We sought to establish whether an oxidised zirconium (OxZr) femoral
component causes less loss of polyethylene volume than a cobalt
alloy (CoCr) femoral component in total knee arthroplasty. A total of 20 retrieved tibial inserts that had articulated with
OxZr components were matched with 20 inserts from CoCr articulations
for patient age, body mass index, length of implantation, and revision
diagnosis. Changes in dimensions of the articular surfaces were compared
with those of pristine inserts using laser scanning. The differences
in volume between the retrieved and pristine surfaces of the two
groups were calculated and compared.Aims
Materials and Methods
Aims. The optimal
The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.Aims
Methods
Distal femoral resection in conventional total knee arthroplasty (TKA) utilizes an intramedullary guide to determine coronal alignment, commonly planned for 5° of valgus. However, a standard 5° resection angle may contribute to malalignment in patients with variability in the femoral anatomical and mechanical axis angle. The purpose of the study was to leverage deep learning (DL) to measure the femoral mechanical-anatomical axis angle (FMAA) in a heterogeneous cohort. Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A DL workflow was created to measure the FMAA and validated against human measurements. To reflect potential intramedullary guide placement during manual TKA, two different FMAAs were calculated either using a line approximating the entire diaphyseal shaft, and a line connecting the apex of the femoral intercondylar sulcus to the centre of the diaphysis. The proportion of FMAAs outside a range of 5.0° (SD 2.0°) was calculated for both definitions, and FMAA was compared using univariate analyses across sex, BMI, knee alignment, and femur length.Aims
Methods
Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium. Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia.Aims
Methods
Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model. We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.Aims
Methods
The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.Aims
Methods
Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia. In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee.Aims
Methods
The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime risk of revision may be more meaningful to patients when gauging risks, especially in younger patients. We aimed to assess the lifetime risk of revision for patients in different age categories at the time of undergoing primary TKA. The New Zealand Joint Registry database was used to obtain revision rates, mortality, and the indications for revision for all primary TKAs performed during an 18-year period between January 1999 and December 2016. Patients were stratified into age groups at the time of the initial TKA, and the lifetime risk of revision was calculated according to age, sex, and the American Society of Anesthesiologists (ASA) grade. The most common indications for revision were also analyzed for each age group.Aims
Methods
The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower
The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components. Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components.Aims
Methods
We carried out 60 revision procedures for failed porous coated anatomic total knee replacements in 54 patients, which were divided into two groups. The 14 knees in group I had a well-fixed femoral component at surgery which was retained, and in the 46 knees in group II both tibial and femoral components were loose and were revised using a variety of implants. Our review comprised clinical and radiological assessment. A total of 13 knees required a second revision. Six (42%) in group I failed very early (mean 2.1 years) when compared with seven (15%) in group II (mean 6.8 years). Failure was due to wear of the polyethylene insert by the abraded, retained femoral component (crude odds ratio 4.07; 95% CI 1.07 to 15.5). We recommend a complete change of primary
A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship. A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.Aims
Methods
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment. A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.Aims
Methods
The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups. This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.Aims
Methods
Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties. Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosisAims
Methods
Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty with isolated medial or lateral compartment osteoarthritis. However, polyethylene wear can significantly reduce the lifespan of UKA. Different bearing designs and materials for UKA have been developed to change the rate of polyethylene wear. Therefore, the objective of this study is to investigate the effect of insert conformity and material on the predicted wear in mobile-bearing UKA using a previously developed computational wear method. Two different designs were tested with the same femoral component under identical kinematic input: anatomy mimetic design (AMD) and conforming design inserts with different conformity levels. The insert materials were standard or crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). We evaluated the contact pressure, contact area, wear rate, wear depth, and volumetric wear under gait cycle loading conditions.Objectives
Methods
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
The medially spherical GMK Sphere (Medacta International AG, Castel San Pietro, Switzerland) total knee arthroplasty (TKA) was previously shown to accommodate lateral rollback while pivoting around a stable medial compartment, aiming to replicate native knee kinematics in which some coronal laxity, especially laterally, is also present. We assess coronal plane kinematics of the GMK Sphere and explore the occurrence and pattern of articular separation during static and dynamic activities. Using pulsed fluoroscopy and image matching, the coronal kinematics and articular surface separation of 16 well-functioning TKAs were studied during weight-bearing and non-weight-bearing, static, and dynamic activities. The closest distances between the modelled articular surfaces were examined with respect to knee position, and proportions of joint poses exhibiting separation were computed.Objectives
Methods
Oxidised zirconium was introduced as a material for femoral components
in total knee arthroplasty (TKA) as an attempt to reduce polyethylene
wear. However, the long-term survival of this component is not known. We performed a retrospective review of a prospectively collected
database to assess the ten year survival and clinical and radiological
outcomes of an oxidised zirconium total knee arthroplasty with the
Genesis II prosthesis. The Western Ontario and McMaster Universities Osteoarthritis
Index (WOMAC), Knee Injury and Osteoarthritis Outcome Score (KOOS)
and a patient satisfaction scale were used to assess outcome.Aims
Methods
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean migration was 0.75Â mm ( For the MB group, the mean linear wear rate on the under-surface
was 0.026Â mm/year (
The outcome of total knee replacement (TKR) using
components designed to increase the range of flexion is not fully
understood. The short- to mid-term risk of aseptic revision in high
flexion TKR was evaluated. The endpoint of the study was aseptic
revision and the following variables were investigated: implant
design (high flexion In a cohort of 64 000 TKRs, high flexion components were used
in 8035 (12.5%). The high flexion knees with tibial liners of thickness
>
14 mm had a density of revision of 1.45/100 years of observation,
compared with 0.37/100 in non-high flexion TKR with liners ≤ 14
mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer,
Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR
had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence
interval (CI) 1.48 to 3.50)), an effect that was magnified when
a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)). Surgeons should be cautious when choosing high flexion TKRs,
particularly when thicker tibial liners might be required. Cite this article:
Focal femoral inlay resurfacing has been developed
for the treatment of full-thickness chondral defects of the knee. This
technique involves implanting a defect-sized metallic or ceramic
cap that is anchored to the subchondral bone through a screw or
pin. The use of these experimental caps has been advocated in middle-aged
patients who have failed non-operative methods or biological repair
techniques and are deemed unsuitable for conventional arthroplasty
because of their age. This paper outlines the implant design, surgical
technique and biomechanical principles underlying their use. Outcomes
following implantation in both animal and human studies are also reviewed. Cite this article:
The aim of this prospective multicentre study
was to report the patient satisfaction after total knee replacement (TKR),
undertaken with the aid of intra-operative sensors, and to compare
these results with previous studies. A total of 135 patients undergoing
TKR were included in the study. The soft-tissue balance of each
TKR was quantified intra-operatively by the sensor, and 18 (13%)
were found to be unbalanced. A total of 113 patients (96.7%) in
the balanced group and 15 (82.1%) in the unbalanced group were satisfied
or very satisfied one year post-operatively (p = 0.043). A review of the literature identified no previous study with
a mean level of satisfaction that was greater than the reported
level of satisfaction of the balanced TKR group in this study. Ensuring
soft-tissue balance by using intra-operative sensors during TKR
may improve satisfaction. Cite this article:
The aims of this retrospective study were to
compare the mid-term outcomes following revision total knee replacement
(TKR) in 76 patients (81 knees) <
55 years of age with those
of a matched group of primary TKRs based on age, BMI, gender and
comorbid conditions. We report the activity levels, functional scores,
rates of revision and complications. Compared with patients undergoing
primary TKR, those undergoing revision TKR had less improvement
in the mean Knee Society function scores (8.14 (–55 to +60) Young patients undergoing revision TKR should be counselled that
they can expect somewhat less improvement and a higher risk of complications
than occur after primary TKR. Cite this article:
The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years ( These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.
Debate has raged over whether a cruciate retaining
(CR) or a posterior stabilised (PS) total knee replacement (TKR) provides
a better range of movement (ROM) for patients. Various sub-sets
of CR design are frequently lumped together when comparing outcomes.
Additionally, multiple factors have been proven to influence the
rate of manipulation under anaesthetic (MUA) following TKR. The
purpose of this study was to determine whether different CR bearing
insert designs provide better ROM or different MUA rates. All primary
TKRs performed by two surgeons between March 2006 and March 2009
were reviewed and 2449 CR-TKRs were identified. The same CR femoral
component, instrumentation, and tibial base plate were consistently
used. In 1334 TKRs a CR tibial insert with 3° posterior slope and
no posterior lip was used (CR-S). In 803 there was an insert with
no slope and a small posterior lip (CR-L) and in 312 knees the posterior
cruciate ligament (PCL) was either resected or lax and a deep-dish,
anterior stabilised insert was used (CR-AS). More CR-AS inserts
were used in patients with less pre-operative ROM and greater pre-operative
tibiofemoral deformity and flexion contracture (p <
0.05). The
mean improvement in ROM was highest for the CR-AS inserts (5.9°
(-40° to 55°) Cite this article:
Antibiotic impregnated articulating spacers are
used in two-stage revision total knee arthroplasty to deliver local antibiotic
therapy while preserving function. We have observed infection control
in greater than 95% of cases with functional outcomes approaching
those seen in revision for aseptic loosening. Higher failure has
been observed with methicillin resistant organisms.
We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period. We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure.
We compared patient-reported outcomes of the Kinemax fixed- and mobile-bearing total knee replacement in a multi-centre randomised controlled trial. Patients were randomised to the fixed- or the mobile-bearing prosthesis via a sealed envelope method after the bone cuts had been made in the operating theatre. Randomisation was stratified by centre and diagnosis. Patients were assessed pre-operatively and at eight to 12 weeks, one year and two years post-operatively. Validated questionnaires were used which included the Western Ontario MacMasters University, Short-Form 12, Mental Health Index-5, Knee Injury and Osteoarthritis Outcome Score for Knee-Related Quality of Life and Function in Sport and Recreation scales and a validated scale of satisfaction post-operatively. In total, 242 patients (250 knees) with a mean age of 68 years (40 to 80) were recruited from four NHS orthopaedic centres. Of these, 132 patients (54.5%) were women. No statistically significant differences could be identified in any of the patient-reported outcome scores between patients who received the fixed-bearing or the mobile-bearing knee up to two-years post-operatively.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
We describe the survivorship of the Medial Rotation total knee replacement (TKR) at ten years in 228 cemented primary replacements implanted between October 1994 and October 2006, with their clinical and radiological outcome. This implant has a highly congruent medial compartment, with the femoral component represented by a portion of a sphere which articulates with a matched concave surface on the medial side of the tibial insert. There were 78 men (17 bilateral TKRs) and 111 women (22 bilateral TKRs) with a mean age of 67.9 years (28 to 90). All the patients were assessed clinically and radiologically using the American Knee Society scoring systems. The mean follow-up was for six years (1 to 13) with only two patients lost to follow-up and 34 dying during the period of study, one of whom had required revision for infection. There were 11 revisions performed in total, three for aseptic loosening, six for infection, one for a periprosthetic fracture and one for a painful but well-fixed replacement performed at another centre. With revision for any cause as the endpoint, the survival at ten years was 94.5% (95% CI 85.1 to 100), and with aseptic loosening as the endpoint 98.4% (95% CI 93 to 100). The mean American Knee Society score improved from 47.6 (0 to 88) to 72.2 (26 to 100) and for function from 45.1 (0 to 100) to 93.1 (45 to 100). Radiological review failed to detect migration in any of the surviving knees. The clinical and radiological results of the Medial Rotation TKR are satisfactory at ten years. The increased congruence of the medial compartment has not led to an increased rate of loosening and continued use can be supported.
The purpose of this study was to determine objectively the outcome of total knee replacement in patients with ankylosed knees. There were 82 patients (99 knees) with ankylosed knees who underwent total knee replacement with a condylar constrained or a posterior stabilised prosthesis. Their mean age was 41.9 years (23 to 60) and the mean follow-up was for 8.9 years (6.6 to 14). Pre- and post-operative data included the Hospital for Special Surgery (HSS), the Knee Society (KS) and the Western Ontario and McMaster University Osteoarthritis index (WOMAC) scores. The mean HSS, KS and WOMAC scores improved from 60, 53, and 79 pre-operatively to 81, 85, and 37 at follow-up. These improvements were statistically significant (p = 0.018, 0.001 and 0.014 respectively). The mean physical, social and emotional WOMAC scores also improved significantly (p = 0.032, p = 0.023 and p <
0.001 respectively). The mean satisfaction score was 8.5 ( Total knee replacement gives good mid-term results in patients with ankylosed knees.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the In this The results from this