Advertisement for orthosearch.org.uk
Results 1 - 50 of 597
Results per page:
Bone & Joint Research
Vol. 10, Issue 8 | Pages 467 - 473
2 Aug 2021
Rodríguez-Collell JR Mifsut D Ruiz-Sauri A Rodríguez-Pino L González-Soler EM Valverde-Navarro AA

Aims. The main objective of this study is to analyze the penetration of bone cement in four different full cementation techniques of the tibial tray. Methods. In order to determine the best tibial tray cementation technique, we applied cement to 40 cryopreserved donor tibiae by four different techniques: 1) double-layer cementation of the tibial component and tibial bone with bone restrictor; 2) metallic cementation of the tibial component without bone restrictor; 3) bone cementation of the tibia with bone restrictor; and 4) superficial bone cementation of the tibia and metallic keel cementation of the tibial component without bone restrictor. We performed CT exams of all 40 subjects, and measured cement layer thickness at both levels of the resected surface of the epiphysis and the endomedular metaphyseal level. Results. At the epiphyseal level, Technique 2 gave the greatest depth compared to the other investigated techniques. At the endomedular metaphyseal level, Technique 1 showed greater cement penetration than the other techniques. Conclusion. The best metaphyseal cementation technique of the tibial component is bone cementation with cement restrictor. Additionally, if full tibial component cementation is to be done, the cement volume used should be about 40 g of cement, and not the usual 20 g. Cite this article: Bone Joint Res 2021;10(8):467–473


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 444 - 451
1 Apr 2022
Laende EK Mills Flemming J Astephen Wilson JL Cantoni E Dunbar MJ

Aims. Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA. Methods. Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation. Results. In the overall group, only fixation was significantly associated with migration (p < 0.001). For uncemented tibial components in males, smoking was significantly associated with lower migration (p = 0.030) and BMI approached significance (p = 0.061). For females with uncemented components, smoking (p = 0.081) and age (p = 0.063) approached significance and were both associated with increased migration. The small number of self-reported smokers in this study warrants cautious interpretation and further investigation. For cemented components in females, larger sizes of tibial component were significantly associated with increased migration (p = 0.004). No factors were significant for cemented components in males. Conclusion. The migration of uncemented tibial components was more sensitive to patient factors than cemented implants. These differences were not consistent by sex, suggesting that it may be of value to evaluate female and male patients separately following TKA. Cite this article: Bone Joint J 2022;104-B(4):444–451


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1118 - 1125
4 Oct 2022
Suda Y Hiranaka T Kamenaga T Koide M Fujishiro T Okamoto K Matsumoto T

Aims. A fracture of the medial tibial plateau is a serious complication of Oxford mobile-bearing unicompartmental knee arthroplasty (OUKA). The risk of these fractures is reportedly lower when using components with a longer keel-cortex distance (KCDs). The aim of this study was to examine how slight varus placement of the tibial component might affect the KCDs, and the rate of tibial plateau fracture, in a clinical setting. Methods. This retrospective study included 255 patients who underwent 305 OUKAs with cementless tibial components. There were 52 males and 203 females. Their mean age was 73.1 years (47 to 91), and the mean follow-up was 1.9 years (1.0 to 2.0). In 217 knees in 187 patients in the conventional group, tibial cuts were made orthogonally to the tibial axis. The varus group included 88 knees in 68 patients, and tibial cuts were made slightly varus using a new osteotomy guide. Anterior and posterior KCDs and the origins of fracture lines were assessed using 3D CT scans one week postoperatively. The KCDs and rate of fracture were compared between the two groups. Results. Medial tibial fractures occurred after surgery in 15 patients (15 OUKAs) in the conventional group, but only one patient (one OUKA) had a tibial fracture after surgery in the varus group. This difference was significant (6.9% vs 1.1%; p = 0.029). The mean posterior KCD was significantly shorter in the conventional group (5.0 mm (SD 1.7)) than in the varus group (6.1 mm (SD 2.1); p = 0.002). Conclusion. In OUKA, the distance between the keel and posterior tibial cortex was longer in our patients with slight varus alignment of the tibial component, which seems to decrease the risk of postoperative tibial fracture. Cite this article: Bone Joint J 2022;104-B(10):1118–1125


Bone & Joint Research
Vol. 11, Issue 5 | Pages 252 - 259
1 May 2022
Cho BW Kang K Kwon HM Lee W Yang IH Nam JH Koh Y Park KK

Aims. This study aimed to identify the effect of anatomical tibial component (ATC) design on load distribution in the periprosthetic tibial bone of Koreans using finite element analysis (FEA). Methods. 3D finite element models of 30 tibiae in Korean women were created. A symmetric tibial component (STC, NexGen LPS-Flex) and an ATC (Persona) were used in surgical simulation. We compared the FEA measurements (von Mises stress and principal strains) around the stem tip and in the medial half of the proximal tibial bone, as well as the distance from the distal stem tip to the shortest anteromedial cortical bone. Correlations between this distance and FEA measurements were then analyzed. Results. The distance from the distal stem tip to the shortest cortical bone showed no statistically significant difference between implants. However, the peak von Mises stress around the distal stem tip was higher with STC than with ATC. In the medial half of the proximal tibial bone: 1) the mean von Mises stress, maximum principal strain, and minimum principal strain were higher with ATC; 2) ATC showed a positive correlation between the distance and mean von Mises stress; 3) ATC showed a negative correlation between the distance and mean minimum principal strain; and 4) STC showed no correlation between the distance and mean measurements. Conclusion. Implant design affects the load distribution on the periprosthetic tibial bone, and ATC can be more advantageous in preventing stress-shielding than STC. However, under certain circumstances with short distances, the advantage of ATC may be offset. Cite this article: Bone Joint Res 2022;11(5):252–259


Bone & Joint Open
Vol. 5, Issue 1 | Pages 20 - 27
17 Jan 2024
Turgeon TR Vasarhelyi E Howard J Teeter M Righolt CH Gascoyne T Bohm E

Aims. A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Methods. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits. Results. The patient cohort mean age was 66 years (SD seven years), 59% were female, and the mean BMI was 32 kg/m. 2. (SD 6 kg/m. 2. ). Mean two-year subsidence of the EF-TKA was 0.056 mm (95% confidence interval (CI) 0.025 to 0.086) versus 0.006 mm (95% CI -0.029 to 0.040) for the Std-TKA, and the two-year maximum total point motion (MTPM) was 0.285 mm (95% upper confidence limit (UCL) ≤ 0.363) versus 0.346 mm (95% UCL ≤ 0.432), respectively, for a mean difference of -0.061 mm (95% CI -0.196 to 0.074). Inducible displacement also did not differ between groups. The MTPMs between 12 and 24 months for each group was below the published threshold of 0.2 mm for predicting early aseptic loosening (p < 0.001 and p = 0.001, respectively). Conclusion. Both the enhanced fixation and the standard tibial implant design showed fixation with a predicted low risk of long-term aseptic loosening. Cite this article: Bone Jt Open 2024;5(1):20–27


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 165 - 170
1 Jun 2021
Larson DJ Rosenberg JH Lawlor MA Garvin KL Hartman CW Lyden E Konigsberg BS

Aims. Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components. Methods. This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples t-test. Logistic regression was used to adjust for age, sex, and preoperative bone loss. Results. No statistically significant differences in stem length, operative side, or indications for revision were found between the two cohorts. Tibial pain at the end of the stem was present in 25.3% (16/63) of cemented stems and 25.5% (12/47) of uncemented stems (p = 1.000); 74.6% (47/63) of cemented patients and 78.7% (37/47) of uncemented patients were satisfied following revision TKA (p = 0.657). Conclusion. There were no differences in patient satisfaction, overall pain, and diaphyseal tibial pain in cemented and uncemented stemmed tibial components in revision TKA. Patient factors, rather than implant selection and surgical technique, likely play a large role in the presence of postoperative pain. Stemmed tibial components have been shown to be a possible source of pain in revision TKA. There is no difference in patient satisfaction or postoperative pain with cemented or uncemented stemmed tibial components in revision TKA. Cite this article: Bone Joint J 2021;103-B(6 Supple A):165–170


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 2 | Pages 252 - 258
1 Mar 1990
Albrektsson B Ryd L Carlsson L Freeman M Herberts P Regner L Selvik G

We studied the effect of a metal tray with an intramedullary stem on the micromotion of the tibial component in total knee arthroplasty. Of 32 uncemented Freeman-Samuelson knee arthroplasties performed in London and Gothenburg, nine had a metal backing and stem added to the tibial component. Micromotion of the tibial components, expressed as migration and inducible displacement, was analysed using roentgen stereophotogrammetric analysis up to two years follow-up. The addition of a metal back and a 110 mm stem to the standard polyethylene component significantly reduced both migration over two years and inducible displacement


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 170 - 176
1 Feb 2020
Bernthal NM Burke ZDC Hegde V Upfill-Brown A Chen CJ Hwang R Eckardt JJ

Aims. We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. Methods. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up. Results. A total of 278 patients were identified with 289 all-polyethylene tibial components. Mechanical survival was 98.4%, 91.1%, and 85.2% at five, ten and 15 years, respectively. A total of 15 mechanical failures were identified at the final follow-up. Of the 13 all-polyethylene tibial components used for revision of a previous tibial component, five (38.5%) failed mechanically. Younger patients (< 18 years vs > 18 years; p = 0.005) and those used as revision components (p < 0.001) had significantly increased rates of failure. Multivariate logistic regression modelling showed revision status to be a positive risk factor for failure (odds ratio (OR) 19.498, 95% confidence interval (CI) 4.598 to 82.676) and increasing age was a negative risk factor for failure (OR 0.927, 95% CI 0.872 to 0.987). Age-stratified risk analysis showed that age > 24 years was no longer a statistically significant risk factor for failure. The final mean MSTS score for all patients was 89% (8.5% to 100.0%). Conclusion. The long-term mechanical survivorship of all-polyethylene tibial components when used for tumour endoprostheses was excellent. Tumour surgeons should consider using these components for their durability and the secondary benefits of reduced cost and ease of removal and revision. However, caution should be taken when using all-polyethylene tibial components in the revision setting as a significantly higher rate of mechanical failure was seen in this group of patients. Cite this article: Bone Joint J. 2020;102-B(2):170–176


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 855 - 863
1 May 2021
Koster LA Meinardi JE Kaptein BL Van der Linden - Van der Zwaag E Nelissen RGHH

Aims. The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). Methods. A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively. Results. A linear mixed-effect model using RSA data of 31 asymmetrical and 38 symmetrical TKAs did not show a difference in mean MTPM migration pattern of the tibial or femoral components. Mean tibial component MTPM at two years postoperative of the asymmetrical TKA design was 0.93 mm and 1.00 mm for the symmetrical design. For the femoral component these values were 1.04 mm and 1.14 mm, respectively. No significant differences were observed in other migration parameters or in clinical and PROM measurements. Conclusion. The TKA design with an asymmetrical tibial component has comparable component migration with the proven TKA with a symmetrical tibial component. This suggests the risk of long-term aseptic loosening of the two designs is comparable. Cite this article: Bone Joint J 2021;103-B(5):855–863


Bone & Joint Research
Vol. 8, Issue 3 | Pages 126 - 135
1 Mar 2019
Sekiguchi K Nakamura S Kuriyama S Nishitani K Ito H Tanaka Y Watanabe M Matsuda S

Objectives. Unicompartmental knee arthroplasty (UKA) is one surgical option for treating symptomatic medial osteoarthritis. Clinical studies have shown the functional benefits of UKA; however, the optimal alignment of the tibial component is still debated. The purpose of this study was to evaluate the effects of tibial coronal and sagittal plane alignment in UKA on knee kinematics and cruciate ligament tension, using a musculoskeletal computer simulation. Methods. The tibial component was first aligned perpendicular to the mechanical axis of the tibia, with a 7° posterior slope (basic model). Subsequently, coronal and sagittal plane alignments were changed in a simulation programme. Kinematics and cruciate ligament tensions were simulated during weight-bearing deep knee bend and gait motions. Translation was defined as the distance between the most medial and the most lateral femoral positions throughout the cycle. Results. The femur was positioned more medially relative to the tibia, with increasing varus alignment of the tibial component. Medial/lateral (ML) translation was smallest in the 2° varus model. A greater posterior slope posteriorized the medial condyle and increased anterior cruciate ligament (ACL) tension. ML translation was increased in the > 7° posterior slope model and the 0° model. Conclusion. The current study suggests that the preferred tibial component alignment is between neutral and 2° varus in the coronal plane, and between 3° and 7° posterior slope in the sagittal plane. Varus > 4° or valgus alignment and excessive posterior slope caused excessive ML translation, which could be related to feelings of instability and could potentially have negative effects on clinical outcomes and implant durability. Cite this article: K. Sekiguchi, S. Nakamura, S. Kuriyama, K. Nishitani, H. Ito, Y. Tanaka, M. Watanabe, S. Matsuda. Bone Joint Res 2019;8:126–135. DOI: 10.1302/2046-3758.83.BJR-2018-0208.R2


Bone & Joint Research
Vol. 7, Issue 3 | Pages 226 - 231
1 Mar 2018
Campi S Mellon SJ Ridley D Foulke B Dodd CAF Pandit HG Murray DW

Objectives. The primary stability of the cementless Oxford Unicompartmental Knee Replacement (OUKR) relies on interference fit (or press fit). Insufficient interference may cause implant loosening, whilst excessive interference could cause bone damage and fracture. The aim of this study was to identify the optimal interference fit by measuring the force required to seat the tibial component of the cementless OUKR (push-in force) and the force required to remove the component (pull-out force). Materials and Methods. Six cementless OUKR tibial components were implanted in 12 new slots prepared on blocks of solid polyurethane foam (20 pounds per cubic foot (PCF), Sawbones, Malmo, Sweden) with a range of interference of 0.1 mm to 1.9 mm using a Dartec materials testing machine HC10 (Zwick Ltd, Herefordshire, United Kingdom) . The experiment was repeated with cellular polyurethane foam (15 PCF), which is a more porous analogue for trabecular bone. Results. The push-in force progressively increased with increasing interference. The pull-out force was related in a non-linear fashion to interference, decreasing with higher interference. Compared with the current nominal interference, a lower interference would reduce the push-in forces by up to 45% (p < 0.001 One way ANOVA) ensuring comparable (or improved) pull-out forces (p > 0.05 Bonferroni post hoc test). With the more porous bone analogue, although the forces were lower, the relationship between interference and push-in and pull-out force were similar. Conclusions. This study suggests that decreasing the interference fit of the tibial component of the cementless OUKR reduces the push-in force and can increase the pull-out force. An optimal interference fit may both improve primary fixation and decrease the risk of fracture. Cite this article: S. Campi, S. J. Mellon, D. Ridley, B. Foulke, C. A. F. Dodd, H. G. Pandit, D. W. Murray. Optimal interference of the tibial component of the cementless Oxford Unicompartmental Knee Replacement. Bone Joint Res 2018;7:226–231. DOI: 10.1302/2046-3758.73.BJR-2017-0193.R1


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1579 - 1584
1 Dec 2018
Turgeon TR Gascoyne TC Laende EK Dunbar MJ Bohm ER Richardson CG

Aims. The introduction of a novel design of total knee arthroplasty (TKA) must achieve outcomes at least as good as existing designs. A novel design of TKA with a reducing radius of the femoral component and a modified cam-post articulation has been released and requires assessment of the fixation to bone. Radiostereometric analysis (RSA) of the components within the first two postoperative years has been shown to be predictive of medium- to long-term fixation. The aim of this study was to assess the stability of the tibial component of this system during this period of time using RSA. Patients and Methods. A cohort of 30 patients underwent primary, cemented TKA using the novel posterior stabilized fixed-bearing (ATTUNE) design. There was an even distribution of men and women (15:15). The mean age of the patients was 64 years (sd 8) at the time of surgery; their mean body mass index (BMI) was 35.4 kg/m2 (sd 7.9). RSA was used to assess the stability of the tibial component at 6, 12, and 24 months compared with a six-week baseline examination. Patient-reported outcome measures were also assessed. Results. The mean maximum total point motion (MTPM) of the tibial component between 12 and 24 months postoperatively was 0.08 mm (sd 0.08), which is well below the published threshold of 0.2 mm (p < 0.001). Patient-reported outcome measures consistently improved. Conclusion. The tibial component of this novel design of TKA showed stability between assessment 12 and 24 months postoperatively, suggesting an acceptably low risk of medium- to long-term failure due to aseptic loosening


Bone & Joint Research
Vol. 6, Issue 11 | Pages 623 - 630
1 Nov 2017
Suh D Kang K Son J Kwon O Baek C Koh Y

Objectives. Malalignment of the tibial component could influence the long-term survival of a total knee arthroplasty (TKA). The object of this study was to investigate the biomechanical effect of varus and valgus malalignment on the tibial component under stance-phase gait cycle loading conditions. Methods. Validated finite element models for varus and valgus malalignment by 3° and 5° were developed to evaluate the effect of malalignment on the tibial component in TKA. Maximum contact stress and contact area on a polyethylene insert, maximum contact stress on patellar button and the collateral ligament force were investigated. Results. There was greater total contact stress in the varus alignment than in the valgus, with more marked difference on the medial side. An increase in ligament force was clearly demonstrated, especially in the valgus alignment and force exerted on the medial collateral ligament also increased. Conclusion. These results highlight the importance of accurate surgical reconstruction of the coronal tibial alignment of the knee joint. Varus and valgus alignments will influence wear and ligament stability, respectively in TKA. Cite this article: D-S. Suh, K-T. Kang, J. Son, O-R. Kwon, C. Baek, Y-G. Koh. Computational study on the effect of malalignment of the tibial component on the biomechanics of total knee arthroplasty: A Finite Element Analysis. Bone Joint Res 2017;6:623–630. DOI: 10.1302/2046-3758.611.BJR-2016-0088.R2


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1596 - 1602
1 Dec 2017
Dunbar MJ Laende EK Collopy D Richardson CG

Aims. Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation. Patients and Methods. A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component. Results. While there was significant variation in individual migration patterns, the overall migration of the tibial component in the study group demonstrated a pattern of initial migration followed by stabilisation after one year, with mean maximum total point motion (MTPM) of 0.02 mm (standard deviation (. sd. ) 0.20) between one and two years post-operatively. The direction of greatest motion was subsidence, which stabilised at three months post-operatively (mean translation of 0.21 mm, . sd. 0.40). Conclusion. The tibial component migration pattern of stabilisation in the second post-operative year is indicative of successful long-term fixation for this PA-coated tibial component. Cite this article: Bone Joint J 2017;99-B:1596–1602


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1466 - 1470
1 Oct 2010
Didden K Luyckx T Bellemans J Labey L Innocenti B Vandenneucker H

The biomechanics of the patellofemoral joint can become disturbed during total knee replacement by alterations induced by the position and shape of the different prosthetic components. The role of the patella and femoral trochlea has been well studied. We have examined the effect of anterior or posterior positioning of the tibial component on the mechanisms of patellofemoral contact in total knee replacement. The hypothesis was that placing the tibial component more posteriorly would reduce patellofemoral contact stress while providing a more efficient lever arm during extension of the knee. We studied five different positions of the tibial component using a six degrees of freedom dynamic knee simulator system based on the Oxford rig, while simulating an active knee squat under physiological loading conditions. The patellofemoral contact force decreased at a mean of 2.2% for every millimetre of posterior translation of the tibial component. Anterior positions of the tibial component were associated with elevation of the patellofemoral joint pressure, which was particularly marked in flexion > 90°. From our results we believe that more posterior positioning of the tibial component in total knee replacement would be beneficial to the patellofemoral joint


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 45 - 49
1 Jan 2007
Ikeuchi M Yamanaka N Okanoue Y Ueta E Tani T

We prospectively assessed the benefits of using either a range-of-movement technique or an anatomical landmark method to determine the rotational alignment of the tibial component during total knee replacement. We analysed the cut proximal tibia intraoperatively, determining anteroposterior axes by the range-of-movement technique and comparing them with the anatomical anteroposterior axis. We found that the range-of-movement technique tended to leave the tibial component more internally rotated than when anatomical landmarks were used. In addition, it gave widely variable results (mean 7.5°; 2° to 17°), determined to some extent by which posterior reference point was used. Because of the wide variability and the possibilities for error, we consider that it is inappropriate to use the range-of-movement technique as the sole method of determining alignment of the tibial component during total knee replacement


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 140 - 143
1 Nov 2013
Scott RD

At least four ways have been described to determine femoral component rotation, and three ways to determine tibial component rotation in total knee replacement (TKR). Each method has its advocates and each has an influence on knee kinematics and the ultimate short and long term success of TKR. Of the four femoral component methods, the author prefers rotating the femoral component in flexion to that amount that establishes a stable symmetrical flexion gap. This judgement is made after the soft tissues of the knee have been balanced in extension. Of the three tibial component methods, the author prefers rotating the tibial component into congruency with the established femoral component rotation with the knee is in extension. This yields a rotationally congruent articulation during weight-bearing and should minimise the torsional forces being transferred through a conforming tibial insert, which could lead to wear to the underside of the tibial polyethylene. Rotating platform components will compensate for any mal-rotation, but can still lead to pain if excessive tibial insert rotation causes soft-tissue impingement. Cite this article: Bone Joint J 2013;95-B, Supple A:140–3


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1025 - 1032
1 Aug 2020
Hampton M Mansoor J Getty J Sutton PM

Aims. Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Methods. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m. 2. (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up. Results. At final review all patients were between 11 and 15 years following surgery. In total, 41 of the knees were cemented and 36 uncemented. There were no revisions in the cemented group and one revision in the uncemented group for fracture. The uncemented group reported better outcomes with both statistically and clinically significant (p = 0.001) improvements in knee-specific Oxford and Knee Society scores compared with the cemented group. The global SF-12 scores demonstrated no statistical difference (p = 0.812). Uncemented knees had better radiological analysis compared with the cemented group (p < 0.001). Conclusion. Use of an uncemented trabecular metal tibial implant can afford better long-term clinical outcomes when compared to cemented tibial components of a matched design. However, both have excellent survivorship up to 15 years after implantation. Cite this article: Bone Joint J 2020;102-B(8):1025–1032


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 497 - 503
1 Apr 2012
Bercovy M Beldame J Lefebvre B Duron A

This prospective study compares the outcome of 157 hydroxyapatite (HA)-coated tibial components with 164 cemented components in the ROCC Rotating Platform total knee replacement in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There were two revisions for loosening: one for an HA-coated and one for a cemented tibial component. Radiological evaluation demonstrated no radiolucent lines with the HA-coated femoral components. A total of three HA-coated tibial components exhibited radiolucent lines at three months post-operatively and these disappeared after three further months of protected weight-bearing. With HA-coated components the operating time was shorter (p <  0.006) and the radiological assessment of the tibial interface was more stable (p < 0.01). Using revision for aseptic loosening of the tibial component as the end point, the survival rates at nine years was identical for both groups at 99.1%. Our results suggest that HA-coated components perform at least as well as the same design with cemented components and compare favourably with those of series describing cemented or porous-coated knee replacements, suggesting that fixation of both components with hydroxyapatite is a reliable option in primary total knee replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 222 - 227
1 Mar 2000
Fukuoka S Yoshida K Yamano Y

Accurate quantitative measurements of micromovement immediately after operation would be a reliable indicator of the stability of an individual component. We have therefore developed a system for measuring micromovement of the tibial component using three non-contact displacement transducers attached to the tibial cortex during total knee arthroplasty (TKA). Using this system we measured the initial stability in 31 uncemented TKAs. All the tibial components were fixed by a stem and four screws. The initial stability was defined as the amount of displacement when a load of 20 kg was applied. The mean subsidence was 60.7 μm and the mean lift-off was 103.3 μm. We also studied the migration of the tibial component using roentgen stereophotogrammetric analysis (RSA) for up to two years after operation. Most migration occurred during the first six months, after which all prostheses remained stable. We defined migration as the maximum total point motion (MTPM) at two years after operation. The mean migration was 1.29 mm at two years. Our results show that there was a significant correlation between the initial stability and migration (p < 0.05) and emphasise the importance of the initial stability of the tibial component


Aims. The tibial component of total knee arthroplasty can either be an all-polyethylene (AP) implant or a metal-backed (MB) implant. This study aims to compare the five-year functional outcomes of AP tibial components to MB components in patients aged over 70 years. Secondary aims are to compare quality of life, implant survivorship, and cost-effectiveness. Methods. A group of 130 patients who had received an AP tibial component were matched for demographic factors of age, BMI, American Society of Anesthesiologists (ASA) grade, sex, and preoperative Knee Society Score (KSS) to create a comparison group of 130 patients who received a MB tibial component. Functional outcome was assessed prospectively by KSS, quality of life by 12-Item Short-Form Health Survey questionnaire (SF-12), and range of motion (ROM), and implant survivorships were compared. The SF six-dimension (6D) was used to calculate the incremental cost effectiveness ratio (ICER) for AP compared to MB tibial components using quality-adjusted life year methodology. Results. The AP group had a mean KSS-Knee of 83.4 (standard deviation (SD) 19.2) and the MB group a mean of 84.9 (SD 18.2; p = 0.631), while mean KSS-Function was 75.4 (SD 15.3) and 73.2 (SD 16.2 p = 0.472), respectively. The mental (44.3 vs 45.1; p = 0.464) and physical (44.8 vs 44.9; p = 0.893) dimensions of the SF-12 and ROM (97.9° vs 99.7°; p = 0.444) were not different between the groups. Implant survivorship at five years were 99.2% and 97.7% (p = 0.321). The AP group had a greater SF-6D gain of 0.145 compared to the MB group, with an associated cost saving of £406, which resulted in a negative ICER of -£406/0.145 = -£2,800. Therefore, the AP tibial component was dominant, being a more effective and less expensive intervention. Conclusion. There were no differences in functional outcomes or survivorship at five years between AP and MB tibial components in patients aged 70 years and older, however the AP component was shown to be more cost-effective. In the UK, only 1.4% of all total knee arthroplasties use an AP component; even a modest increase in usage nationally could lead to significant financial savings. Cite this article: Bone Jt Open 2022;3(12):969–976


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 6 | Pages 825 - 831
1 Aug 2001
Adalberth G Nilsson KG Byström S Kolstad K Milbrink J

We studied the quality of fixation of the tibial component using radiostereometric analysis (RSA) in 40 patients who had undergone a cemented Freeman-Samuelson total knee arthroplasty. They were prospectively randomised to either a stemmed metal-backed (MB) or non-stemmed all-polyethylene (AP) tibial component. The articulating geometry of the implants was identical, as was the operative technique and the postoperative regime. The study showed no complications of fixation using AP tibial components, and the migration was the same as that of their metal-backed counterparts. There was no bony collapse or increased subsidence of any part of the tibial component or increased incidence of radiolucent lines in the knees with AP components. Most AP implants were stable between one and two years after surgery, a finding known to be of positive prognostic significance when predicting future aseptic loosening


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1238 - 1244
1 Sep 2010
Nicoll D Rowley DI

This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 609 - 614
1 May 2007
Himanen A Belt EA Lehto MUK Hämäläinen MMJ

We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7). The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory


Bone & Joint Research
Vol. 11, Issue 4 | Pages 229 - 238
11 Apr 2022
Jaeger S Eissler M Schwarze M Schonhoff M Kretzer JP Bitsch RG

Aims. One of the main causes of tibial revision surgery for total knee arthroplasty is aseptic loosening. Therefore, stable fixation between the tibial component and the cement, and between the tibial component and the bone, is essential. A factor that could influence the implant stability is the implant design, with its different variations. In an existing implant system, the tibial component was modified by adding cement pockets. The aim of this experimental in vitro study was to investigate whether additional cement pockets on the underside of the tibial component could improve implant stability. The relative motion between implant and bone, the maximum pull-out force, the tibial cement mantle, and a possible path from the bone marrow to the metal-cement interface were determined. Methods. A tibial component with (group S: Attune S+) and without (group A: Attune) additional cement pockets was implanted in 15 fresh-frozen human leg pairs. The relative motion was determined under dynamic loading (extension-flexion 20° to 50°, load-level 1,200 to 2,100 N) with subsequent determination of the maximum pull-out force. In addition, the cement mantle was analyzed radiologically for possible defects, the tibia base cement adhesion, and preoperative bone mineral density (BMD). Results. The BMD showed no statistically significant difference between both groups. Group A showed for all load levels significantly higher maximum relative motion compared to group S for 20° and 50° flexion. Group S improved the maximum failure load significantly compared to group A without additional cement pockets. Group S showed a significantly increased cement adhesion compared to group A. The cement penetration and cement mantle defect analysis showed no significant differences between both groups. Conclusion. From a biomechanical point of view, the additional cement pockets of the component have improved the fixation performance of the implant. Cite this article: Bone Joint Res 2022;11(4):229–238


Bone & Joint Open
Vol. 3, Issue 6 | Pages 495 - 501
14 Jun 2022
Keohane D Sheridan GA Masterson E

Aims. Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. Methods. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option). Results. Between 2013 and 2016, 352 NexGen TKAs were carried out on 331 patients. A total of 62 TKAs have been revised to date, giving an all-cause revision rate of 17.6% at a minimum of five years. Three of these revisions were due to infection. Overall, 59 of the revisions were performed for aseptic loosening (16.7%) of the tibial component. The tibial component was removed intraoperatively without instrumentation due to significant tibial debonding between the implant-cement interface. Conclusion. While overall, we believe that early aseptic loosening is multi-factorial in nature, the significantly high aseptic revision rate, as seen by an experienced fellowship-trained arthroplasty surgeon, has led us to believe that there is a fundamental issue with this NexGen implant design. Continued implant surveillance and rigorous review across all regions using this particular implant is warranted based on the concerning findings described here. Cite this article: Bone Jt Open 2022;3(6):495–501


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 417 - 425
1 May 1998
Önsten I Nordqvist A Carlsson ÅS Besjakov J Shott S

In a single-blind, randomised series of knee replacements in 116 patients, we used radiostereometric analysis (RSA) to measure micromotion in three types of tibial implant fixation for two years after knee replacement. We compared hydroxyapatite-augmented porous coating, porous coating, and cemented fixation of the same design of tibial component. At one to two years, porous-coated implants migrated at a statistically significantly higher rate than hydroxyapatite-augmented or cemented implants. There was no significant difference between hydroxyapatite-coated and cemented implants. We conclude that hydroxyapatite augmentation may offer a clinically relevant advantage over a simple porous coating for tibial component fixation, but is no better than cemented fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 66-B, Issue 4 | Pages 523 - 528
1 Aug 1984
Tibrewal S Grant K Goodfellow J

Radiolucent lines at the bone-cement interface beneath the tibial components were assessed in 91 consecutive Oxford meniscal knee replacements in 78 patients. Of 80 knees in which radio-opaque cement was used, a radiolucent line was observed in 77, with a radiodense line in the bone immediately adjoining. Radiolucent lines developed in the majority of patients within one year after operation. In 11 knees fixed with radiolucent cement (which precluded assessment of the radiolucent line) a radiodense line was observed beneath the lucent cement in all cases. Histological examination of the interface obtained from secure tibial components showed the lucent zone to be composed of fibrocartilaginous connective tissue and the radiodense line to be a thick lamella of bone. It is suggested that the living bone under a rigid prosthesis requires a layer of relatively compliant fibrocartilaginous material at its interface to accommodate load-bearing. Attention is drawn to the importance of the radiodense line: its presence may constitute positive evidence that healing at the level of bone section is complete and that equilibrium is established; its absence at a mature interface may indicate disequilibrium and impending failure


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 875 - 883
1 Jul 2022
Mills K Wymenga AB van Hellemondt GG Heesterbeek PJC

Aims. Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). Methods. This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction). Results. The median total femoral translation and rotation at ten years were 0.39 mm (interquartile range (IQR) 0.20 to 0.54) and 0.59° (IQR 0.46° to 0.73°) for the cemented group and 0.70 mm (IQR 0.15 to 0.77) and 0.78° (IQR 0.47° to 1.43°) for the hybrid group. For the tibial components this was 0.38 mm (IQR 0.33 to 0.85) and 0.98° (IQR 0.38° to 1.34°) for the cemented group and 0.42 mm (IQR 0.30 to 0.52) and 0.72° (IQR 0.62° to 0.82°) for the hybrid group. None of these values were significantly different between the two groups and there were no significant differences between the clinical scores in the two groups at this time. There was only one re-revision, in the hybrid group, for infection and not for aseptic loosening. Conclusion. These results show good long-term fixation with no difference in micromotion and clinical outcome between fully cemented and hybrid fixation in rTKA, which builds on earlier short- to mid-term results. The patients all had type I or II osseous defects, which may in part explain the good results. Cite this article: Bone Joint J 2022;104-B(7):875–883


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1168 - 1176
1 Nov 2023
Yüksel Y Koster LA Kaptein BL Nelissen RGHH den Hollander P

Aims. Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. Methods. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively. Results. At five-year follow-up, RSA data from 61 patients were available and the mean maximum total point motion (MTPM) of the femoral components were: ATTUNE: 0.96 mm (95% confidence interval (CI) 0.79 to 1.14) and PFC-Sigma 1.37 mm (95% CI 1.18 to 1.59) (p < 0.001). The PFC-Sigma femoral component migrated more in the first postoperative year, but stabilized thereafter. MPTM of the tibial components were comparable at five-year follow-up: ATTUNE 1.12 mm (95% CI 0.95 to 1.31) and PFC-Sigma 1.25 mm (95% CI 1.07 to 1.44) (p = 0.438). RLL at the medial tibial implant-cement interface remained more prevalent for the ATTUNE at five-year follow-up compared to the PFC-Sigma (20% vs 3%). RLL did not progress over time, and varied between patients at different timepoints for both TKA systems. Clinical outcomes and PROMs improved compared with preoperative scores, and were not different between groups. Conclusion. MTPM migration at five-year follow-up of the femoral and tibial component of the ATTUNE were similar and as low as that of the PFC-Sigma. MTPM migration of both knee implants did not significantly change from one year post-surgery, indicating stable fixation. Long-term ATTUNE performance may be expected to be comparable to the clinically well-performing PFC-Sigma. We have not found evidence of increased tibial component migration as measured by RSA to support concerns about cement debonding and a higher risk of aseptic loosening with the ATTUNE TKA. Cite this article: Bone Joint J 2023;105-B(11):1168–1176


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims. The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs. Methods. A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses. Results. The mean MTPM two years postoperatively (95% confidence interval (CI)) of the Attune femoral component (0.92 mm (0.75 to 1.11)) differed significantly from that of the LCS TKA (1.72 mm (1.47 to 2.00), p < 0.001). The Attune femoral component subsided, tilted (anteroposteriorly), and rotated (internal-external) significantly less. The mean tibial MTPM two years postoperatively did not differ significantly, being 1.11 mm (0.94 to 1.30) and 1.17 mm (0.99 to 1.36, p = 0.447) for the Attune and LCS components, respectively. The rate of migration in the second postoperative year was negligible for the femoral and tibial components of both designs. The mean pain-at-rest (numerical rating scale (NRS)-rest) in the Attune group was significantly less compared with that in the LCS group during the entire follow-up period. At three months postoperatively, the Knee injury and Osteoarthritis Outcome Physical Function Shortform score, the Oxford Knee Score, and the NRS-activity scores were significantly better in the Attune group. Conclusion. The mean MTPM of the femoral components of the cementless rotating platform Attune was significantly less compared with that of the LCS design. This was reflected mainly in significantly less subsidence, posterior tilting, and internal rotation. The mean tibial MTPMs were not significantly different. During the second postoperative year, the components of both designs stabilized and low risks for the development of aseptic loosening are expected. Cite this article: Bone Joint J 2023;105-B(2):148–157


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 66 - 73
1 Jan 1996
Hilding MB Lanshammar H Ryd L

We report a prospective study of gait and tibial component migration in 45 patients with osteoarthritis treated by total knee arthroplasty (TKA). Migration was measured over two years using roentgen stereophotogrammetry. We used the previously established threshold of 200 μm migration in the second postoperative year to distinguish two groups: a risk group of 15 patients and a stable group of 28 patients. We performed gait analysis before operation and at six months and at two years after TKA. On all three occasions we found significant differences between the two groups in the mean sagittal plane moments of the knee joint. The risk group walked with higher peak flexion moments than the stable group. The two groups were not discriminated by any clinical or radiological criteria or other gait characteristics. The relationship which we have found between gait with increased flexion moments and risk of tibial component loosening warrants further study as regards the aetiology of prosthetic loosening and possible methods of influencing its incidence


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 31 - 36
1 Jan 2008
Yoo JH Kang YG Chang CB Seong SC Kim TK

We examined the placement of the stem in relation to the medial tibial cortex when using total knee replacements (TKRs) with medially-offset tibial stems in Korean patients. Measurements were performed on the pre- and post-operative radiographs of 246 osteoarthritic knees replaced between January 2005 and May 2006 using the Genesis II or E-motion TKR with a medially-offset stem. Pre-operatively, we measured the distance between the mechanical axis and that of the tibial shaft and post-operatively, that between the midline of the tibial stem and the axis of the shaft. Knees were identified in which there was radiological contact between the tip of the stem and the medial tibial cortex. The mechanical axis was located medial to the axis of the shaft in 203 knees (82.5%). Post-operatively, the midline of the tibial stem was located medial to the tibial shaft axis in 196 knees (79.7%). In 16 knees (6.5%) there was radiological contact between the tibial stem or cement mantle and the medial tibial cortex. Our study has shown that the medially-offset stem in the tibial component may not be a good option for knees undergoing replacement for advanced osteoarthritis in some Korean patients


The Journal of Bone & Joint Surgery British Volume
Vol. 60-B, Issue 2 | Pages 256 - 261
1 May 1978
Bargren J Day W Freeman M Swanson S

Cadaveric knees replaced with the Geomedic, ICLH, Marmor and Total Condylar prostheses were tested in axial compression, in rotation and in hyperextension in order to observe the strength of fixation of the tibial components. In axial compression the strengths at failure varied widely, both with any one prosthesis and between prostheses. This is attributed largely to the strength of the cancellous bone of the tibia, which was measured in each case and also varied widely. Three natural knees failed at loads of 7300, 7600 and 8300 newtons respectively, whereas the strengths of replaced knees ranged from 3000 to 15750 newtons. At least one example of each design failed at less than 7300 newtons, suggesting little or no reserve of strength. The strength of fixation was greater when the tibial prosthesis was large enough to rest on the whole cross-section of the tibia. In rotation the three prostheses embodying rollers in troughs were stiffer than the Marmor which had a nearly flat tibial-bearing surface. The presence or absence of the cruciate ligaments had a negligible effect on torsional stiffness. In hyperextension, knees replaced with the ICLH, Marmor and Total Condylar prostheses failed by rupture of the posterior capsule at moments of about 60 newton-metres, compared with about 100 for natural knees. With the Marmor prosthesis the anterior cruciate ligament was avulsed at about 20 newton-metres compared with about 75 in natural knees, suggesting that in this respect the retention of the cruciate ligaments contributes little. None of the four knees tested after inserting a Geomedic prosthesis showed strengths as high as those replaced with the other three designs


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 295 - 297
1 Mar 1998
Toksvig-Larsen S Magyar G Önsten I Ryd L Lindstrand A

To assess migration of the tibial component we used roentgen stereophotogrammetric analysis in 40 patients who had had a total knee arthroplasty after failure of a closing wedge osteotomy and compared them with 40 matched patients after primary total knee arthroplasty. We found no difference in migration over time or in the tendency for continuous migration between the two groups. There were no differences in alignment or position of the knee prosthesis or in the clinical outcome. Our findings show that revision of a failed high tibial osteotomy to a total knee arthroplasty is effective


Aims. The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients. Methods. A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up. Results. There was no clinically relevant difference in terms of migration of the femoral component or PROMs between the HXLPE and PE groups. The mean difference in migration (maximum total point motion), five years postopeatively, was 0.04 mm (95% CI -0.06 to 0.16) in favour of the PE group. Conclusion. There was no clinically relevant difference in migration of the femoral component, for up to five years between the two groups. These findings will help to establish a benchmark for future studies on migration of femoral components in TKA. Cite this article: Bone Joint J 2024;106-B(8):826–833


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 221 - 226
1 Mar 2023
Wilton T Skinner JA Haddad FS

Recent publications have drawn attention to the fact that some brands of joint replacement may contain variants which perform significantly worse (or better) than their ‘siblings’. As a result, the National Joint Registry has performed much more detailed analysis on the larger families of knee arthroplasties in order to identify exactly where these differences may be present and may hitherto have remained hidden. The analysis of the Nexgen knee arthroplasty brand identified that some posterior-stabilized combinations have particularly high revision rates for aseptic loosening of the tibia, and consequently a medical device recall has been issued for the Nexgen ‘option’ tibial component which was implicated. More elaborate signal detection is required in order to identify such variation in results in a routine fashion if patients are to be protected from such variation in outcomes between closely related implant types. Cite this article: Bone Joint J 2023;105-B(3):221–226


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 2 | Pages 239 - 242
1 Mar 1992
Grewal R Rimmer M Freeman M

We determined the cumulative survival rates, in arthroplasty of the knee, of three designs of tibial component, using a change of position on standard radiographs or revision for aseptic loosening as criteria of failure. The average migration of each of the three designs in the first postoperative year is known from roentgen stereophotogrammetric analysis reported by other authors. The ranking order of the components as judged by cumulative survival is the same as that determined by early migration. This finding supports the view that the measurement of early migration can predict late aseptic loosening and therefore that such measurements are clinically of value


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1231 - 1239
1 Nov 2024
Tzanetis P Fluit R de Souza K Robertson S Koopman B Verdonschot N

Aims. The surgical target for optimal implant positioning in robotic-assisted total knee arthroplasty remains the subject of ongoing discussion. One of the proposed targets is to recreate the knee’s functional behaviour as per its pre-diseased state. The aim of this study was to optimize implant positioning, starting from mechanical alignment (MA), toward restoring the pre-diseased status, including ligament strain and kinematic patterns, in a patient population. Methods. We used an active appearance model-based approach to segment the preoperative CT of 21 osteoarthritic patients, which identified the osteophyte-free surfaces and estimated cartilage from the segmented bones; these geometries were used to construct patient-specific musculoskeletal models of the pre-diseased knee. Subsequently, implantations were simulated using the MA method, and a previously developed optimization technique was employed to find the optimal implant position that minimized the root mean square deviation between pre-diseased and postoperative ligament strains and kinematics. Results. There were evident biomechanical differences between the simulated patient models, but also trends that appeared reproducible at the population level. Optimizing the implant position significantly reduced the maximum observed strain root mean square deviations within the cohort from 36.5% to below 5.3% for all but the anterolateral ligament; and concomitantly reduced the kinematic deviations from 3.8 mm (SD 1.7) and 4.7° (SD 1.9°) with MA to 2.7 mm (SD 1.4) and 3.7° (SD 1.9°) relative to the pre-diseased state. To achieve this, the femoral component consistently required translational adjustments in the anterior, lateral, and proximal directions, while the tibial component required a more posterior slope and varus rotation in most cases. Conclusion. These findings confirm that MA-induced biomechanical alterations relative to the pre-diseased state can be reduced by optimizing the implant position, and may have implications to further advance pre-planning in robotic-assisted surgery in order to restore pre-diseased knee function. Cite this article: Bone Joint J 2024;106-B(11):1231–1239


Aims

The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs).

Methods

A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 277 - 285
8 Apr 2024
Khetan V Baxter I Hampton M Spencer A Anderson A

Aims. The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. Methods. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series. Results. No patients in our series had evidence of radiolucent lines on their latest radiological assessment. Only eight patients out of 454 required revision arthroplasty, and none of these revisions were indicated for aseptic loosening of the tibial baseplate. When compared to data from the NJR annual report, Kaplan-Meier estimates from our series (2.94 (95% confidence interval (CI) 1.24 to 5.87)) show a significant reduction in cumulative estimates of revision compared to all cemented (4.82 (95% CI 4.69 to 4.96)) or cementless TKA (5.65 (95% CI 5.23 to 6.10)). Our data (2.94 (95% CI 1.24 to 5.87)) also show lower cumulative revision rates compared to the most popular implant (PFC Sigma Cemented Knee implant fixation, 4.03 (95% CI 3.75 to 4.33)). The prosthesis time revision rate (PTIR) estimates for our series (2.07 (95% CI 0.95 to 3.83)) were lower than those of cemented cases (4.53 (95% CI 4.49 to 4.57)) from NJR. Conclusion. The NexGen trabecular (tantalum) cementless implant has lower revision rates in our series compared to all cemented implants and other types of cementless implants, and its use in younger patients should be encouraged. Cite this article: Bone Jt Open 2024;5(4):277–285


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1009 - 1015
1 Aug 2012
Scott CEH Biant LC

Stems improve the mechanical stability of tibial components in total knee replacement (TKR), but come at a cost of stress shielding along their length. Their advantages include resistance to shear, reduced tibial lift-off and increased stability by reducing micromotion. Longer stems may have disadvantages including stress shielding along the length of the stem with associated reduction in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic fracture and end-of-stem pain. These features make long stems unattractive in the primary TKR setting, but often desirable in revision surgery with bone loss and instability. In the revision scenario, stems are beneficial in order to convey structural stability to the construct and protect the reconstruction of bony defects. Cemented and uncemented long stemmed implants have different roles depending on the nature of the bone loss involved.

This review discusses the biomechanics of the design of tibial components and stems to inform the selection of the component and the technique of implantation.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 786 - 792
1 Jun 2015
Hutt JRB Farhadnia P Massé V Lavigne M Vendittoli P

This randomised trial evaluated the outcome of a single design of unicompartmental arthroplasty of the knee (UKA) with either a cemented all-polyethylene or a metal-backed modular tibial component. A total of 63 knees in 45 patients (17 male, 28 female) were included, 27 in the all-polyethylene group and 36 in the metal-backed group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and two metal-backed components were revised (at one and five years). One revision in both groups was for unexplained pain, one in the metal-backed group was for progression of osteoarthritis. The others in the all-polyethylene group were for aseptic loosening. The survivorship at seven years calculated by the Kaplan–Meier method for the all-polyethylene group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at risk 16) This difference was statistically significant (p <  0.001). At the most recent follow-up, significantly better mean Western Ontario and McMaster Universities Arthritis Index Scores were found in the all-polyethylene group (13.4 vs 23.0, p = 0.03) but there was no difference in the mean Knee injury and Osteoarthritis Outcome scores (68.8; 41.4 to 99.0 vs 62.6; 24.0 to 100.0), p = 0.36). There were no significant differences for range of movement (p = 0.36) or satisfaction (p = 0.23).

This randomised study demonstrates that all-polyethylene components in this design of fixed bearing UKA had unsatisfactory results with significantly higher rates of failure before ten years compared with the metal-back components.

Cite this article: Bone Joint J 2015;97-B:786–92.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1356 - 1361
1 Oct 2012
Streit MR Walker T Bruckner T Merle C Kretzer JP Clarius M Aldinger PR Gotterbarm T

The Oxford mobile-bearing unicompartmental knee replacement (UKR) is an effective and safe treatment for osteoarthritis of the medial compartment. The results in the lateral compartment have been disappointing due to a high early rate of dislocation of the bearing. A series using a newly designed domed tibial component is reported.

The first 50 consecutive domed lateral Oxford UKRs in 50 patients with a mean follow-up of three years (2.0 to 4.3) were included. Clinical scores were obtained prospectively and Kaplan-Meier survival analysis was performed for different endpoints. Radiological variables related to the position and alignment of the components were measured.

One patient died and none was lost to follow-up. The cumulative incidence of dislocation was 6.2% (95% confidence interval (CI) 2.0 to 17.9) at three years. Survival using revision for any reason and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85 to 99) at three years, respectively. Outcome scores, visual analogue scale for pain and maximum knee flexion showed a significant improvement (p < 0.001). The mean Oxford knee score was 43 (sd 5.3), the mean Objective American Knee Society score was 91 (sd 13.9) and the mean Functional American Knee Society score was 90 (sd 17.5). The mean maximum flexion was 127° (90° to 145°). Significant elevation of the lateral joint line as measured by the proximal tibial varus angle (p = 0.04) was evident in the dislocation group when compared with the non-dislocation group.

Clinical results are excellent and short-term survival has improved when compared with earlier series. The risk of dislocation remains higher using a mobile-bearing UKR in the lateral compartment when compared with the medial compartment. Patients should be informed about this complication. To avoid dislocations, care must be taken not to elevate the lateral joint line.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 4 | Pages 619 - 619
1 May 2002
RODRIGUEZ JA


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 368 - 369
1 Mar 1999
NELISSEN RGHH VALSTAR ER ROZING PM


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 9 - 17
1 Jan 1992
Engh G Dwyer K Hanes C

We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation. Severe wear, with delamination or deformation, was observed in 51% of the implants, and was associated with time in situ, lack of congruency, thin polyethylene, third-body wear debris, and heat-pressed polyethylene. Significant under-surface cold flow was identified in some areas of unsupported polyethylene, and was associated with delamination in the load-bearing areas of thin inserts above screw holes in the underlying metal tray. We recommend the use of thicker polyethylene inserts, particularly in young, active patients and in designs with screw holes in the tibial baseplate. Thin polyethylene inserts which are at risk for accelerated wear and premature failure should be monitored radiographically at annual intervals.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal.


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.