Advertisement for orthosearch.org.uk
Results 1 - 11 of 11
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 4 | Pages 546 - 549
1 Jul 1996
Haddad FS Cobb AG Bentley G Levell NJ Dowd PM

Aseptic loosening is seen in a significant proportion of cemented total hip replacements (THR). In a small subgroup of patients who suffer early loosening polyethylene debris is unlikely to be responsible. We recently reported one case of allergic contact dermatitis to N,N-dimethylparatoluidine (DMT), an accelerator used in bone cement. We have therefore investigated this using skin-patch tests to a variety of substances including metals, polyethylene and the separated individual components of Simplex cement.

We studied 70 patients, 15 with aseptic loosening less than two years after THR, 25 with satisfactory long-term cemented fixation, five with infected loosening of cemented THRs and 25 awaiting hip arthroplasty. We found seven positive reactions to DMT, all of them in patients with the rapid onset of aseptic loosening.

Allergy to DMT is recognised in the dental profession in respect of the ‘denture sore mouth’ syndrome, and could also be an occupational hazard since some industrial glues contain DMT. Our results suggest the need for awareness of possible previous dental or occupational exposure to the constituents of bone cement. We recommend the use of skin-patch testing in high-risk cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 66 - 73
1 Jan 1996
Hilding MB Lanshammar H Ryd L

We report a prospective study of gait and tibial component migration in 45 patients with osteoarthritis treated by total knee arthroplasty (TKA). Migration was measured over two years using roentgen stereophotogrammetry. We used the previously established threshold of 200 μm migration in the second postoperative year to distinguish two groups: a risk group of 15 patients and a stable group of 28 patients.

We performed gait analysis before operation and at six months and at two years after TKA. On all three occasions we found significant differences between the two groups in the mean sagittal plane moments of the knee joint. The risk group walked with higher peak flexion moments than the stable group. The two groups were not discriminated by any clinical or radiological criteria or other gait characteristics.

The relationship which we have found between gait with increased flexion moments and risk of tibial component loosening warrants further study as regards the aetiology of prosthetic loosening and possible methods of influencing its incidence.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 1 | Pages 108 - 115
1 Jan 2000
Garcia-Cimbrelo E Diaz-Martin A Madero R Munuera L

Between 1972 and 1990, we performed 168 primary low-friction arthroplasties in 125 patients with acetabular protrusion. Twelve hips were lost to follow-up within eight years and eight which became infected were excluded from the final study. Of the 148 hips remaining, 62 with a mild protrusion were classified as group 1, 54 with moderate or severe protrusion as group 2 and, after 1985, 32 with moderate and severe protrusion which required bone grafts as group 3. The mean follow-up was 18.3 years (3 to 24) for group 1, 17.4 years (8 to 22) for group 2 and ten years (8 to 13) for group 3.

There were 31 revisions of the cup, 12 in group 1 and 19 in group 2. According to the Kaplan-Meier analysis the overall rates at 20 years were 21 ± 10.79% in group 1 and 37 ± 11.90% in group 2. There have been 43 radiological loosenings: 22 in group 1, 21 in group 2 and none so far in group 3, at ten years. The overall loosening rates at 20 years were 42 ± 14.76% in group 1 and 49 ± 19.50% in group 2. The grafts were well incorporated in all group-3 hips, and the bone structure appeared normal after one year.

The distance between the centre of the head of the femoral prosthesis and the approximate true centre of the femoral head was less in group 3 than in groups 1 and 2 (p < 0.01). According to the Cox proportional-hazards regression this was the single most important factor in loosening of the cup (odds ratio 1.11; 95% CI 1.05 to 1.18/mm). Better results were obtained in moderate and severe protrusions reconstructed with bone grafting than in hips with mild protrusion which were not grafted.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 367 - 374
5 May 2022
Sinagra ZP Davis JS Lorimer M de Steiger RN Graves SE Yates P Manning L

Aims

National joint registries under-report revisions for periprosthetic joint infection (PJI). We aimed to validate PJI reporting to the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) and the factors associated with its accuracy. We then applied these data to refine estimates of the total national burden of PJI.

Methods

A total of 561 Australian cases of confirmed PJI were captured by a large, prospective observational study, and matched to data available for the same patients through the AOANJRR.


The Bone & Joint Journal
Vol. 105-B, Issue 4 | Pages 373 - 381
15 Mar 2023
Jandl NM Kleiss S Mussawy H Beil FT Hubert J Rolvien T

Aims

The aim of this study was to evaluate the diagnostic accuracy of the absolute synovial polymorphonuclear neutrophil cell (PMN) count for the diagnosis or exclusion of periprosthetic joint infection (PJI) after total hip (THA) or knee arthroplasty (TKA).

Methods

In this retrospective cohort study, 147 consecutive patients with acute or chronic complaints following THA and TKA were included. Diagnosis of PJI was established based on the 2018 International Consensus Meeting criteria. A total of 39 patients diagnosed with PJI (32 chronic and seven acute) and 108 patients with aseptic complications were surgically revised.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years. The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial. Cite this article: Bone Joint J 2014;96-B:263–9


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 850 - 858
1 Sep 1998
Schai PA Thornhill TS Scott RD

A consecutive series of 235 total knee arthroplasties using the PFC system was followed prospectively for at least ten years in 186 patients. The operation was for osteoarthritis in 150 knees, for rheumatoid arthritis in 83, and for Paget’s disease and femoral osteonecrosis in one knee each. At the latest review 56 patients had died, fi ve were too ill to assess and three could not be traced. The PFC knee replacement utilised was a nonconforming posterior-cruciate-retaining prosthesis with a polyethylene insert which is fl at in the sagittal plane. The patella was resurfaced using a metal-backed component in 170 cases, but later in the series we used an all-polyethylene component in 22 knees; 43 patellae were not resurfaced. The survival without need for reoperation for any reason was 90% at ten years. Nineteen revisions were component-related due to failure of nine metal-backed patellae, nine polyethylene inserts, and one unresurfaced patella; two reoperations were for synovectomy (one for recurrent haemarthrosis and one for recurrent rheumatoid synovitis) and three were for metastatic joint infection. There were no revisions for aseptic loosening of femoral or tibial components, or the all-polyethylene patellar replacement. The PFC system provides good and predictable results in tricompartmental arthritis of the knee. Loosening appeared to be negligible, but there were wear-related problems in 8%. The change from a metal-backed patella and an increase in the contact area of the tibial insert should provide further improvement by minimising wear


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims

The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer.

Materials and Methods

Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 413 - 418
1 Mar 2010
Rothwell AG Hooper GJ Hobbs A Frampton CM

We analysed data from the Oxford hip and knee questionnaires collected by the New Zealand Joint Registry at six months and five years after joint replacement, to determine if there was any relationship between the scores and the risk of early revision. Logistic regression of the six-month scores indicated that for every one-unit decrease in the Oxford score, the risk of revision within two years increased by 9.7% for total hip replacement (THR), 9.9% for total knee replacement (TKR) and 12.0% for unicompartmental knee replacement (UKR). Our findings showed that 70% of the revisions within two years for TKR and 67% for THR and UKR would have been captured by monitoring the lowest 22%, 28% and 28%, respectively, of the Oxford scores. When analysed using the Kalairajah classification a score of < 27 (poor) was associated with a risk of revision within two years of 7.6% for THR, 7.0% for TKR and 24.3% for UKR, compared with risks of 0.7%, 0.7% and 1.8%, respectively, for scores > 34 (good or excellent).

Our study confirms that the Oxford hip and knee scores at six months are useful predictors of early revision after THR and TKR and we recommend their use for the monitoring of the outcome and potential failure in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 229 - 233
1 Feb 2009
Kempshall PJ Metcalfe A Forster MC

As part of the government’s initiative to reduce waiting times for major joint surgery in Wales, the Cardiff and Vale NHS Trust sent 224 patients (258 knees) to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement. The Kinemax total knee replacement system was used in all cases. The cumulative survival rate at three years was 79.2% (95% confidence interval (CI) 69.2 to 86.8) using re-operation for any cause as an endpoint and 85.3% (95% CI 75.9 to 91.8) using aseptic revision as an endpoint. This is significantly worse than that recorded in the published literature. These poor results have resulted in a significant impact on our service.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 856 - 861
1 Jun 2010
Emms NW Stockley I Hamer AJ Wilkinson JM

Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis.

At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p < 0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate.

There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.