Tapered, fluted, modular, titanium stems are
increasingly popular in the operative management of Vancouver B2 and
selected B3 peri-prosthetic femoral fractures. We have reviewed
the results at our institution looking at stem survival and clinical
outcomes and compared this with reported outcomes in the literature.
Stem survival at a mean of 54 months was 96% in our series and 97%
for combined published cases. Review of radiology showed maintenance
or improvement of bone stock in 89% of cases with high rates of
femoral union. Favourable clinical outcome scores have reported
by several authors. No difference in survival or clinical scores
was observed between B2 and B3 fractures. Tapered stems are a useful
option in revision for femoral fracture across the spectrum of femoral
bone deficiency. Cite this article:
The use of allograft struts and cerclage wire, possibly augmented by plate fixation, for the treatment of Vancouver type-B1
In this retrospective study, we investigated
the results of revision total hip replacement (THR) using a cemented long-stemmed
Exeter femoral component, with a minimum length of 205 mm in patients
with extensive femoral bone defects. The study included 37 consecutive
patients with a mean age of 76 years (39 to 93) and a mean follow-up
of nine years (5 to 16). A total of 26 patients (70%) had a pre-operative
Endo-Klinik score of 3 or 4. Impaction bone grafting was used in
24 patients (65%). At the time of evaluation, 22 patients (59%)
were still alive and were evaluated clinically and radiologically.
A total of 14 patients died during follow-up and their data were
included until the time of their death. One reconstruction failed
after five years and five months owing to recurrent dislocation:
the hip was converted to an excision arthroplasty. Intra-operative
fractures or fissures were encountered in nine patients (24%), but
none occurred during impaction of the bone graft. Post-operative
peri-prosthetic fractures occurred in two patients (5%); both were
treated with plate fixation. At nine years, survival with the endpoint
of all-cause re-revision was 96.3% (95% CI 76.4 to 99.5); including
re-operations for any reason, it was 80.7% (95% CI 56.3 to 92.3%).
There were no re-revisions for aseptic loosening. The survival of long stem cemented femoral components following
revision THR is satisfactory in a fragile population with extensive
femoral defects. Cite this article:
We determined the midterm survival, incidence
of
Revision arthroplasty of the hip is expensive
owing to the increased cost of pre-operative investigations, surgical implants
and instrumentation, protracted hospital stay and drugs. We compared
the costs of performing this surgery for aseptic loosening, dislocation,
deep infection and
The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and
An 81-year-old woman presented with a fracture
in the left femur. She had well-fixed bilateral hip replacements
and had received long-term bisphosphonate treatment. Prolonged bisphosphonate
use has been recently linked with atypical subtrochanteric and diaphyseal
femoral fractures. While the current definition of an atypical fracture
of the femur excludes
Revision total hip arthroplasty (THA) is challenging
when there is severe loss of bone in the proximal femur. The purpose
of this study was to evaluate the clinical and radiographic outcomes
of revision THA in patients with severe proximal femoral bone loss
treated with a fluted, tapered, modular femoral component. Between
January 1998 and December 2004, 92 revision THAs were performed
in 92 patients using a single fluted, tapered, modular femoral stem
design. Pre-operative diagnoses included aseptic loosening, infection
and
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs.
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. . Four patients with late complications (14%) have since undergone
surgery (two for a
We present our experience with a double-mobility
acetabular component in 155 consecutive revision total hip replacements
in 149 patients undertaken between 2005 and 2009, with particular
emphasis on the incidence of further dislocation. The mean age of
the patients was 77 years (42 to 89) with 59 males and 90 females.
In all, five patients died and seven were lost to follow-up. Indications
for revision were aseptic loosening in 113 hips, recurrent instability
in 29,
We reviewed 44 consecutive revision hip replacements in 38 patients performed using the cement-in-cement technique. All were performed for acetabular loosening in the presence of a well-fixed femoral component. The mean follow-up was 5.1 years (2 to 10.1). Radiological analysis at final follow-up indicated no loosening of the femoral component, except for one case with a continuous radiolucent line in all zones and
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) vs 5.8%
(6 of 104), p = 0.569) or re-operation rates without revision (5.1%
(5 of 99) vs 2.9% (3 of 104), p = 0.428). In the
THR group one recurrent dislocation, two late deep infections and
one
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for
If a surgeon is faced with altered lesser trochanter
anatomy when revising the femoral component in revision total hip
replacement, a
The ideal bearing surface for young patients
undergoing total hip replacement (THR) remains controversial. We report
the five-year results of a randomised controlled trial comparing
the clinical and radiological outcomes of 102 THRs in 91 patients
who were <
65 years of age. These patients were randomised to
receive a cobalt–chrome on ultra-high-molecular-weight polyethylene,
cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic
bearing. In all, 97 hip replacements in 87 patients were available
for review at five years. Two hips had been revised, one for infection
and one for
We report the clinical and radiographic outcomes
of 208 consecutive femoral revision arthroplasties performed in 202
patients (119 women, 83 men) between March 1991 and December 2007
using the X-change Femoral Revision System, fresh-frozen morcellised
allograft and a cemented polished Exeter stem. All patients were
followed prospectively. The mean age of the patients at revision
was 65 years (30 to 86). At final review in December 2013 a total
of 130 patients with 135 reconstructions (64.9%) were alive and
had a non re-revised femoral component after a mean follow-up of
10.6 years (4.7 to 20.9). One patient was lost to follow-up at six
years, and their data were included up to this point.
Re-operation for any reason was performed in 33 hips (15.9%), in
13 of which the femoral component was re-revised (6.3%). The mean
pre-operative Harris hip score was 52 (19 to 95) (n = 73) and improved
to 80 (22 to 100) (n = 161) by the last follow-up. Kaplan–Meier
survival with femoral re-revision for any reason as the endpoint
was 94.9% (95% confidence intervals (CI) 90.2 to 97.4) at ten years;
with femoral re-revision for aseptic loosening as the endpoint it was
99.4% (95% CI 95.7 to 99.9); with femoral re-operation for any reason
as the endpoint it was 84.5% (95% CI 78.3 to 89.1); and with subsidence ≥ 5
mm it was 87.3% (95% CI 80.5 to 91.8). Femoral revision with the
use of impaction allograft bone grafting and a cemented polished
stem results in a satisfying survival rate at a mean of ten years’ follow-up. Cite this article:
The Unified Classification System (UCS) emphasises
the key principles in the assessment and management of peri-prosthetic
fractures complicating partial or total joint replacement. We tested the inter- and intra-observer agreement for the UCS
as applied to the pelvis and femur using 20 examples of peri-prosthetic
fracture in 17 patients. Each subtype of the UCS was represented
by at least one case. Specialist orthopaedic surgeons (experts)
and orthopaedic residents (pre-experts) assessed reliability on
two separate occasions. For the pelvis, the UCS showed inter-observer agreement of 0.837
(95% confidence intervals (CI) 0.798 to 0.876) for the experts and
0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer
agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and
0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the
UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765
to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773)
for the pre-experts. The intra-observer agreement was 0.920 (95%
CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892)
for the pre-experts. This corresponds to a substantial and ‘almost
perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic
fractures of the pelvis and femur. We hope that unifying the terminology of these injuries will
assist in their assessment, treatment and outcome. Cite this article:
Data on early morbidity and complications after
revision total hip replacement (THR) are limited. The aim of this nationwide
study was to describe and quantify early morbidity after aseptic
revision THR and relate the morbidity to the extent of the revision
surgical procedure. We analysed all aseptic revision THRs from 1st
October 2009 to 30th September 2011 using the Danish National Patient
Registry, with additional information from the Danish Hip Arthroplasty
Registry. There were 1553 procedures (1490 patients) performed in
40 centres and we divided them into total revisions, acetabular
component revisions, femoral stem revisions and partial revisions.
The mean age of the patients was 70.4 years (25 to 98) and the median
hospital stay was five days (interquartile range 3 to 7). Within 90
days of surgery, the readmission rate was 18.3%, mortality rate
1.4%, re-operation rate 6.1%, dislocation rate 7.0% and infection
rate 3.0%. There were no differences in these outcomes between high-
and low-volume centres. Of all readmissions, 255 (63.9%) were due
to ‘surgical’ complications Cite this article:
The outcome of 219 revision total hip arthroplasties
(THAs) in 98 male and 121 female patients, using 137 long length
and 82 standard length cemented collarless double-taper femoral
stems in 211 patients, with a mean age of 72 years (30 to 90) and
mean follow-up of six years (two to 18) have been described previously.
We have extended the follow-up to a mean of 13 years (8 to 20) in
this cohort of patients in which the pre-operative bone deficiency Paprosky
grading was IIIA or worse in 79% and 73% of femurs with long and
standard stems, respectively. For the long stem revision group, survival to re-revision for
aseptic loosening at 14 years was 97% (95% confidence interval (CI)
91 to 100) and in patients aged >
70 years, survival was 100%. Two
patients (two revisions) were lost to follow-up and 86 patients
with 88 revisions had died. Worst-case analysis for survival to
re-revision for aseptic loosening at 14 years was 95% (95% CI 89
to 100) and 99% (95% CI 96 to 100) for patients aged >
70 years. One
additional long stem was classified as loose radiographically but
not revised. For the standard stem revision group, survival to re-revision
for aseptic loosening at 14 years was 91% (95% CI 83 to 99). No
patients were lost to follow-up and 49 patients with 51 hips had
died. No additional stems were classified as loose radiographically. Femoral revision using a cemented collarless double-taper stem,
particularly with a long length stem, and in patients aged >
70
years, continues to yield excellent results up to 20 years post-operatively,
including in hips with considerable femoral metaphyseal bone loss. Cite this article:
The optimal management of intracapsular fractures of the femoral
neck in independently mobile patients remains open to debate. Successful
fixation obviates the limitations of arthroplasty for this group
of patients. However, with fixation failure rates as high as 30%,
the outcome of revision surgery to salvage total hip arthroplasty
(THA) must be considered. We carried out a systematic review to
compare the outcomes of salvage THA and primary THA for intracapsular
fractures of the femoral neck. We performed a Preferred Reporting Items for Systematic Reviews
and Meta-Analysis (PRISMA) compliant systematic review, using the
PubMed, EMBASE and Cochrane libraries databases. A meta-analysis
was performed where possible, and a narrative synthesis when a meta-analysis
was not possible.Aims
Patients and Methods
The early failure and revision of bimodular primary
total hip arthroplasty prostheses requires the identification of the
risk factors for material loss and wear at the taper junctions through
taper wear analysis. Deviations in taper geometries between revised
and pristine modular neck tapers were determined using high resolution
tactile measurements. A new algorithm was developed and validated
to allow the quantitative analysis of material loss, complementing
the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals ( Cite this article:
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
Many different lengths of stem are available
for use in primary total hip replacement, and the morphology of
the proximal femur varies greatly. The more recently developed shortened
stems provide a distribution of stress which closely mimics that
of the native femur. Shortening the femoral component potentially
comes at the cost of decreased initial stability. Clinical studies
on the performance of shortened cemented and cementless stems are promising,
although long-term follow-up studies are lacking. We provide an
overview of the current literature on the anatomical features of
the proximal femur and the biomechanical aspects and clinical outcomes
associated with the length of the femoral component in primary hip
replacement, and suggest a classification system for the length
of femoral stems. Cite this article:
The treatment of substantial proximal femoral
bone loss in young patients with developmental dysplasia of the
hip (DDH) is challenging. We retrospectively analysed the outcome
of 28 patients (30 hips) with DDH who underwent revision total hip
replacement (THR) in the presence of a deficient proximal femur,
which was reconstructed with an allograft prosthetic composite.
The mean follow-up was 15 years (8.5 to 25.5). The mean number of
previous THRs was three (1 to 8). The mean age at primary THR and
at the index reconstruction was 41 years (18 to 61) and 58.1 years
(32 to 72), respectively. The indication for revision included mechanical
loosening in 24 hips, infection in three and peri-prosthetic fracture
in three. Six patients required removal and replacement of the allograft
prosthetic composite, five for mechanical loosening and one for
infection. The survivorship at ten, 15 and 20 years was 93% (95%
confidence interval (CI) 91 to 100), 75.5% (95% CI 60 to 95) and
75.5% (95% CI 60 to 95), respectively, with 25, eight, and four
patients at risk, respectively. Additionally, two junctional nonunions
between the allograft and host femur required bone grafting and
plating. An allograft prosthetic composite affords a good long-term outcome
in the management of proximal femoral bone loss in revision THR
in patients with DDH, while preserving distal host bone.
We investigated the incidence of soft-tissue
lesions after small head metal-on-metal total hip replacement (MoM THR).
Between December 1993 and May 1999, 149 patients (195 hips) underwent
primary cementless MoM THR. During the follow-up period, three patients (five THRs) died
and eight patients (14 THRs) were lost to follow-up. We requested
that all patients undergo CT evaluation. After exclusion of five
patients (six THRs) who had undergone a revision procedure, and
22 (28 THRs) who were unwilling to take part in this study, 111
patients (142 THRs) were evaluated. There were 63 men (88 THRs)
and 48 women (54 THRs) with a mean age of 45.7 years (37 to 56)
at the time of surgery. The mean follow-up was 15.4 years (13 to
19). A soft-tissue lesion was defined as an abnormal peri-prosthetic
collection of fluid, solid lesion or asymmetrical soft-tissue mass. At final follow-up, soft-tissue lesions were found in relation
to 28 THRs (19.7%), including 25 solid and three cystic lesions.
They were found in 20 men and eight women; 26 lesions were asymptomatic
and two were symptomatic. The mean maximal diameter of the soft-tissue
lesion was 42.3 mm (17 to 135). The relatively high rate of soft-tissue lesions
observed with small head MoM THR remains a concern. Cite this article:
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of ceramic fracture and dislocation between
28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men
and 58 women, mean age 49.2 years) were randomised to THR with either
a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned
to the 28 mm group and 55/60 hips assigned to the 32 mm group were
followed for at least five years. No ceramic component fractures
occured in any patient in either group. There was one dislocation
in the 32 mm group and none in the 28 mm group (p = 0.464). No hip
had detectable wear, focal osteolysis or prosthetic loosening. In
our small study the 32 mm ceramic articulation appeared to be safe
in terms of ceramic liner fracture. Cite this article:
Until now, there has been no consensus as to whether stemmed
acetabular components are appropriate for use in patients undergoing
revision total hip arthroplasty (THA) who have major acetabular
defects or pelvic discontinuity. We wished to address this deficiency
in the literature. We carried out a retrospective study of 35 patients (six men
and 29 women) with a mean age of 68 years (37 to 87), with major
acetabular defects who underwent revision THA between 2000 and 2012.Aim
Patients and Methods
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved revision stem with distal cone-in-cone fixation. A successful
outcome depends on careful pre-operative planning and the use of
a transfemoral approach when the anatomy is distorted or a fracture
is imminent, or residual cement or a partially-secured existing
stem cannot be removed. The shortest appropriate stem should, in
our opinion, be used and secured with >
3 cm fixation at the femoral
isthmus, and distal interlocking screws should be used for additional
stability when this goal cannot be realised. Cite this article:
Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article:
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
The purpose of this prospective study was to
evaluate the long-term clinical and radiological outcomes of revision of
the femoral component of a total hip replacement using impaction
bone grafting. Femoral revision with an impacted allograft was performed
on 29 patients (31 hips). In all, 21 hips (68%) had grade III or
IV femoral defects according to the Endo-Klinik classification.
A total of 11 patients (12 hips) died before the ten-year follow-up
period. Of the remaining patients, 18 patients (19 hips) were followed
for 10 to 15 years; three further patients died during this time.
None of the 31 stems underwent further revision of their stem. However,
four stems showed extensive subsidence (>
15 mm). One of these patients
had a femoral fracture that required fixation. Three other patients
had a femoral fracture, two of which required fixation and the other
was treated conservatively. Patients with a femoral fracture and/or
severe subsidence had significantly more grade IV defects (six of
seven hips; p = 0.004). One patient needed a closed reduction for
dislocation. Impaction allografting in revision hip surgery gives good long-term
results for femora with grades I, II and III Endo-Klinik-classified
defects. Extensive subsidence and femoral fractures were seen mainly
in patients with grade IV damaged femora.
We performed a case–control study to compare
the rates of further surgery, revision and complications, operating time
and survival in patients who were treated with either an uncemented
hydroxyapatite-coated Corail bipolar femoral stem or a cemented
Exeter stem for a displaced intracapsular fracture of the hip. The
mean age of the patients in the uncemented group was 82.5 years
(53 to 97) and in the cemented group was 82.7 years (51 to 99) We used
propensity score matching, adjusting for age, gender and the presence
or absence of dementia and comorbidities, to produce a matched cohort
receiving an Exeter stem (n = 69) with which to compare the outcome of
patients receiving a Corail stem (n = 69). The Corail had a significantly
lower all-cause rate of further surgery (p = 0.016; odds ratio (OR)
0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further
surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating
time was significantly less for the Corail group than for the cemented Exeter
group (59 min [12 to 136] Cite this article:
The Vancouver classification has been shown by its developers to be a valid and reliable method for categorising the configuration of periprosthetic proximal femoral fractures and for planning their management. We have re-validated this classification system independently using the radiographs of 30 patients with periprosthetic fractures. These were reviewed by six experienced consultant orthopaedic surgeons, six trainee surgeons and six medical students in order to assess intra- and interobserver reliability and reproducibility. Each observer read the radiographs on two separate occasions. The results were subjected to weighted kappa statistical analysis. The respective kappa values for interobserver agreement were 0.72 and 0.74 for consultants, 0.68 and 0.70 for trainees on the first and second readings of the radiographs and 0.61 for medical students. The intra-observer agreement for the consultants was 0.64 and 0.67, for the trainees 0.61 and 0.64, and for the medical students 0.59 and 0.60 for the first and second readings, respectively. The validity of the classification was studied by comparing the pre-operative radiological findings within B subgroups with the operative findings. This revealed agreement for 77% of these type-B fractures, with a kappa value of 0.67. Our data confirm the reliability and reproducibility of this classification system in a European setting and for inexperienced staff. This is a reliable system which can be used by non-experts, between centres and across continents.
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some brands in specific subgroups,
but high failure rates and local soft-tissue reactions in others.
The National Joint Registry for England and Wales (NJR) has collected
data on all hip resurfacings performed since 2003. This retrospective
cohort study recorded survival time to revision from a resurfacing
procedure, exploring risk factors independently associated with
failure. All patients with a primary diagnosis of osteoarthritis
who underwent resurfacing between 2003 and 2010 were included in
the analyses. Cox’s proportional hazard models were used to analyse
the extent to which the risk of revision was related to patient,
surgeon and implant covariates. A total of 27 971 hip resurfacings were performed during the
study period, of which 1003 (3.59%) underwent revision surgery.
In the final adjusted model, we found that women were at greater
risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007),
but the risk of revision was independent of age. Of the implant-specific
predictors, five brands had a significantly greater risk of revision
than the Birmingham Hip Resurfacing (BHR) (ASR: HR = 2.82, p <
0.001,
Conserve: HR = 2.03, p <
0.001, Cormet: HR = 1.43, p = 0.001,
Durom: HR = 1.67, p <
0.001, Recap: HR = 1.58, p = 0.007). Smaller
femoral head components were also significantly more likely to require
revision (≤ 44 mm: HR = 2.14, p <
0.001, 45 to 47 mm: HR = 1.48,
p = 0.001) than medium or large heads, as were operations performed
by low-volume surgeons (HR = 1.36, p <
0.001). Once these influences
had been removed, in 4873 male patients <
60 years old undergoing
resurfacing with a BHR, the five-year estimated risk of revision
was 1.59%. In summary, after adjustment for a range of covariates we found
that there were significant differences in the rate of failure between
brands and component sizes. Younger male patients had good five-year
implant survival when the BHR was used.
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
We have managed 27 patients (16 women and 11 men) with a mean age of 68.4 years (50 to 84), with failed total hip replacement and severe proximal femoral bone loss by revision using a distal fix/proximal wrap prosthesis. The mean follow-up was for 55.3 months (25 to 126). The mean number of previous operations was 2.2 (1 to 4). The mean Oxford hip score decreased from 46.2 (38 to 60) to 28.5 (17 to 42) (paired t-test, p <
0.001) and the mean Harris Hip score increased from 30.4 (3 to 57.7) to 71.7 (44 to 99.7) (paired t-test, p <
0.001). There were two dislocations, and in three patients we failed to eradicate previous infection. None required revision of the femoral stem. This technique allows instant distal fixation while promoting biological integration and restoration of bone stock. In the short term, the functional outcome is encouraging and the complication rates acceptable in this difficult group of patients.
Tranexamic acid (TXA), an inhibitor of fibrinolysis,
reduces blood loss after total knee arthroplasty. However, its effect
on minimally invasive total hip arthroplasty (THA) is not clear.
We performed a prospective, randomised double-blind study to evaluate
the effect of two intravenous injections of TXA on blood loss in
patients undergoing minimally invasive THA. In total, 60 patients (35 women and 25 men with a mean age of
58.1 years; 17 to 84) who underwent unilateral minimally invasive
uncemented THA were randomly divided into the study group (30 patients,
20 women and ten men with a mean age of 56.5 years; 17 to 79) that
received two intravenous injections 1 g of TXA pre- and post-operatively
(TXA group), and a placebo group (30 patients, 15 women and 15 men
with a mean age of 59.5 years; 23 to 84). We compared the peri-operative
blood loss of the two groups. Actual blood loss was calculated from
the maximum reduction in the level of haemoglobin. All patients
were followed clinically for the presence of venous thromboembolism. The TXA group had a lower mean intra-operative blood loss of
441 ml (150 to 800) This prospective, randomised controlled study showed that a regimen
of two intravenous injections of 1 g TXA is effective for blood
conservation after minimally invasive THA. Cite this article:
Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
Despite excellent results, the use of cemented
total hip replacement (THR) is declining. This retrospective cohort study
records survival time to revision following primary cemented THR
using the most common combination of components that accounted for
almost a quarter of all cemented THRs, exploring risk factors independently associated
with failure. All patients with osteoarthritis who had an Exeter
V40/Contemporary THR (Stryker) implanted before 31 December 2010
and recorded in the National Joint Registry for England and Wales
were included in the analysis. Cox’s proportional hazard models
were used to analyse the extent to which risk of revision was related
to patient, surgeon and implant covariates, with a significance
threshold of p <
0.01. A total of 34 721 THRs were included in
the study. The overall seven-year rate of revision for any reason
was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final
adjusted model the risk of revision was significantly higher in
THRs with the Contemporary hooded component (hazard ratio (HR) 1.88,
p <
0.001) than with the flanged version, and in smaller head
sizes (<
28 mm) compared with 28 mm diameter heads (HR 1.50,
p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69
to 1.63) with a 28 mm diameter head and flanged component. The overall
risk of revision was independent of age, gender, American Society
of Anesthesiologists grade, body mass index, surgeon volume, surgical
approach, brand of cement/presence of antibiotic, femoral head material
(stainless steel/alumina) and stem taper size/offset. However, the
risk of revision for dislocation was significantly higher with a
‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component
(HR 2.34, p <
0.001). In summary, we found that there were significant differences
in failure between different designs of acetabular component and
sizes of femoral head after adjustment for a range of covariates.
We compared the clinical and radiological outcomes
of two cementless femoral stems in the treatment of patients with
a Garden III or IV fracture of the femoral neck. A total of 70 patients At final follow-up there were no statistically significant differences
between the short anatomical and the conventional stems with regard
to the mean Harris hip score (85.7 (66 to 100) Our study demonstrated that despite the poor bone quality in
these elderly patients with a fracture of the femoral neck, osseo-integration
was obtained in all hips in both groups. However, the incidence
of thigh pain, pulmonary microemboli and peri-prosthetic fracture
was significantly higher in the conventional stem group than in
the short stem group.
We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
We treated 34 patients with recurrent dislocation of the hip with a constrained acetabular component. Roentgen stereophotogrammetric analysis was performed to assess migration of the prosthesis. The mean clinical follow-up was 3.0 years (2.2 to 4.8) and the radiological follow-up was 2.7 years (2.0 to 4.8). At the latest review six patients had died and none was lost to follow-up. There were four acetabular revisions, three for aseptic loosening and one for deep infection. Another acetabular component was radiologically loose with progressive radiolucent lines in all Gruen zones and was awaiting revision. The overall rate of aseptic loosening was 11.8% (4 of 34). Roentgen stereophotogrammetric analysis in the non-revised components confirmed migration of up to 1.06 mm of translation and 2.32° of rotation at 24 months. There was one case of dislocation and dissociation of the component in the same patient. Of the 34 patients, 33 (97.1%) had no further episodes of dislocation. The constrained acetabular component reported in our study was effective in all but one patient with instability of the hip, but the rate of aseptic loosening was higher than has been reported previously and requires further investigation.
Revision total hip replacement (THR) for young
patients is challenging because of technical complexity and the potential
need for subsequent further revisions. We have assessed the survivorship,
functional outcome and complications of this procedure in patients
aged <
50 years through a large longitudinal series with consistent treatment
algorithms. Of 132 consecutive patients (181 hips) who underwent
revision THR, 102 patients (151 hips) with a mean age of 43 years
(22 to 50) were reviewed at a mean follow-up of 11 years (2 to 26)
post-operatively. We attempted to restore bone stock with allograft
where indicated. Using further revision for any reason as an end point,
the survival of the acetabular component was 71% ( This overall perspective on the mid- to long-term results is
valuable when advising young patients on the prospects of revision
surgery at the time of primary replacement. Cite this article:
An atypical femoral fracture (AFF), with a transverse
fracture radiologically through the lateral cortex is a rare but serious
condition. In order to improve our ability to identify patients
with this condition, we retrospectively surveyed all subtrochanteric,
peri-implant and diaphyseal femoral fractures in patients aged ≥ 65
years who underwent surgical treatment at our hospital between 2004
and 2011. We describe the incidence of atypical fractures and their characteristics,
with observational data including a review of the hospital and general
practitioner records. Clinical outcomes were evaluated using the
Harris hip score (HHS) and the timed up-and-go (TUG) test. Atypical fractures only occurred in women with an incidence of
9.8 per 100 000 person-years. The incidence in those who were treated
with bisphosphonates was 79.0 per 100 000 person-years; eight of
17 fractures occurred around metal implants. There was a high incidence
of delayed union and revision surgery. A total of nine patients (ten
AFFs) were available for review at a mean follow-up of 36.5 months
(10 to 104). The clinical outcome was poor with a mean HHS of 58.9
(95% CI 47.4 to 70.4) and a mean TUG test of 25.7 s (95% CI 12.7
to 38.8). The delay in diagnosis and treatment of AFF may result from a
lack of knowledge of this condition. Cite this article:
In March 2012, an algorithm for the treatment
of intertrochanteric fractures of the hip was introduced in our academic
department of Orthopaedic Surgery. It included the use of specified
implants for particular patterns of fracture. In this cohort study,
102 consecutive patients presenting with an intertrochanteric fracture
were followed prospectively (post-algorithm group). Another 117
consecutive patients who had been treated immediately prior to the
implementation of the algorithm were identified retrospectively
as a control group (pre-algorithm group). The total cost of the
implants prior to implementation of the algorithm was $357 457 (mean:
$3055 (1947 to 4133)); compared with $255 120 (mean: $2501 (1052
to 4133)) after its implementation. There was a trend toward fewer complications
in patients who were treated using the algorithm (33% pre- The implementation of an evidence-based algorithm for the treatment
of intertrochanteric fractures reduced costs while maintaining quality
of care with a lower rate of complications and re-admissions. Cite this article:
Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge. We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108). The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no mechanical failures, no new infections and no re-operations. We suggest that in highly selected cases, in which conventional fixation is not feasible, this technique offers a durable option and avoids the morbidity of a total femoral replacement.
Iontophoresis is a novel technique which may be used to facilitate the movement of antibiotics into the substance of bone using an electrical potential applied externally. We have examined the rate of early infection in allografts following application of this technique in clinical practice. A total of 31 patients undergoing revision arthroplasty or surgery for limb salvage received 34 iontophoresed sequential allografts, of which 26 survived for a minimum of two years. The mean serum antibiotic levels after operation were low (gentamicin 0.37 mg/l (0.2 to 0.5); flucloxacillin 1 mg/l (0 to 1) and the levels in the drains were high (gentamicin 40 mg/l (2.5 to 131); flucloxacillin 17 mg/l (1 to 43). There were no early deep infections. Two late infections were presumed to be haemotogenous; 28 of the 34 allografts were retained. In 12 patients with pre-existing proven infection further infection has not occurred at a mean follow-up of 51 months (24 to 82).
Femoral revision after cemented total hip replacement
(THR) might include technical difficulties, following essential cement
removal, which might lead to further loss of bone and consequently
inadequate fixation of the subsequent revision stem. Femoral impaction allografting has been widely used in revision
surgery for the acetabulum, and subsequently for the femur. In combination
with a primary cemented stem, impaction grafting allows for femoral
bone restoration through incorporation and remodelling of the impacted
morsellized bone graft by the host skeleton. Cavitary bone defects
affecting meta-physis and diaphysis leading to a wide femoral shaft,
are ideal indications for this technique. Cancellous allograft bone
chips of 1 mm to 2 mm size are used, and tapered into the canal
with rods of increasing diameters. To impact the bone chips into
the femoral canal a prosthesis dummy of the same dimensions of the definitive
cemented stem is driven into the femur to ensure that the chips
are very firmly impacted. Finally, a standard stem is cemented into
the neo-medullary canal using bone cement. To date several studies have shown favourable results with this
technique, with some excellent long-term results reported in independent
clinical centres worldwide. Cite this article:
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
Juvenile idiopathic arthritis (JIA) is a chronic disease of childhood; it causes joint damage which may require surgical intervention, often in the young adult. The aim of this study was to describe the long-term outcome and survival of hip replacement in a group of adult patients with JIA and to determine predictors of survival for the prosthesis. In this retrospective comparative study patients were identified from the database of a regional specialist adult JIA clinic. This documented a series of 47 hip replacements performed in 25 adult patients with JIA. Surgery was performed at a mean age of 27 years (11 to 47), with a mean follow-up of 19 years (2 to 36). The mean Western Ontario and McMaster Universities osteoarthritis index questionnaire (WOMAC) score at the last follow-up was 53 (19 to 96) and the mean Health Assessment Questionnaire score was 2.25 (0 to 3). The mean pain component of the WOMAC score (60 (20 to 100)) was significantly higher than the mean functional component score (46 (0 to 97)) (p = 0.02). Kaplan-Meier survival analysis revealed a survival probability of 46.6% (95% confidence interval 37.5 to 55.7) at 19 years, with a trend towards enhanced survival with the use of a cemented acetabular component and a cementless femoral component. This was not, however, statistically significant (acetabular component, p = 0.76, femoral component, p = 0.45). Cox’s proportional hazards regression analysis showed an implant survival rate of 54.9% at 19 years at the mean of covariates. Survival of the prosthesis was significantly poorer (p = 0.001) in patients who had been taking long-term corticosteroids and significantly better (p = 0.02) in patients on methotrexate.
Total hip replacement (THR) still is a rare intervention
in many African countries. In Burkina Faso it is not performed on
a regular basis. A visiting programme for THR was started in a district
hospital with no previous relevant experience. In this paper we
present an analysis of the surgical technical problems and peri-operative
complications of 152 THRs in 136 patients and three bipolar hemiarthroplasties
in three patients undertaken in this new programme with limited
orthopaedic equipment. There were 86 male and 53 female patients
with a mean age of 49 years (21 to 78). We identified 77 intra-operative
technical problems in 51 operations. There were 24 peri-operative complications
in 21 patients, 17 of which were bony in nature. So far, ten revision
THRs have been performed in nine patients. Regular analysis of the technical problems and complications
was used to improve quality, and we identified patient selection
adapted to the local circumstances as important to avoid complications.
Our reflections on the problems encountered in initiating such a
programme may be of help to other teams planning similar projects. Cite this article:
We undertook a randomised prospective follow-up study of changes in peri-prosthetic bone mineral density (BMD) after hip resurfacing and compared them with the results after total hip replacement. A total of 59 patients were allocated to receive a hip resurfacing (n = 29) or an uncemented distally fixed total hip replacement (n = 30). The BMD was prospectively determined in four separate regions of interest of the femoral neck and in the calcar region corresponding to Gruen zone 7 for the hip resurfacing group and compared only to the calcar region in the total hip replacement group. Standardised measurements were performed pre-operatively and after three, six and 12 months. The groups were well matched in terms of gender distribution and mean age. The mean BMD in the calcar region increased after one year to 105.2% of baseline levels in the resurfaced group compared with a significant decrease to 82.1% in the total hip replacement group (p <
0.001) by 12 months. For the resurfaced group, there was a decrease in bone density in all four regions of the femoral neck at three months which did not reach statistical significance and was followed by recovery to baseline levels after 12 months. Hip resurfacing did indeed preserve BMD in the inferior femoral neck. In contrast, a decrease in the mean BMD in Gruen zone 7 followed uncemented distally fixed total hip replacement. Long term follow-up studies are necessary to see whether this benefit in preservation of BMD will be clinically relevant at future revision surgery.
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and offset, shell, liner and stem type, and surgeon characteristics. We found significant differences in failure between bearing surfaces
and femoral stem size after adjustment for a range of covariates
in a large cohort of single-brand cementless THRs. In this study
of procedures performed since 2003, hard bearings had significantly
higher rates of revision, but we found no evidence that head size
had an effect. Patient characteristics, such as BMI and American
Society of Anesthesiologists grade, also influence the survival
of cementless components. Cite this article:
Using the Trent regional arthroplasty register, we analysed the survival at ten years of 1198 consecutive Charnley total hip replacements carried out across a single health region of the United Kingdom in 1990. At ten years, information regarding outcome was available for 1001 hips (83.6%). The crude revision rate was 6.2% (62 of 1001) and the cumulative survival rate with revision of the components as an end-point was 93.1%. At five years, a review of this series of patients identified gross radiological failure in 25 total hip replacements which had previously been unrecognised. At ten years the outcome was known for 18 of these 25 patients (72%), of whom 13 had not undergone revision. This is the first study to assess the survival at ten years for the primary Charnley total hip replacement performed in a broad cross-section of hospitals in the United Kingdom, as opposed to specialist centres. Our results highlight the importance of the arthroplasty register in identifying the long-term outcome of hip prostheses.
In 2012 we reviewed a consecutive series of 92
uncemented THRs performed between 1986 and 1991 at our institution
using the CLS Spotorno stem, in order to assess clinical outcome
and radiographic data at a minimum of 21 years. The series comprised
92 patients with a mean age at surgery of 59.6 years (39 to 77)
(M:F 43;49). At the time of this review, seven (7.6%) patients had died and
two (2.2%) were lost to follow-up. The 23-year Kaplan–Meier survival
rates were 91.5% (95% confidence intervals (CI) 85.4% to 97.6%;
55 hips at risk) and 80.3% (95% CI, 71.8% to 88.7%; 48 hips at risk)
respectively, with revision of the femoral stem or of any component
as endpoints. At the time of this review, 76 patients without stem
revision were assessed clinically and radiologically (mean follow-up
24.0 years (21.5 to 26.5)). For the 76 unrevised hips the mean Harris
hip score was 87.1 (65 to 97). Femoral osteolysis was detected in
five hips (6.6%) only in Gruen zone 7. Undersized stems were at
higher risk of revision owing to aseptic loosening (p = 0.0003).
Patients implanted with the stem in a varus position were at higher risk
of femoral cortical hypertrophy and thigh pain (p = 0.0006 and p
= 0.0007, respectively). In our study, survival, clinical outcome and radiographic data
remained excellent in the third decade after implantation. Nonetheless,
undersized stems were at higher risk of revision owing to aseptic
loosening. Cite this article:
We describe the clinical and radiological results of 120 consecutive revision hip replacements in 107 patients, using the JRI Furlong hydroxyapatite-ceramic-coated femoral component. The mean age of the patients at operation was 71 years (36 to 92) and the mean length of follow-up 8.0 years (5.0 to 12.4). We included patients on whom previous revision hip surgery had taken place. The patients were independently reviewed and scored using the Harris hip score, the Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and the Charnley modification of the Merle d’Aubigné and Postel score. Radiographs were assessed by three reviewers for the formation of new bone, osteolysis, osseointegration and radiolucent lines in each Gruen zone. The mean Harris hip score was 85.8 (42 to 100) at the latest post-operative review. The mean WOMAC and Merle d’Aubigné and Postel scores were 34.5 and 14.8, respectively. The mean visual analogue score for pain (possible range 0 to 10) was 1.2 overall, but 0.5 specifically for mid-thigh pain. There were no revisions of the femoral component for aseptic loosening. There were four re-revisions, three for infection and one for recurrent dislocation. Radiological review of all the femoral components, including the four re-revisions showed stable bony ingrowth and no new radiolucent lines in any zone. Using revision or impending revision for aseptic loosening as an end-point, the cumulative survival of the femoral component at ten years was 100% (95% confidence interval 94 to 100). We present excellent medium- to long-term clinical, radiological and survivorship results with the fully hydroxyapatite-ceramic-coated femoral component in revision hip surgery.
In a time of limited resources, the debate continues
over which types of hip prosthesis are clinically superior and more
cost-effective. Orthopaedic surgeons increasingly need robust economic
evidence to understand the full value of the operation, and to aid
decision making on the ‘package’ of procedures that are available
and to justify their practice beyond traditional clinical preference. In this paper we explore the current economic debate about the
merits of cemented and cementless total hip replacement, an issue
that continues to divide the orthopaedic community. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
Impaction bone grafting for the reconstitution
of bone stock in revision hip surgery has been used for nearly 30 years.
Between 1995 and 2001 we used this technique in acetabular reconstruction,
in combination with a cemented component, in 304 hips in 292 patients
revised for aseptic loosening. The only additional supports used
were stainless steel meshes placed against the medial wall or laterally
around the acetabular rim to contain the graft. All Paprosky grades
of defect were included. Clinical and radiographic outcomes were
collected in surviving patients at a minimum of ten years after
the index operation. Mean follow-up was 12.4 years ( Cite this article:
Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
The purpose of this study was to examine the
complications and outcomes of total hip replacement (THR) in super-obese
patients (body mass index (BMI) >
50 kg/m2) compared
with class I obese (BMI 30 to 34.9 kg/m2) and normal-weight
patients (BMI 18.5 to 24.9 kg/m2), as defined by the
World Health Organization. A total of 39 THRs were performed in 30 super-obese patients
with a mean age of 53 years (31 to 72), who were followed for a
mean of 4.2 years (2.0 to 11.7). This group was matched with two
cohorts of normal-weight and class I obese patients, each comprising
39 THRs in 39 patients. Statistical analysis was performed to determine differences
among these groups with respect to complications and satisfaction
based on the Western Ontario and McMaster Universities (WOMAC) osteoarthritis
index, the Harris hip score (HHS) and the Short-Form (SF)-12 questionnaire. Super-obese patients experienced significantly longer hospital
stays and higher rates of major complications and readmissions than
normal-weight and class I obese patients. Although super-obese patients
demonstrated reduced pre-operative and post-operative satisfaction
scores, there was no significant difference in improvement, or change in
the score, with respect to HHS or the WOMAC osteoarthritis index. Super-obese patients obtain similar satisfaction outcomes as
class I obese and normal-weight patients with respect to improvement
in their scores. However, they experience a significant increase
in length of hospital stay and major complication and readmission
rates. Cite this article:
To quantify and compare peri-acetabular bone mineral density
(BMD) between a monoblock acetabular component using a metal-on-metal
(MoM) bearing and a modular titanium shell with a polyethylene (PE)
insert. The secondary outcome was to measure patient-reported clinical
function. A total of 50 patients (25 per group) were randomised to MoM
or metal-on-polyethlene (MoP). There were 27 women (11 MoM) and
23 men (14 MoM) with a mean age of 61.6 years (47.7 to 73.2). Measurements
of peri-prosthetic acetabular and contralateral hip (covariate)
BMD were performed at baseline and at one and two years’ follow-up.
The Western Ontario and McMaster Universities osteoarthritis index
(WOMAC), University of California, Los Angeles (UCLA) activity score,
Harris hip score, and RAND-36 were also completed at these intervals.Objectives
Methods
The technique of femoral cement-in-cement revision
is well established, but there are no previous series reporting its
use on the acetabular side at the time of revision total hip replacement.
We describe the technique and report the outcome of 60 consecutive
acetabular cement-in-cement revisions in 59 patients at a mean follow-up
of 8.5 years (5 to 12). All had a radiologically and clinically
well-fixed acetabular cement mantle at the time of revision. During
the follow-up 29 patients died, but no hips were lost to follow-up.
The two most common indications for acetabular revision were recurrent
dislocation (46, 77%) and to complement femoral revision (12, 20%). Of the 60 hips, there were two cases of aseptic loosening of
the acetabular component (3.3%) requiring re-revision. No other
hip was clinically or radiologically loose (96.7%) at the latest
follow-up. One hip was re-revised for infection, four for recurrent
dislocation and one for disarticulation of a constrained component.
At five years the Kaplan-Meier survival rate was 100% for aseptic
loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause
as the endpoint. These results support the use of cement-in-cement revision on
the acetabular side in appropriate cases. Theoretical advantages
include preservation of bone stock, reduced operating time, reduced
risk of complications and durable fixation.
Revision total hip arthroplasty (THA) is projected
to increase by 137% from the years 2005 to 2030. Reconstruction of
the femur with massive bone loss can be a formidable undertaking.
The goals of revision surgery are to create a stable construct,
preserve bone and soft tissues, augment deficient host bone, improve
function, provide a foundation for future surgery, and create a
biomechanically restored hip. Options for treatment of the compromised femur
include: resection arthroplasty, allograft prosthetic composite
(APC), proximal femoral replacement, cementless fixation with a
modular tapered fluted stem, and impaction grafting. The purpose
of this article is to review the treatment options along with their
associated outcomes in the more severe femoral defects (Paprosky types
IIIb and IV) in revision THA.
Two Durasul highly crosslinked polyethylene liners were exchanged during revision surgery four and five years after implantation, respectively. The retrieved liners were evaluated macroscopically and surface analysis was performed using optical and electron microscopy. A sample of each liner was used to determine the oxidation of the material by Fourier transform infrared spectroscopy. Samples of the capsule were examined histologically. The annual wear rate was found to be 0.010 and 0.015 mm/year, respectively. Surface analysis showed very little loss of material caused by wear. Histological evaluation revealed a continuous neosynovial lining with single multinucleated foreign-body giant cells. Our findings showed no unexpected patterns of wear on the articulating surfaces up to five years after implantation and no obvious failure of material.
Between 1999 and 2001, 90 patients underwent
total hip replacement using the same uncemented acetabular and femoral
components with a 28 mm metallic femoral head but with prospective
randomisation of the acetabular liner to either Durasul highly cross-linked
polyethylene or nitrogen-sterilised Sulene polyethylene. We assessed
83 patients at a minimum follow-up of ten years. Linear penetration
of the femoral head was estimated at six weeks, six and 12 months
and annually thereafter, using the Dorr method, given the non-spherical
shape of the acetabular component. There was no loosening of any component; only one hip in the
Sulene group showed proximal femoral osteolysis. The mean penetration
of the femoral head at six weeks was 0.08 mm (0.02 to 0.15) for
the Durasul group and 0.16 mm (0.05 to 0.28) for the Sulene group
(p = 0.001). The mean yearly linear penetration was 64.8% lower
for the Durasul group at 0.05 mm/year ( Cite this article:
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score. There was a significant improvement ( Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
The best treatment for the active and lucid elderly patient with a displaced intracapsular fracture of the femoral neck is still controversial. Randomised controlled trials have shown that a primary total hip replacement is superior to internal fixation as regards the need for secondary surgery, hip function and health-related quality of life. Despite good results achieved with total hip replacement in this group, most orthopaedic surgeons still advocate hemiarthroplasty for this injury. We studied 120 patients with a mean age of 81 years (70 to 90) with an acute displaced intracapsular fracture of the femoral neck. They were randomly allocated to be treated with either a bipolar hemiarthroplasty or total hip replacement. Outcome measurements included peri-operative data, general and hip-specific complications, hip function and health-related quality of life. The patients were reviewed at four and 12 months. The duration of surgery was longer in the total hip replacement group (102 minutes (70 to 151)) These results indicate that a total hip replacement provides better function than a bipolar hemiarthroplasty as soon as one year post-operatively, without increasing the complication rate. We recommend total hip replacement as the primary treatment for this group of patients.
Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6)
Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications. We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre. At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip. The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
We reviewed the seven- to ten-year results of our previously reported prospective randomised controlled trial comparing total hip replacement and hemiarthroplasty for the treatment of displaced intracapsular fracture of the femoral neck. Of our original study group of 81 patients, 47 were still alive. After a mean follow up of nine years (7 to 10) overall mortality was 32.5% and 51.2% after total hip replacement and hemiarthroplasty, respectively (p = 0.09). At 100 months postoperatively a significantly greater proportion of hemiarthroplasty patients had died (p = 0.026). Three hips dislocated following total hip replacement and none after hemiarthroplasty. In both the total hip replacement and hemiarthroplasty groups a deterioration had occurred in walking distance (p = 0.02 and p <
0.001, respectively). One total hip replacement required revision compared with four hemiarthroplasties which were revised to total hip replacements. All surviving patients with a total hip replacement demonstrated wear of the cemented polyethylene component and all hemiarthroplasties had produced acetabular erosion. There was lower mortality (p = 0.013) and a trend towards superior function in patients with a total hip replacement in the medium term.
Persistent groin pain after seemingly successful
total hip replacement (THR) appears to have become more common.
Recent studies have indicated a high incidence after metal-on-polyethylene
and metal-on-metal conventional THR and it has been documented in
up to 18% of patients after metal-on-metal resurfacing. There are many
causes, including acetabular loosening, stress fracture, and iliopsoas
tendonitis and impingement. The evaluation of this problem requires
a careful history and examination, plain radiographs and an algorithmic approach
to special diagnostic imaging and tests. Non-operative treatment
is not usually successful. Specific operative treatment depending
on the cause of the pain usually involves revision of the acetabular
component, iliopsoas tenotomy or other procedures, and is usually
successful. Here, an appropriate algorithm is described.
Increased femoral head size may reduce dislocation rates following total hip replacement. The National Joint Registry for England and Wales has highlighted a statistically significant increase in the use of femoral heads ≥ 36 mm in diameter from 5% in 2005 to 26% in 2009, together with an increase in the use of the posterior approach. The aim of this study was to determine whether rates of dislocation have fallen over the same period. National data for England for 247 546 procedures were analysed in order to determine trends in the rate of dislocation at three, six, 12 and 18 months after operation during this time. The 18-month revision rates were also examined. Between 2005 and 2009 there were significant decreases in cumulative dislocations at three months (1.12% to 0.86%), six months (1.25% to 0.96%) and 12 months (1.42% to 1.11%) (all p <
0.001), and at 18 months (1.56% to 1.31%) for the period 2005 to 2008 (p <
0.001). The 18-month revision rates did not significantly change during the study period (1.26% to 1.39%, odds ratio 1.10 (95% confidence interval 0.98 to 1.24), p = 0.118). There was no evidence of changes in the coding of dislocations during this time. These data have revealed a significant reduction in dislocations associated with the use of large femoral head sizes, with no change in the 18-month revision rate.
We reviewed the results of 84 total hip replacements performed with a short metaphyseal-fitting anatomical cementless femoral component in 84 unselected consecutive patients with a mean age of 78.9 years (70 to 88). The mean follow-up was 4.6 years (4 to 5). The mean pre-operative Harris hip score was 26 points (0 to 56), which improved to 89 (61 to 100) at the final follow-up. No patient had thigh pain. The mean pre-operative Western Ontario and McMaster Universities osteoarthritis index score was 61 points (48 to 75), which improved to 21 (6 to 46). The mean University of California, Los Angeles activity score was 5.5 points (3 to 7) at the final follow-up. Osseointegration was seen in all femoral and acetabular components. All hips had grade 1 stress shielding of the proximal femur. No acetabular or femoral osteolysis was identified. These results demonstrate that a short metaphyseal-fitting femoral component achieves optimal fixation without diaphyseal anchorage in elderly patients.
Between January 2000 and December 2007, 31 patients 90 years of age or older underwent total hip replacement at our hospital. Their data were collected prospectively. The rate of major medical complications was 9%. The surgical re-operation rate was 3%. The requirement for blood transfusion was 71% which was much higher than for younger patients. The 30-day, one-year and current mortality figures were 6.4% (2 of 31), 9.6% (3 of 31) and 55% (17 of 31), respectively, with a mean follow-up for the 14 surviving patients of six years. Cox’s regression analysis revealed no significant independent predictors of mortality. Only 52% of patients returned immediately to their normal abode, with 45% requiring a prolonged period of rehabilitation. This is the first series to assess survival five years after total hip replacement for patients in their 90th year and beyond. Hip replacement in the extreme elderly should not be discounted on the grounds of age alone, although the complication rate exceeds that for younger patients. It can be anticipated that almost half of the patients will survive five years after surgery.
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS Spotorno replacement with an uncemented, straight, collarless tapered titanium stem between January 1985 and December 1989. The mean follow-up was 17 years (15 to 20). The occurrence of peri-prosthetic femoral fracture during follow-up was noted. Kaplan-Meier survival analysis was used to estimate the cumulative incidence of fracture. At the last follow-up, 86 patients (89 hips) had died and eight patients (eight hips) had been lost to follow-up. A total of 14 fractures in 14 patients had occurred. In ten hips, the femoral component had to be revised and in four the fracture was treated by open reduction and internal fixation. The cumulative incidence of peri-prosthetic femoral fracture was 1.6% (95% confidence interval 0.7 to 3.8) at ten years and 4.5% (95% confidence interval 2.6 to 8.0) at 17 years after the primary THR. There was no association between the occurrence of fracture and gender or age at the time of the primary replacement. Our findings indicate that peri-prosthetic femoral fracture is a significant mode of failure in the long term after the insertion of an uncemented CLS Spotorno stem. Revision rates for this fracture rise in the second decade. Further research is required to investigate the risk factors involved in the occurrence of late peri-prosthetic femoral fracture after the implantation of any uncemented stem, and to assess possible methods of prevention.
We prospectively evaluated the long-term outcome of 158 consecutive patients who underwent revision total hip replacement using uncemented computer-assisted design-computer-assisted manufacture femoral components. There were 97 men and 61 women. Their mean age was 63.1 years (34.6 to 85.9). The mean follow-up was 10.8 years (10 to 12). The mean Oxford, Harris and Western Ontario and McMaster hip scores improved from 41.1, 44.2 and 52.4 pre-operatively to 18.2, 89.3 and 12.3, respectively (p <
0.0001, for each). Six patients required further surgery. The overall survival of the femoral component was 97% (95% confidence interval 94.5 to 99.7). These results are comparable to those of previously published reports for revision total hip replacement using either cemented or uncemented components.
We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully porous-coated threaded hemispheric titanium shell (T-Tap ST) with a calcium ion stearate-free, isostatically compression-moulded polyethylene liner. The titanium femoral component used was a Taperloc with a reduced distal stem. No patient was lost to follow-up. Complete clinical and radiological follow-up was obtained for all 123 hips at a mean of 14 years (12 to 16). One femoral component was revised after a fracture, and three acetabular components for aseptic loosening. No additional femoral or acetabular components were judged loose by radiological criteria. Mild proximal femoral osteolysis was identified in two hips and minor acetabular osteolysis was present in four. The mean rate of penetration of the femoral head was 0.036 mm/year (0.000 to 0.227). These findings suggest that refinements in component design may be associated with excellent long-term fixation in cementless primary total hip replacement.
Dislocation is a common and well-studied complication after total hip replacement. However, subluxation, which we define as a clinically recognised episode of incomplete movement of the femoral head outside the acetabulum with spontaneous reduction, has not been studied previously. Out of a total of 2521 hip replacements performed over 12 years by one surgeon, 30 patients experienced subluxations which occurred in 31 arthroplasties. Data were collected prospectively with a minimum follow-up of two years. Subluxation occurred significantly more frequently after revision than after primary hip replacement, and resolved in 19 of 31 cases (61.3%). In six of the 31 hips (19.4%) the patient subsequently dislocated the affected hip, and in six hips (19.4%) intermittent subluxation continued. Four patients had a revision operation for instability, three for recurrent dislocation and one for recurrent subluxation. Clinical and radiological comparisons with a matched group of stable total hips showed no correlation with demographic or radiological parameters. Patients with subluxing hips reported significantly more concern that their hip would dislocate, more often changed their behaviour to prevent instability and had lower postoperative Harris hip scores than patients with stable replacements.
We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips). Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant. This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging. Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.
The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising. Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft.
We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip Resurfacing arthroplasties in 98 patients. The procedures were performed between October 1999 and June 2002 by one surgeon. All patients were followed up clinically and radiologically. The mean follow-up was 71 months (60 to 93). Revision of either component was defined as failure. The mean Harris Hip score at follow-up was 96.4 (53 to 100). The mean Oxford hip score was 41.9 (16 to 57) pre-operatively and 15.4 (12 to 49) post-operatively (p <
0.001). The mean University of California Los Angeles activity score was 3.91 (1 to 10) pre-operatively and 7.5 (4 to 10) post-operatively (p <
0.001). There were four failures giving a survival at five years of 96.3% (95% confidence interval 92.8 to 99.8). When applying a new method to estimate narrowing of the femoral neck we identified a 10% thinning of the femoral neck in 16 hips (14.5%), but the relevance of this finding to the long-term outcome remains unclear. These good medium-term results from an independent centre confirm the original data from Birmingham.
Between 1986 and 1991 we implanted 331 consecutive Furlong hydroxyapatite-coated femoral components of a total hip replacement in 291 patients. A cemented acetabular prosthesis was used in 217 hips and a hydroxyapatite-coated component in 114. We describe the long-term clinical and radiological survival of the femoral component at a mean follow-up of 17.5 years (15 to 21). Only two patients (0.68%) were lost to follow-up. With revision of the femoral component for any reason as the endpoint, the survival at a mean of 17 years was 97.4% (95% confidence interval 94.1 to 99.5), and with revision for aseptic loosening as the endpoint it was 100%. The survival at a maximum of 21 years with revision of the femoral component for any reason as the endpoint was 97.4% (95% confidence interval 81.0 or 99.5). These results compare favourably with the best long-term results of cemented or uncemented femoral components used in total hip replacement.
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p <
0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).
Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of >
10 mm in five (7.2%). The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.
Between April 1999 and April 2004, 3497 Birmingham hips were inserted by 89 surgeons. Fracture of the neck of the femur occurred in 50 patients, an incidence of 1.46%. Complete data were available for 45 patients in whom the incidence was 1.91% for women and 0.98% for men. The relative risk of fracture for women The mean time to fracture was 15.4 weeks. In women the fractures occurred at a mean of 18.5 weeks and were more likely to have been preceded by a pro-dromal phase of pain and limping. In men the mean time to fracture was 13.5 weeks. Significant varus placement of the femoral component, intra-operative notching of the femoral neck and technical problems were common factors in 85% of cases.