We analysed in-vivo migration and wear over a long period of all-polyethylene acetabular cups which had not been affected by mechanical loosening. The selection criteria of regular radiological follow-up, good clinical outcome (Charnley score of 5 or 6), continued walking without crutches and no radiological signs of loosening of the acetabular cups were fulfilled by 25 Charnley total hip arthroplasties. Mean migration, measured by the Nunn method, was 0.6 mm in the medial and 0.2 mm in the cranial direction. The mean yearly rate of wear was 0.05 mm and 0.04 mm, with six and two cups having no detectable wear, as measured by the Livermore and Charnley-Cupic methods, respectively. The maximal detected wear was 3.7 mm. There were no changes in the rate of wear with time. Computerised We conclude that long-term successful cups do not migrate and have a very low rate of wear which was not affected by ageing of the polyethylene. There was no evidence that polyethylene wear alone caused mechanical loosening of the cup but high rates of wear seem to have an adverse prognostic value in terms of the long-term survival of the prosthesis.
We compared the outcome of total knee arthroplasty in 19 patients who had had previous patellectomy with the results in a matched series of arthroplasties performed on knees in which the patella was intact. The mean follow-up was 63 months (21 to 114). In the study group, the outcome was poor in five patients. There was instability in the coronal plane in three patients and persistent pain in four. Three supracondylar fractures occurred. The overall complication rate was 36%. In the control group, pain was relieved in every case and there were no complications. Total knee arthroplasty has a higher complication rate and inferior results if the knee has undergone prior patellectomy.
We report the long-term outcome of 218 Charnley low-friction arthroplasties in 141 patients who were 40 years old or younger at the time of surgery. The minimum follow-up was ten years with a mean of 16 years. The probability of the femoral component surviving 20 years was 86% and of the acetabular component, 84%. The chance that both components would survive for this period was 75%. The pathological diagnosis significantly influenced implant survival. In rheumatoid patients the probability of both components surviving at 20 years was 96% compared with 51% in patients with osteoarthritis. Clinical assessment of 103 patients (166 hips) in whom the arthroplasty was still functioning showed that 94% of hips had minimal pain or none. We conclude that in young patients cemented total hip replacement is a good procedure for those with rheumatoid arthritis but that the results are much less reliable in those with osteoarthritis.