Hip fractures are some of the most common fractures encountered in orthopaedic practice. We aimed to identify whether perioperative hypotension is a predictor of 30-day mortality, and to stratify patient groups that would benefit from closer monitoring and early intervention. While there is literature on intraoperative blood pressure, there are limited studies examining pre- and postoperative blood pressure. We conducted a prospective observational cohort study over a one-year period from December 2021 to December 2022. Patient demographic details, biochemical results, and haemodynamic observations were taken from electronic medical records. Statistical analysis was conducted with the Cox proportional hazards model, and the effects of independent variables estimated with the Wald statistic. Kaplan-Meier survival curves were estimated with the log-rank test.Aims
Methods
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
Louis Pasteur once said that: “Fortune favours
the prepared mind.” As one of the great scientists who contributed
to the fight against infection, he emphasised the importance of
being prepared at all times to recognise infection and deal with
it. Despite the many scientific discoveries and technological advances,
such as the advent of antibiotics and the use of sterile techniques,
infection continues to be a problem that haunts orthopaedic surgeons
and inflicts suffering on patients. The medical community has implemented many practices with the
intention of preventing infection and treating it effectively when
it occurs. Although high-level evidence may support some of these
practices, many are based on little to no scientific foundation.
Thus, around the world, there is great variation in practices for
the prevention and management of periprosthetic joint infection. This paper summaries the instigation, conduct and findings of
a recent International Consensus Meeting on Surgical Site and Periprosthetic
Joint Infection. Cite this article:
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods