Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries. Results. The distribution of liner geometries were neutral (39.4%; 79,822), 10° (34.5%; 69,894), 15° (21.6%; 43,722), offset reorientating (2.8%; 5705), offset neutral (0.9%; 1,767), and 20° (0.8%; 1,601). There were 690 (0.34%) revisions for instability. Compared to neutral liners, the adjusted subhazard ratios of revision for instability were: 10°, 0.64 (p < 0.001); 15°, 0.48 (p < 0.001); and offset reorientating, 1.6 (p = 0.010). No association was found with other geometries. 10° and 15° liners had a time-dependent lower risk of revision for instability within the first 1.2 years. In posterior approaches, 10° and 15° liners had a lower risk of revision for instability, with no significant difference between them. The protective effect of lipped over neutral liners was not observed in laterally approached
Aims. Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. Methods. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m. 2. (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97). Results. There were nine cemented DM cup revisions: three for periprosthetic joint infection, three for acetabular aseptic loosening from bone, two for dislocation, and one for a broken cup-cage construct. The two-year survivals free of aseptic DM revision and dislocation were both 92%. There were five postoperative dislocations, all in patients with prior dislocation or abductor deficiency. On radiological review, the DM cup remained well-fixed at the cemented interface in all but one case. Conclusion. While dislocation was not eliminated in this series of complex revision
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
Aims. Implant waste during total hip arthroplasty (THA) represents a significant cost to the USA healthcare system. While studies have explored methods to improve THA cost-effectiveness, the literature comparing the proportions of implant waste by intraoperative technology used during THA is limited. The aims of this study were to: 1) examine whether the use of enabling technologies during THA results in a smaller proportion of wasted implants compared to navigation-guided and conventional manual THA; 2) determine the proportion of wasted implants by implant type; and 3) examine the effects of surgeon experience on rates of implant waste by technology used. Methods. We identified 104,420 implants either implanted or wasted during 18,329 primary
Aims. The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary
Aims. The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). Methods. We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression. Results. A total of 213,146
Aims. The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups. Methods. Between 2010 and 2019, 2,075 primary
Aims. Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision
Aims. The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA). Methods. We included all primary 394
Aims. The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. Methods. We reviewed the results of 165
Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing
Aims. Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. Methods. In a high-volume, tertiary referral centre, we identified 10,112
Aims. Periprosthetic femoral fracture (PPF) is a major complication following total hip arthroplasty (THA). Uncemented femoral components are widely preferred in primary THA, but are associated with higher PPF risk than cemented components. Collared components have reduced PPF rates following uncemented primary THA compared to collarless components, while maintaining similar prosthetic designs. The purpose of this study was to analyze PPF rate between collarless and collared component designs in a consecutive cohort of posterior approach
Aims. Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation
Aims. Our main aim was to describe the trend in the comorbidities of patients undergoing elective total hip arthroplasties (THAs) and knee arthroplasties (KAs) between 1 January 2005 and 31 December 2018 in England. Methods. We combined data from the National Joint Registry (NJR) on primary elective hip and knee arthroplasties performed between 2005 and 2018 with pre-existing conditions recorded at the time of their primary operation from Hospital Episodes Statistics. We described the temporal trend in the number of comorbidities identified using the Charlson Comorbidity Index, and how this varied by age, sex, American Society of Anesthesiologists (ASA) grade, index of multiple deprivation, and type of KA. Results. We included 696,504 and 833,745 elective primary
Aims. Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Methods. Using the database of
a single academic centre, 126 revision
Aims. The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. Methods. We performed a retrospective review of a consecutive series of 11,018
Aims. The current study aimed to compare robotic arm-assisted (RA-THA), computer-assisted (CA-THA), and manual (M-THA) total hip arthroplasty regarding in-hospital metrics including length of stay (LOS), discharge disposition, in-hospital complications, and cost of RA-THA versus M-THA and CA-THA versus M-THA, as well as trends in use and uptake over a ten-year period, and future projections of uptake and use of RA-THA and CA-THA. Methods. The National Inpatient Sample was queried for primary
Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of
Aims. A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision