Yoshitomi H Kakinoki R Nakamura T

We compared the outcome of peri-operative humeral condylar fractures in patients undergoing a Coonrad-Morrey semiconstrained total elbow replacement with that of patients with rheumatoid arthritis undergoing the same procedure without fractures. In a consecutive series of 40 elbows in 33 patients, 13 elbows had a fracture in either condyle peri-operatively, and 27 elbows were intact. The fractured condyle was either fixed internally or excised. We found no statistical difference in the patients’ background, such as age, length of follow-up, immobilisation period, Larsen’s radiological grade, or Steinbrocker’s stage and functional class. There was also no statistical difference between the groups in relation to the Mayo Elbow Performance Score, muscle strength, range of movement, or radiolucency around the implants at a mean of 4.8 years (1.1 to 8.0) follow-up.

We conclude that fractured condyles can be successfully treated with either internal fixation or excision, and cause no harmful effect.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1198 - 1204
1 Sep 2008
Peden JP Morrey BF

This study reports our experience with total elbow replacement for fused elbows.

Between 1982 and 2004, 13 patients with spontaneously ankylosed elbows were treated with a linked semi-constrained non-custom total elbow implant. The mean age at operation was 54 years (24 to 80). The stiffness was a result of trauma in ten elbows, juvenile rheumatoid arthritis in one, and rheumatoid arthritis in two. The patients were followed for a mean of 12 years (2 to 26) and were evaluated clinically using the Mayo Elbow Performance Score, as well as radiologically.

A mean arc from 37° of extension to 118° of flexion was achieved. Outcomes were good or excellent for seven elbows at final review. Ten patients felt better or much better after total elbow replacement. However, there was a high complication rate and re-operation was required in over half of patients. Two developed peri-operative soft-tissue breakdown requiring debridement. A muscle flap with skin grafting was used for soft-tissue cover in one. Revision was undertaken in one elbow following fracture of the ulnar component. Three patients developed a deep infection. Three elbows were manipulated under anaesthesia for post-operative stiffness. Prophylactic measures for heterotopic ossification were unsuccessful.

Total elbow replacement for the ankylosed elbow should be performed with caution. However, the outcome can be reliable in the long term and have a markedly positive impact on patient function and satisfaction. The high potential for complications must be considered. We consider total elbow replacement to be an acceptable procedure in selected patients with reasonable expectations.